This Medtronic Midas Rex Legend AA10 is a pneumatic surgical drill attachment used in bone procedures. The 29 reports mostly describe the attachment overheating, being hard to insert or lock, and losing ball bearings or other parts, sometimes causing noise and vibration; none report patient harm. There are no recalls for this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00763000440220), received between Aug 21, 2024 and Jun 19, 2026 — classified as 29 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (12 reports).
9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device12
- Difficult to Insert10
- Detachment of Device or Device Component4
- Loss of Power4
- Electrical /Electronic Property Problem2
- No Apparent Adverse Event2
- Noise, Audible2
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WAS FOUND DURING EVALUATION..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25557162 ↗MalfunctionJun 15, 2026Difficult to Insert
IT WAS REPORTED THAT DEVICE IS NOT HOLDING THE BURR PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25498500 ↗MalfunctionJun 1, 2026Difficult to Insert; Vibration; Noise, Audible
IT WAS REPORTED THAT DEVICE, NOISY, VIBRATES, AND IS DIFFICULT TO LOCK. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25346196 ↗MalfunctionMay 29, 2026Difficult to Insert
IT WAS REPORTED THAT DURING USE THE ATTACHMENT CAN'T LOCK THE BUR. AT THE TIME OF REPORT, THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25334564 ↗MalfunctionMay 21, 2026Difficult to Remove
IT WAS REPORTED THAT THE ATTACHMENT FAILS TO RETAIN THE BURR PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARING WAS DETACHED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25239360 ↗MalfunctionMay 3, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE FROM ONE OF THE ATTACHMENTS THE BALL BEARING HAS FALLEN OUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25063104 ↗MalfunctionApr 3, 2026Difficult to Insert; Overheating of Device; Noise, Audible
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING, SOUND ROUGH, CANNOT HOLD BURR AND LOCKING MECHANISM FAULTY AFTER THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO KNOWN PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777018 ↗MalfunctionMar 25, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT CANNOT BE INSERTED - DETECTABLE BLOCKAGE. NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24684797 ↗MalfunctionJan 21, 2026Loss of Power
IT WAS REPORTED THAT ATTACHMENT HAS BEEN FROZEN. IT WAS REPORTED THAT THERE WAS NO KNOWN PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS IN TUBE WERE FOUND DURING EVALUATION.
Patient problems: Insufficient Information
MAUDE report 24130506 ↗MalfunctionNov 13, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT OVERHEATED WHEN ATTACHED TO THE MICROMOTOR. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23542738 ↗MalfunctionNov 10, 2025Overheating of Device; Loss of Power; Overheating of Device; Loss of Power
IT WAS REPORTED THAT THE ATTACHMENT IS OVERHEATING AND SEIZED. AT THE TIME OF REPORT, NO KNOWN PATIENT IMPACT REPORTED. ON FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23507957 ↗MalfunctionAug 12, 2025Difficult to Insert; Difficult to Insert
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DIFFICULTY WITH ASSEMBLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22780380 ↗MalfunctionMay 29, 2025No Apparent Adverse Event; No Apparent Adverse Event
REPAIR REQUEST INITIATED FOR THE DEVICE WITH THE REPORT OF NOT WORKING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO BEARING DETACHMENT. ON FOLLOW, IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22103102 ↗MalfunctionMar 19, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT HAS A BEARING PROBLEM. WHEN YOU PUT IT INTO OPERATION, THE HANDPIECE OVERHEATS. THERE WAS NO PATIENT INVOLVED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21639517 ↗MalfunctionMar 17, 2025Electrical /Electronic Property Problem; Overheating of Device; Electrical /Electronic Property Problem; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND SEIZED/NOT RUNNING. NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21616981 ↗MalfunctionMar 14, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21602945 ↗MalfunctionMar 7, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21549336 ↗MalfunctionMar 6, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21534676 ↗MalfunctionOct 7, 2024Loss of Power; Loss of Power
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF LOSS OF POWER. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20384123 ↗MalfunctionOct 4, 2024Difficult to Insert; Difficult to Insert
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF LOCKING MECHANISM DEFECTIVE. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20370935 ↗MalfunctionSep 26, 2024Loss of Power; Loss of Power
IT WAS REPORTED THAT THE DEVICE BROKEN AND THAT IT DOESN'T WORK. THERE WAS NO PATIENT INVOLVED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20313492 ↗MalfunctionSep 25, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT IT IS NOT POSSIBLE TO ROTATE AND SEIZED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20300056 ↗MalfunctionSep 24, 2024Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF ATTACHMENT DOES NOT WORK. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT DUE TO EVALUATION FINDING OF DETACHED BEARINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292448 ↗MalfunctionSep 24, 2024Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DIFFICULTY WITH ASSEMBLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT DUE TO EVALUATION FINDING OF BEARING INSIDE THE TUBE WAS DAMAGED .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20288890 ↗MalfunctionSep 17, 2024Electrical /Electronic Property Problem; Overheating of Device; Electrical /Electronic Property Problem; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND SEIZED/NOT RUNNING. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20234856 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.
- Z-1602-2016 · Apr 8, 2016 · TerminatedMedtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
- Z-0321-2016 · Oct 21, 2015 · TerminatedTool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
- Z-0546-2010 · Nov 30, 2009 · TerminatedMedtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
- Z-2156-2008 · May 30, 2008 · TerminatedMedtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
- Z-1574-05 · Sep 1, 2005 · TerminatedMedtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
- Z-1067-05 · Jun 29, 2005 · TerminatedMedtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
- Z-0278-05 · Nov 11, 2004 · TerminatedMedtronic Midas Rex Legend High Speed Pneumatic System
- Z-0213-05 · Nov 3, 2004 · TerminatedMedtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
- Z-0419-03 · Nov 11, 2002 · TerminatedLegend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page