Skip to content

CD Horizon™ ModuLeX™ Spinal System 559200009Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SET SCREW 559200009 STERILE BREAK OFF · Class 2 · Implant

All models of CD Horizon ModuLeX Spinal System (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000435714), received between Nov 17, 2022 and Dec 16, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Break1
  • Device Damaged by Another Device1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Deformation1

Patient problems

  • Pain2
  • Foreign Body In Patient1
  • Intervertebral Disc Compression or Protrusion1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PATIENT UNDERGONE L2-S1 PSF WITH L4-S1 TLIF. IT WAS REPORTED THAT WHILE FINAL TIGHTENING OF THE SET CAPS ONE OF THE SET CAPS SPLAYED AND WAS UNABLE TO BE TORQUED DOWN, LEAVING THE RING OF THE SET CAP ATTACHED. THE BREAK OFF RING ON THE SET CAP IS STILL ATTACHED TO THE SET CAP AND IS IN THE PATIENT. NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23818912 ↗
  • MalfunctionMar 3, 2025Device Damaged by Another Device; Material Deformation; Device Damaged by Another Device; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OP DIAGNOSIS OF PATIENT WAS ADOLESCENT IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT, SET SCREWS WERE STRIPPED DURING FINAL TIGHTENING. HAD TO REMOVE AND REPLACE WITH NEW SET SCREWS. IT WAS FOUND THAT THE TIP OF POWEREASE BREAKOFF DRIVER LOOKS WORN AND IT WAS CONTRIBUTING TO STRIPPING OF SET SCREWS. PROCEDURE/TECHNIQUE PERFORMED WAS POSTERIOR SPINAL FUSION. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21502212 ↗
  • InjuryMar 3, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA DISTRIBUTOR REGARDING A PATIENT UNDERWENT SPINAL THERAPY. PRE-OP DIAGNOSIS OF PATIENT WAS ADOLESCENT IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT, PATIENT HAD SOME MILD INCREASE IN PAIN NEARLY FOR 4 MONTHS AFTER THE SURGERY. SET CAPS DID NOT HOLD THE RODS AS NEEDED. THE ROD CAME OUT OF THE SET CAP AND SCREWHEAD AFTER THE SURGERY. X-RAY PERFORMED ON (B)(6) 2025 SHOWS RODS NO LONGER HELD BY ORIGINAL CONSTRUCT AT L3. THE LEVELS IMPLANTED WERE T3-L3. PROCEDURE/TECHNIQUE PERFORMED DURING INITIAL SURGERY WAS ADOLESCENT IDIOPATHIC SCOLIOSIS SURGERY. PATIENT UNDERWENT REVISION SURGERY, AND THE REPORTED PRODUCTS WERE EXPLANTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Pain

    MAUDE report 21502034 ↗
  • InjuryJun 10, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR SPINAL FUSION, PSF SEGMENTAL INSTRUMENTATION FOR SCOLIOSIS. IT WAS REPORTED THAT THE ROD SLIPPED FROM THE BOTTOM SCREW OF A SCOLIOSIS CONSTRUCT. THE SET CAP IS STILL TIGHTENED TO THE SCREW. TULIP OR SET SCREW HALFWAY OUT OF ROD. REVISION SURGERY PERFORMED ON (B)(6) 2024 AND PRODUCTS WERE REPLACED. LEVELS IMPLANTED WERE T2-L2 IN INITIAL SURGERY AND L2-L3 IN REVISION SURGERY. PATIENT HAD PAIN. NO NEUROLOGICAL COMPLICATIONS NOTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Pain

    MAUDE report 19496432 ↗
  • InjuryNov 17, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A SET SCREW USED IN TLIF THERAPY. LEVELS IMPLANTED: L3-4 IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED WITH A REDUCTION OF A SPONDYLOLISTHESIS AT L3/4. AFTER THE SPONDYLOLISTHESIS WAS REDUCED, THE SURGEON COMPRESSED TO GAIN LORDOSIS AT THIS SEGMENT. IMAGES WERE TAKEN IN THE ROOM AFTER FINAL TIGHTENING AND IT SHOWED THAT THERE WAS A REDUCTION PERFORMED WITH THE ROD ABOVE AND BELOW THE SCREWS. FOLLOW-UP IMAGES WERE THEN TAKEN IN THE CLINIC ON (B)(6) 2022, WHICH SHOWED THAT THE LEFT - L3 SET SCREW CAME LOOSE AND WAS NO LONGER ATTACHED TO THE POLY HEAD OF THE LEFT - L3 SCREW. THIS FAILURE CAUSED THE LORDOSIS GAINED FROM THE COMPRESSION TO BE LOST AS WELL AS THE SPONDYLOLISTHESIS TO REOCCUR ONCE THE SET SCREW FAILED. AN EXPLANT SURGERY WAS PERFORMED TO CORRECT THIS ISSUE ON (B)(6) 2022. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.

    Patient problems: Intervertebral Disc Compression or Protrusion

    MAUDE report 15813358 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →