Quick summary

This Medtronic InterStim X implanted neurostimulator is used to treat fecal incontinence. The reports often describe the device not delivering stimulation as expected, moving out of place, and patients reporting pain or electric shock sensations. FDA records list 5,994 reports, no deaths, 2,644 injuries and 3,350 malfunctions. No recalls name this device. Reports fell from 1,616 in the prior 12 months to 862 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,944
Deaths reported
0
Injuries reported
2,623
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5,944 adverse event reports that list this exact device (DI 00763000484668), received between Mar 8, 2022 and Sep 30, 2026 — classified as 2,623 injuries, 3,321 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1,313 reports).

7 other recalls cover devices cleared under the same FDA submission (P080025, P970004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 111 reportsJan 2024: 144 reportsFeb 2024: 143 reportsMar 2024: 142 reportsApr 2024: 160 reportsMay 2024: 135 reportsJun 2024: 180 reportsJul 2024: 187 reportsAug 2024: 171 reportsSep 2024: 223 reportsOct 2024: 245 reportsNov 2024: 177 reportsDec 2024: 113 reportsJan 2025: 223 reportsFeb 2025: 156 reportsMar 2025: 139 reportsApr 2025: 134 reportsMay 2025: 139 reportsJun 2025: 121 reportsJul 2025: 115 reportsAug 2025: 133 reportsSep 2025: 80 reportsOct 2025: 86 reportsNov 2025: 105 reportsDec 2025: 103 reportsJan 2026: 110 reportsFeb 2026: 105 reportsMar 2026: 106 reportsApr 2026: 67 reportsMay 2026: 69 reportsJun 2026: 51 reportsJul 2026: 19 reportsAug 2026: 30 reportsSep 2026: 47 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026245
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1,313
  • Adverse Event Without Identified Device or Use Problem1,201
  • Migration or Expulsion of Device893
  • Energy Output Problem861
  • Failure to Deliver Energy745
  • Inappropriate/Inadequate Shock/Stimulation646
  • Unintended Collision482
  • Electromagnetic Compatibility Problem475

Patient problems

  • Insufficient Information1,840
  • No Clinical Signs, Symptoms or Conditions1,744
  • Pain1,367
  • Incontinence1,131
  • Discomfort664
  • Electric Shock638
  • Urinary Frequency / Polyuria565
  • Urinary Retention498

Recent adverse event reports

The 25 most recent of 5,944 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Premature Discharge of Battery

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for unknown indications for use. It was reported that they had some surgery up at Cleveland Clinic and they were going to turn off the MRI turnoffs that stimulated for the MRI, but it wouldn't work at all. The patient said earlier this year they had an MRI of their prostate, and it worked fine, but between May and September 22nd when they had the surgery, the battery died in the device. Patient asked why the battery would have expired so soon. Agent reviewed battery longevity information with the patient. Patient mentioned that their previous implant always used up and even in the old one that was not MRI compatible, they always had to use about 4.2 source of stimulation. The issue was not resolved through troubleshooting. The patient was redirected to their healthcare provider to further address the issue.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950306 ↗
  • InjurySep 30, 2026Disconnection; Migration or Expulsion of Device; Insufficient Device Problem Information

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that they weren't sure if they were implanted with a new battery, but they were given a new modulator (handset and communicator) to use, and they were told that someone from the manufacturer would be calling them to set the settings up because, apparently, after they put in the INS, they only set it up to a low level. The patient said the new INS was put in on (b)(6)2026. The patient also wanted to check if they were given a new INS. The patient said that the original INS was put in in (b)(6) 2023. The patient said that the packet of the original INS had come loose, so it was kind of moving around and it was causing pain, so they went to their healthcare provider (HCP). The patient said their HCP had gone ahead and created a new pocket, but the HCP said they would use the same battery (INS), but they didn't know because they thought when they gave them a new handset and communicator, they also had a new battery (INS). The agent then walked the caller through connecting to the INS and checking the serial numb

    Patient problems: Insufficient Information; Pain; Urinary Frequency / Polyuria

    MAUDE report 50947493 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction: urge incontinence It was reported that  at least over a year ago they noticed the therapy was not working so they stopped using it. The patient ended up having the INS explanted on (b)(6) 2026. The patient's reason for calling was to ask what they should do with the external equipment. The agent reviewed disposal guidelines and documented reported event.

