CD Horizon™ ModuLeX™ Spinal System 55900105545Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SHANK 55900105545 5.5 X 45 CORTICAL · Class 2 · Implant

All models of CD Horizon ModuLeX Spinal System (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000435523), received between Aug 6, 2024 and Aug 6, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2024Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH SHANK HAVING TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) THERAPY FOR LUMBAR RADICULOPATHY AND SPONDYLOLISTHESIS. IT WAS REPORTED THAT PATIENT UNDERWENT AN L5-S1 MIS TLIF WITH MODULEX CORTICAL SCREWS, AND CATALYFT PL. DR PLACED CORTICAL TRAJECTORY SCREWS USING INTRAOPERATIVE IMAGING FOLLOWED BY S8 STEALTH NAVIGATION AND THEN PROCEEDED WITH A DIRECT DECOMPRESSION OF THE NURAL ELEMENTS FOLLOWED BY INSERTION OF AN INTERBODY DEVICE AT 5-1. PATIENT FOLLOWED UP ON (B)(6) 2024 WITH COMPLAINTS OF BACK PAIN AND A RETURN OF LEFT LEG PAIN. NORMAL BONE DENSITY WAS REPORTED. UPON IMAGING, IT WAS REVEALED THAT BILATERAL S1 SCREW SHANKS WERE FRACTURED. PATIENT DID UNDERGO A REVISION SURGERY WHICH INCLUDED AN ANTERIOR APPROACH, REMOVAL OF THE TLIF CAGE AND SUBSEQUENT INSERTION OF AN ALIF CAGE AND PLATE. THE CATALYFT CAGE WAS DISPOSED AT THE TIME OF REMOVAL. THE FRACTURED SCREWS REMAIN IN THE PATIENT AND POSTERIOR SCREWS MAY BE REMOVED AT A FUTURE DATE. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Pain

    MAUDE report 19918979 ↗