Axonics 1201Recalls & reported problems
Axonics Modulation Technologies, Inc. · Tined Lead Kit · Class 3 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Axonics tined lead is a thin implanted lead that stimulates a nerve to treat fecal incontinence. Many of its 490 reports describe leads that moved out of place, high electrical resistance and fractures. Recent injury reports describe lost therapy, warning lights on the remote and leads being revised or replaced. The reports include 350 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 492
- Deaths reported
- 0
- Injuries reported
- 352
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 492 adverse event reports that list this exact device (DI 10810005340141), received between Feb 13, 2020 and Sep 29, 2026 — classified as 352 injuries, 140 malfunctions.
The most frequently reported specific device problem is “Migration” (198 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Migration198
- High impedance133
- Adverse Event Without Identified Device or Use Problem98
- Fracture44
- Communication or Transmission Problem31
- Impedance Problem25
- Material Twisted/Bent17
- Lack of Effect10
Patient problems
- Failure of Implant112
- No Clinical Signs, Symptoms or Conditions102
- Incontinence100
- Pain53
- Urinary Incontinence38
- Discomfort37
- Undesired Nerve Stimulation35
- Unspecified Infection22
Recent adverse event reports
The 25 most recent of 492 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Migration
It was reported that the patient with this tined lead underwent a replacement of the lead due to it being twisted inside of the patient. The patient was not experiencing symptom relief as a result. The procedure was completed as planned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941141 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with the neurostimulator underwent an explant due to inadequate stimulation of the lead.
Patient problems: Inadequate Pain Relief
MAUDE report 50937699 ↗InjurySep 23, 2026Migration
It was reported that a patient with an SNM Tined Lead saw no symptom improvement after implantation. Lead migration was confirmed via X-ray. A successful revision was performed by replacing the patient's tined lead. There was no additional patient complications reported.
Patient problems: Urinary Incontinence
MAUDE report 50870860 ↗InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient reported a lack of symptom relief despite troubleshooting and reprogramming multiple times. The tined lead was explanted and replaced with a new lead on the opposite site. The lead placement was confirmed to appear fine through a radiograph. The patient later was more relaxed with some motor responses.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50823644 ↗InjurySep 18, 2026Migration
It was reported that the patient said their device had not been working well. They had an x-ray and it showed their Tined Lead migrated. The patient had surgery to revise their Tined Lead. The patient is doing well, and their symptoms have improved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817284 ↗InjurySep 16, 2026Difficult or Delayed Positioning; Malposition of Device
It was reported that a patient with a Recharge-Free SNM IPG stated that their device never worked despite multiple adjustments and reprogramming. It was noted that getting a caudal curve on the lead was unsuccessful and motor responses were undesirable during the initial implantation. The patient elected have a revision with their lead and battery replaced. No additional patient complications were reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788911 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with this Implantable Neurostimulator (INS) Tined Lead has not felt any symptomatic relief since implantation. The patient is still waking up 6-8 times to void. Troubleshooting was performed in programing appointments, testing multiple electrodes and configurations with no success. A device revision was performed, where the lead was replaced and new lead implanted on the contralateral side. No other patient complications were reported as a result of this event.
Patient problems: Urinary Frequency / Polyuria
MAUDE report 50756209 ↗InjurySep 10, 2026High impedance
It was reported that the patient with the neurostimulator (INS) underwent tined lead and INS revision due to pain and an open impedance. There were no patient complications.
Patient problems: Pain
MAUDE report 50722576 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient with a SNM Tined Lead underwent a planned lead revision procedure due to reduced therapeutic benefit compared to that experienced during the trial period. No further patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50719172 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with this Tined Lead turned their device off due to rectal pain and a sensation they described as their foot going to sleep. They also experienced lower back pain and a warm sensation around their implant site. The local care team member confirmed the patient turned off their device. With the stimulation off, the patient said the sensation was still present but better. The patient reported their discomfort got worse when lying or sitting down, but the discomfort was constant. The patient stated they did not experience this during the trial but did experience the discomfort from the procedure. The patient denied experiencing any falls and denied discomfort after adjusting stimulation. The patient believed the implant damaged their nerve, but they were educated that the lead was not attached to a nerve but was placed around it. Additionally, the physician said the patient did okay but still experience the discomfort. The physician does not believe that the concerns the patient had were related to the device. Later, the patient had surgery to explant their Tined Lead and Neurostimulator. The patient was not implanted with a new device. Per the physicia
Patient problems: Burning Sensation; Discomfort; Inappropriate Device Stimulation of Tissue; Pain
MAUDE report 50712955 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with the neurostimulator underwent lead revision due to the patient's overactive bladder symptoms not being controlled. No further patient complications were reported.
