Skip to content

INFINITY™ Occipitocervical Upper Thoracic System 3600215Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SET SCREW 3600215 M6 SETSCREW · Class 2 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000473921), received between Jul 11, 2022 and May 15, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (K163375); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device2
  • Device Dislodged or Dislocated2
  • Material Deformation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR CERVI CAL C4-T3. IT WAS REPORTED THAT A SET SCREW WAS FOUND IN A X-RAY OF THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT AN INFINITY SET SCREW IS VISIBLE ON AN X-RAY NOT IN THE TULIP OF A SCREW. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WAS NO ADDITIONAL SURGERY PERFORMED AND AT THIS TIME NO ADDITIONAL SURGERY IS PLANNED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22025756 ↗
  • MalfunctionFeb 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION RECEIVED FROM THE HEALTH CARE PROFESSIONAL VIA A MANUFACTURING REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT DURING A ROUTINE CHECK-UP AT A 6-WEEK POST-OPERATIVE APPOINTMENT, TWO SET SCREWS FROM THE MULTI-AXIAL SCREW (MAS) WERE DISCOVERED TO HAVE POPPED OUT, AS REVEALED BY AN X-RAY. THE SET SCREWS AT C3 (RIGHT) AND T1 (LEFT) WERE IDENTIFIED AS HAVING POPPED OUT. THERE WERE NO REPORTED SYMPTOMS OR DISCOMFORT FROM THE PATIENT, AND NO INCIDENTS SUCH AS FALLS OR INJURIES WERE NOTED PRIOR TO THE DISCOVERY. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THIS EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THERE WAS NO PLAN/SCHEDULE FOR THE REVISION SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21372734 ↗
  • MalfunctionJul 11, 2022Device Damaged by Another Device; Material Deformation; Device Damaged by Another Device; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL REGARDING A PATIENT HAVING C1/C2 POSTERIOR FIXATION AND DECOMPRESSION. IT WAS REPORTED THAT USE OF 3 NM TOQUE LIMITING HANDLE ON 2 NM DRIVER SHAFT AND SET SCREWS RESULTED IN ROUNDING OF BOTH THE SET SCREWS AND THE DRIVER SHAFT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14974726 ↗
  • MalfunctionJul 11, 2022Device Damaged by Another Device; Material Deformation; Device Damaged by Another Device; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL REGARDING A PATIENT HAVING C1/C2 POSTERIOR FIXATION AND DECOMPRESSION. IT WAS REPORTED THAT USE OF 3 NM TOQUE LIMITING HANDLE ON 2 NM DRIVER SHAFT AND SET SCREWS RESULTED IN ROUNDING OF BOTH THE SET SCREWS AND THE DRIVER SHAFT. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14974725 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K163375. One submission often covers many models, so check the product and lot details in each record.

  • Z-1410-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1413-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1412-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1409-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1411-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.