    Patient problems: Insufficient Information

    MAUDE report 50936298 ↗
  • InjurySep 29, 2026Battery Problem; Communication or Transmission Problem; Insufficient Device Problem Information

    Information was received from a friend/family member regarding a patient who was implanted with an implantable neurostimulator (INS) for unknown indications for use. It was reported that the caller reports that the patient has not used the implant because it has not been doing what it was supposed to be doing since within 6 months of being implanted in 2022. They tried turning it up until the point where it hurts and then back it back down. The caller reported that the patient also has Alzheimer's, so even if the device was working, they know what the impact would be with the Alzheimer's. The patient was scheduled for an MRI yesterday, but when they tried to activate MRI mode they saw a screen that the implant was too low and needed to be charged and they could not due the MRI. Reviewed that the implant is recharge free. The caller noted that the light on the communicator was yellow. Reviewed that it was likely the communicator that needed to be charged. Caller was not with the patient at the time of the call, but will call back to confirm what the screen was. Additional information was received from a friend/family member. The patient representative called back repeating event his

    Patient problems: Incontinence; Pain

    MAUDE report 50935485 ↗
  • InjurySep 29, 2026Failure to Deliver Energy; Energy Output Problem

    Additional information was received from the patient. They reported that they already had their external devices with them and were able to access their therapy settings. The patient considered changing the program and decreasing the therapy settings to a comfortable level. It was reported that the cause of the therapy being off was determined. It was a mistake and was left off after an MRI. No further action would be taken.

    Patient problems: Discomfort

    MAUDE report 50935156 ↗
  • InjurySep 29, 2026Inappropriate/Inadequate Shock/Stimulation

    Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urinary urge incontinence. It was reported that they had a patient coming in for an MRI today and they wanted to confirm the model of the device to make sure the patient was eligible for the shoulder MRI. The caller said the patient felt a big shock when they turned on their implant. The caller said the patient hadn't used therapy in a long time because they felt it wasn't working for them. The caller didn't know how long therapy had not worked for the patient, or how long therapy had been off. The patient was redirected to their healthcare provider to further address the issue. The MRI technician called with the patient later the same day and they stated that the patient felt a shocking sensation just by trying to connect to the implant, thus the patient was hesitant to try again. Technical Services (TS) reviewed that the process of connecting to the implant should not cause a shocking sensation. The patient proceeded to try again and they were able to enter MRI mode, which showed full body eligible, without shocking

    Patient problems: Electric Shock

    MAUDE report 50934364 ↗
  • InjurySep 29, 2026Electromagnetic Interference; Insufficient Device Problem Information

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urinary urge incontinence. It was reported that they'd had their bladder control bowel control stimulator for quite a few years. For some reason they didn't feel as if it was working anymore. They'd been having very bad episodes and they were wondering how they could go about having this checked. They were out of town now, but they were sure there would be a representative here in Rhode Island that could check. An agent responded on (b)(6) 2026, recommending the patient work with their physician. The agent reviewed the role of the manufacturer representative (rep) and provided the Patient Services contact information for questions about their device, general program use, therapy adjustments, etc. The patient called on (b)(6) 2026 and reported that there was something wrong with their therapy. They were having symptoms they have never had before. No matter what they did they had diarrhea all the time, and an extreme amount of gas. The symptoms started about three weeks ago. They said they had had no falls or accidents. They had only had CT scans recen

    Patient problems: Diarrhea; Flatus; Insufficient Information

    MAUDE report 50934101 ↗
  • InjurySep 29, 2026Impedance Problem

    Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the reason for the call was the representative said they were doing a lead revision. The representative said they put the old battery on the new lead and it gave different impedances than with the old lead.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932659 ↗
  • MalfunctionSep 29, 2026High impedance

    Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction. It was reported that the reason for the call was the representative said they were doing a lead revision (see 709415697) and they tested the new lead and 2 of the electrodes came back with high impedance every time. The representative said they put the old battery on and it gave different impedances. The 1 and 2 electrodes were over 4,000 ohms. The representative put the new battery back in and they gave the same value of being high. The representative confirmed they tightened the set screw and tugged on the lead to make sure it was secure. The representative had not tried getting motor response from the lead with the external battery. During the call, the representative was able to get motor response on all 4 electrodes. Electrode 1 & 2 that had high impedances got motor response at 0.4 mA. Technical Services (Ts) reviewed with caller that the impedance was good, even though it registered high. Ts stated that if impedance was truly high then the patient would not have gotten motor response at 0.4 mA. The issue wa

