MIDAS REX LEGEND AF02Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT AF02 LEGEND FOOTED F2-B1 2.4MM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic MIDAS REX Legend AF02 is a pneumatic drill attachment. The 39 reports mention parts wearing out, overheating, components detaching, trouble inserting it, power loss, and punctured material, with no deaths or injuries. There are no recalls naming this device. Reports rose to 24 in the last 12 months from 8 in the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00763000440381), received between Jan 3, 2024 and Sep 16, 2026 — classified as 40 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (18 reports).
9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn18
- Overheating of Device9
- Detachment of Device or Device Component7
- Difficult to Insert6
- Loss of Power3
- No Apparent Adverse Event3
- Material Puncture/Hole2
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request and no alleged product deficiencies were reported. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: Insufficient Information
MAUDE report 50786582 ↗MalfunctionJul 6, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT DRILL GOT STUCK IN THE CRANIOTOME. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25713675 ↗MalfunctionJul 6, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT DRILL GOT STUCK IN THE CRANIOTOME. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25713609 ↗MalfunctionJun 26, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, BEARING DAMAGE WAS OBSERVED WITH THE PRODUCT. THERE WAS NO PATIENT INOVLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636615 ↗MalfunctionJun 23, 2026Naturally Worn
IT WAS REPORTED THAT, PRIOR TO USE THE ATTACHMENT HAD INTERNAL BEARING BROKEN. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25590081 ↗MalfunctionJun 18, 2026Overheating of Device
IT WAS REPORTED THAT A HEATING ISSUE WAS OBSERVED WITH THE DEVICE DURING PRE-OPERATIVE PREPARATION FOR USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25542934 ↗MalfunctionJun 17, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE THE ATTACHMENT HAS BEARING DAMAGE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25530499 ↗MalfunctionJun 17, 2026Naturally Worn
IT WAS REPORTED THAT DURING PRIOR TO USE ATTACHMENT HAD A BEARING ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25530253 ↗MalfunctionJun 11, 2026Naturally Worn
IT WAS REPORTED THAT DURING MAINTENANCE OF THE ATTACHMENT, A BEARING DEFECT WAS IDENTIFIED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25467862 ↗MalfunctionJun 11, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF WORN. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25466202 ↗MalfunctionJun 9, 2026Naturally Worn
IT WAS REPORTED THAT DURING PRE USE THE ATTACHMENT HAD A BEARING ISSUE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BROKEN BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25443512 ↗MalfunctionJun 5, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, THERE WAS BEARING DAMAGE. AT THE TIME OF EVENT, THERE IS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED THAT BEARINGS WERE DETACHED DURING THE EVALUATION .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25404081 ↗MalfunctionMay 29, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DURING DEMO IN THE ATTACHMENT PARTS FALL OFF. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25330711 ↗MalfunctionMay 29, 2026Loss of Power; Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT HAS BEARING DAMAGE AND UNABLE TO RUN PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT FRACTURE OF BEARING FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25329116 ↗MalfunctionMay 29, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT BEARING WAS BROKEN. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25328119 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DEVICE, BEARING OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25298109 ↗MalfunctionMay 27, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT HAS BEARING PROBLEM PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS WERE BROKEN FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25297177 ↗MalfunctionMay 26, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT FOR THE DEVICE HAD BEARING DETACHMENT AND DETACHMENT OF COMPONENTS. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273708 ↗MalfunctionMay 4, 2026Material Puncture/Hole
IT WAS REPORTED THAT THE FRONT OF THE MILLING CUTTER IS BROKEN . AT THE TIME OF REPORT THERE WAS NO PATIENT INVOLVEMENT/IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25073089 ↗MalfunctionMay 3, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE FROM ONE OF THE ATTACHMENTS THE BALL BEARING HAS FALLEN OUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25063053 ↗MalfunctionApr 27, 2026No Apparent Adverse Event
IT WAS REPORTED THAT CUSTOMER HAD A DISSATISFACTION WITH THE ATTACHMENT. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24995735 ↗MalfunctionApr 23, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT DID NOT CLAMP THE CUTTING BLADES. THERE WAS NO PATIENT IMPACT AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24975158 ↗MalfunctionApr 6, 2026Difficult to Insert; Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE THE BEARINGS WERE DAMAGED . AT THE TIME OF REPORT NO PATIENT WAS INVOLVED. ADDITIONAL INFORMATION RECEIVED STATING THAT THE WAS TOOL CAN NOT BE INSERTED IN THE ATTACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24787246 ↗MalfunctionApr 3, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE EXHIBITED BROKEN BEARINGS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777667 ↗MalfunctionMar 31, 2026Naturally Worn
IT WAS REPORTED THAT THE DEVICE BEARING WAS BROKEN. IT WAS REPORTED THERE WAS NO REPORTED PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATED THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24740438 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.
- Z-1602-2016 · Apr 8, 2016 · TerminatedMedtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
- Z-0321-2016 · Oct 21, 2015 · TerminatedTool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
- Z-0546-2010 · Nov 30, 2009 · TerminatedMedtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
- Z-2156-2008 · May 30, 2008 · TerminatedMedtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
- Z-1574-05 · Sep 1, 2005 · TerminatedMedtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
- Z-1067-05 · Jun 29, 2005 · TerminatedMedtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
- Z-0278-05 · Nov 11, 2004 · TerminatedMedtronic Midas Rex Legend High Speed Pneumatic System
- Z-0213-05 · Nov 3, 2004 · TerminatedMedtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
- Z-0419-03 · Nov 11, 2002 · TerminatedLegend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page