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932642 ↗
  • InjurySep 29, 2026Electromagnetic Interference; Electromagnetic Compatibility Problem; Therapeutic or Diagnostic Output Failure

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for Urinary/bowel dysfunction. It was reported that they had an EMG last week and turned therapy off to be on the safe side but when they turned it back on they got the message: The System is Operating at Maximum Settings and their setting was program A, 1.5, they have not tried any other programs. The patient (Pt) said their note about program A said, up to 5.0 but pt did not recall what the amplitude had been before they turned it off for the EMG and pt said they were not sure if the Max Settings message appeared when they turned it off or not because they had been very nervous. Reviewed meaning of Max Settings message and asked pt to synch to their INS and navigate to the battery section, their MyBattery showed OK. Pt said they have not had any falls, they had several MRIs. The Pt also reported their symptoms were 50% improved over their baseline but their results were not the same as they were during the evaluation. The Pt said they still have to wear the heaviest pads all the time and run to the bathroom. The Pt said in the 3 months following the implant procedure they met with

    Patient problems: Anxiety; Incontinence

    MAUDE report 50927552 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urge incontinence. It was reported that caller reported a return of symptoms for the past 3-4 days. Caller stated that they have been having to use their catheter every day, multiple times a day (implanted for urinary retention). Caller added that sometimes it happens for a day or two and then they don't need to use the catheter but it never happens consistently like this. Caller is wondering if the implant battery could have reached the end of its life. Caller stated they were going to use the handset to check the status or possible make an adjustment but they cant because the handset screen is damaged and it wont power on. Caller is worried that it might cause a UTI (stating their UTIs are severe and resistant and require IV antibiotics for 2 weeks) and they are trying to prevent that. The patient was redirected to their healthcare provider to further address the issue. screen is damaged and it wouldn't respond when trying to power on. redirected to sunMed portal called prs to update phone

    Patient problems: Urinary Retention

    MAUDE report 50927071 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and fecal incontinence. It was reported that the reason for the call was to report no therapeutic benefit. The patient stated that the device had never provided meaningful relief since it was implanted and they plan to have it removed. The patient also stated that they do not currently have a physician who can remove it. Agent sent listings. Caller stated they are considering having a colostomy bag placed instead. Additionally, the patient commented that electronic devices such as this should not be given to elderly individuals. The patient was redirected to their healthcare provider.

    Patient problems: Insufficient Information

    MAUDE report 50918285 ↗
  • InjurySep 28, 2026Premature Discharge of Battery; Battery Problem

    On 2026-Oct-20 additional information was received from a healthcare provider (HCP). The HCP requested a manufacturer rep to check the patient's INS. The caller stated the patient contacted the office today to let them know they saw a \"light on their device\" with message of \"low battery\" and caller would like rep to see if this is accurate. Technical services redirected the caller to the National Answering Service to page a rep. The patient called back regarding the low internal battery warning for the implanted device. The patient stated the nurse at their Urology office told them to contact the manufacturer to see what the battery percentage was. The agent reviewed we didn't have a way to remote in to see that information and redirected to the HCP for further assistance.

    Patient problems: Incontinence

    MAUDE report 50917655 ↗
  • InjurySep 28, 2026Insufficient Device Problem Information

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction : Urinary Urge Incontinence The patient reported that they have the device placed in the right back.¿ They stated, ¿6 months ago they experienced burning in the right calf.¿ They went to their provider, and ¿a  representative lowered the setting, which helped with the burning. The burning gets worse when they have their leg hanging; it is better when they keep their leg straight.¿ The device is turned off ¿because the burning was going up to their right buttock.¿ The patient declined a new/list of providers,- will self-schedule,<(>,<)> and was provided with the number for an MDT Patient Representative.

    Patient problems: Burning Sensation

    MAUDE report 50916759 ↗
  • InjurySep 26, 2026Failure to Deliver Energy; Data Problem

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and fecal incontinence. It was reported that maybe within the last 2 weeks they noticed they were feeling like the therapy was not working as well as it had been. Patient said it was such a huge improvement and 95% was gone but in the last couple of weeks it's happened a lot when they've gone for a walk and almost every time they have had an accident which they hadn't had in months. Patient said today they went to check it and it said therapy was turned off. Pt said they have not turned it off nor had any other medical tests or procedures. Pt said when they turned it back on, it was at 1.5 which they don't remember ever setting it that high and it was really painful so they turned it OFF. and that stopped. Reviewed some therapy education information and during the call, patient was successful to sync with their implant and increase stim to a comfortable level. Patient will maintain stimulation level and will cont inue to track symptoms. Redirect to doctor if issue persists. The patient mentioned other medical history. This included patient said lately ,

    Patient problems: Discomfort; Incontinence; Sleep Dysfunction

    MAUDE report 50906778 ↗
  • InjurySep 25, 2026Energy Output Problem; Electromagnetic Compatibility Problem

    Information was received from a friend/family member regarding a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that the patient had been in pain for a while. The pain was going more outside of the bicycle area. It was going up the left side, down their butt area, and down their leg. The issue had been going on since the patient had major back surgery in the middle of July. The patient representative called the manufacturer representative (rep) and talked to them on the phone they think Monday (b)(6) 2026. The rep told the caller to shut the therapy off for 24-48 hours until the pain was totally gone and then try a different program. Yesterday they tried to turn the therapy back on and they went through every single program of the device, and all the programs had pain on all of them. The agent reviewed with the caller how they changed programs. They selected different program numbers but never selected done where it actually switched to a different program. The whole time the patient was still on the same program. The caller said they didn't know if there was something wrong with the w

    Patient problems: Pain

    MAUDE report 50905833 ↗
  • InjurySep 23, 2026Migration or Expulsion of Device

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for fecal incontinence and urinary/bowel dysfunction. It was reported that their device had not been helpful. They continued to have urinary incontinence. They did not use this device anymore because it does not help. Additional information was received from the patient. They reported that the lack of effect was since the implant. The doctor said they should contact their rep. The therapy was still not working, but it was still in their lower back, and the doctor said they should take it out. They have a new doctor and he helps a lot. They have not explanted the device

    Patient problems: Urinary Incontinence

    MAUDE report 50872714 ↗
  • InjurySep 23, 2026Migration or Expulsion of Device; Extracardiac stimulation

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that they lost a bunch of weight so their stimulator had moved too low and was in the wrong place. The patient said that they felt stimulation in the wrong place so they called their manufacturer representative (rep) and the rep told the patient that since their symptom control was sufficient (better than it was before getting the device) they shouldn't change the settings. The rep told the patient that they would run out of programs and not to mess with the equipment. The patient said they had never messed with the settings until they had the issue with stimulation in the wrong spot that started the past few months so the patient ended up changing programs on their own so they could be at a setting where they felt stimulation in the bicycle seat. The patient said that they had been feeling stimulation at their implant site. During the call the patient adjusted therapy to where they felt stimulation in the bicycle seat. The patient was redirected to their healthcare provider to further address the issue. Add

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50872444 ↗
  • InjurySep 18, 2026Migration or Expulsion of Device; Inappropriate/Inadequate Shock/Stimulation; Malposition of Device

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urge incontinence. It was reported that yesterday while at the grocery store, all of a sudden started to get electrical socks in their left buttock and then to the right side of their vagina and it knocked them down to their knees. Patient said had their external devices however they weren't charged so patient went to the care to charge up the external devices. Patient said that as soon as they sat down the shocking stopped. Patient recharged the external devices, connected to the implant and turned therapy off. Patient said received implant to help with pain in their bladder and to go to the bathroom. Patient said that without the stimulator she just don't void. Patient said hasn't yet voided today however said does have catheters. Patient mentioned concern that the wire might have moved. When asked, no falls or trauma or heavy lifting. Patient said that has lost weight and the stimulator sticks out and catches their clothes on it all the time. Patient said that the stimulator site is also more tender and painful. Patient said that has an appointmen

    Patient problems: Electric Shock; Fall; No Clinical Signs, Symptoms or Conditions; Pain; Weight Changes

    MAUDE report 50830955 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery; Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820713 ↗
  • InjurySep 18, 2026Inappropriate/Inadequate Shock/Stimulation

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urge incontinence. It was reported that their bladder stim was turned up too high and it was hurting like crazy which started about a week ago. Prior to the hurting the patient had increased stim because they felt like the device wasn't working as well as it should . Patient was getting a message to charge the communicator. Patient will charge communicator then call back if needed. Additional information was received from the patient on (b)(6) 2026. The patient reported ongoing concerns with therapeutic effect and painful stimulation sensation. The patient was calling from their hospital room, and reported they were admitted for lower right abdominal pain in the same area as their kidney and the patient was concerned the pain was coming from the kidney. The patient noted the pain was severe and they felt it on both back and front of the body. The patient got interitem therapy to help with urinary retention, and shared that prior to implant when they would use a catheter they would drain 2 liters of urine from their bladder. About an hour ago the pati

    Patient problems: Electric Shock; Insufficient Information; Pain; Paresthesia; Urinary Retention

    MAUDE report 50819630 ↗
  • InjurySep 17, 2026Inappropriate/Inadequate Shock/Stimulation

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that their current implant  had never worked well for them. Patient said last night they had to get up 4 times to go to the bathroom almost every 2 hours and it was kind of scary because they had never had that happen before. Patient said that they had talked to the manufacturer representative (rep) about their lack of therapeutic benefit and was told that if the patient had tried all the different settings there was nothing else they could help the patient with. During the call patient made an adjustment to therapy, patient increased stimulation and said they felt stimulation at their implant site. Patient said that they had felt stimulation at their implant site before. Patient said sometimes it felt more like a shock at their implant site so they would have to turn down stimulation. Patient said they had a lot of things on their plate so they would try to get an appointment with their managing physician in October. Patient said they do have an appointment with their managing doctor in February but that wa

    Patient problems: Electric Shock; Sleep Dysfunction; Urinary Frequency / Polyuria

    MAUDE report 50810042 ↗
  • InjurySep 17, 2026Break; Insufficient Device Problem Information

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary urge incontinence and urinary/bowel dysfunction. It was reported that patient had a lead issue and also stated that if the lead would highly be broken again, then it would the second time it would happen.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808860 ↗
  • MalfunctionSep 15, 2026Migration or Expulsion of Device; Unstable; Malposition of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50778758 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    Information was received from a patient who was implanted with an implantable neurostimulator (INS) for urinary/bowel dysfunction and urinary urge incontinence. It was reported that the therapy worked for their symptoms for a couple of months after first getting it, however after that, it just started all over again with their symptoms the way they used to be and every time they went to their managing healthcare provider (HCP), the HCP would tell them everything was working fine and they needed to make more adjustments and the x-rays showed everything was fine and the wires were in place however after a year of going to the HCP the HCP said to the patient they could change out the wire so on (b)(6)2025, of last year the HCP replaced the lead wire and they were released from the hospital at 3:00 pm.

    Patient problems: Insufficient Information

    MAUDE report 50778311 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P080025 on Mar 14, 2011. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Sacral Nerve Stimulation Therapy System”. The FDA review took 899 days. The FDA has since approved 350 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗

Approved through premarket approval (PMA) P970004 on Sep 29, 1997. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Therapy System For Urinary Control”. The FDA review took 242 days. The FDA has since approved 449 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗

The FDA regulates product code QON (implanted electrical device intended for treatment of fecal incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P080025, P970004. One submission often covers many models, so check the product and lot details in each record.

  • Z-1704-2025 · Apr 3, 2025 · Open, Classified
    Enhanced Verify Evaluation Handset (CFN HH90130FA)
  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-2365-2023 · Jul 6, 2023 · Open, Classified
    Medtronic InterStim Ground Pad REF 041826
  • Z-0998-2022 · Feb 28, 2022 · Terminated
    MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-0040-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim
  • Z-0037-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta

Common questions about the InterStim™ X 97800

Has the InterStim™ X 97800 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000484668), based on the FDA recall file dated Nov 27, 2026. However, 7 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the InterStim™ X 97800?

The FDA's MAUDE database lists 5,944 adverse event reports for this device, received between Mar 8, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Migration or Expulsion of Device” (893), “Energy Output Problem” (861) and “Failure to Deliver Energy” (745); the most common patient problems are “Pain” (1,367), “Incontinence” (1,131) and “Discomfort” (664). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the InterStim™ X 97800?

None of the reports are classified as deaths, 2,623 as injuries and 3,321 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the InterStim™ X 97800 FDA approved?

It was approved through premarket approval (PMA) P970004 on Sep 29, 1997; its record lists 2 FDA submissions in total.

Is the InterStim™ X 97800 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the InterStim™ X 97800, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand InterStim™ X, model 97800). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Implanted electrical device intended for treatment of fecal incontinence devices

All Implanted electrical device intended for treatment of fecal incontinence devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page