Patient problems: Incontinence
MAUDE report 50712954 ↗InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem
It was reported the patient experienced infection and pain following a bilateral advanced trial. The right and left lead were both tunneled to the future Neurostimulator (INS) site on the right side. A week into the trial, it was noted by the Clinical Specialist (CS) that the patient had 75 percent success on the left lead (with the right lead off). Then the patient was switched to the right lead on (with the left lead off). The patient experienced pain in the right incision site and the nurse practitioner assessed the site. Per the physician, there were no signs of infection noted. The stimulation was high. The stimulation was turned off, and the pain started to improve right away. When the Clinical Specialist followed up with the patient, they stated with the stimulation off, they were still having very bothersome pain. The pain was described near their hip, upper buttocks on right side. The stimulation remained off and the physician was updated. The plan was to move the INS site to the left side for the Stage 2 procedure and keep the left lead (which was successful). During the Stage 2 procedure, in pre-op, the patient stated they have pain at the right INS site. The patient wan
Patient problems: Pain; Purulent Discharge; Unspecified Infection
MAUDE report 50698736 ↗InjurySep 9, 2026High impedance; Communication or Transmission Problem
It was reported that the patient had a red light on their remote control. The patient attempted a readiness check for an MRI, but they kept getting the red light on the remote control. Troubleshooting attempts were made with the SNM Patient Care team member but they got the same result. The patient had to reschedule their MRI and meet with a local care team member to troubleshoot in person. Additional information reported that the patient had an impedance check which showed high impedance. The patient received an x-ray, and it showed the lead was in a good position. Later, the patient had surgery to replace only their Tined Lead. The patient is doing well post-revision.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50698735 ↗InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem
It was reported the patient experienced infection and pain following a bilateral advanced trial. The right and left lead were both tunneled to the future Neurostimulator (INS) site on the right side. A week into the trial, it was noted by the Clinical Specialist (CS) that the patient had 75 percent success on the left lead (with the right lead off). Then the patient was switched to the right lead on (with the left lead off). The patient experienced pain in the right incision site and the nurse practitioner assessed the site. Per the physician, there were no signs of infection noted. The stimulation was high. The stimulation was turned off, and the pain started to improve right away. When the Clinical Specialist followed up with the patient, they stated with the stimulation off, they were still having very bothersome pain. The pain was described near their hip, upper buttocks on right side. The stimulation remained off and the physician was updated. The plan was to move the INS site to the left side for the Stage 2 procedure and keep the left lead (which was successful). During the Stage 2 procedure, in pre-op, the patient stated they have pain at the right INS site. The patient wan
Patient problems: Pain; Purulent Discharge; Unspecified Infection
MAUDE report 50698224 ↗InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with the neurostimulator (INS) experienced pain, discomfort, and the battery portion being exposed at the incision site. The patient underwent a revision of the device. There were no additional patient complications.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Erosion; Pain
MAUDE report 50695242 ↗InjurySep 9, 2026Fracture; Communication or Transmission Problem; Impedance Problem; Migration
It was reported that a patient with this Implantable Neurostimulator (INS) contacted the clinic requesting assistance with switching therapy programs after observing red indicator lights on programs P1 and P2 on the patient remote. Upon evaluation, an impedance issue affecting all four electrodes was identified. The patient reported that a loss of therapy effectiveness had been present prior to experiencing a recent fall. Visual and physical assessment indicated evidence of possible lead migration and or fracture, as the lead was raised and palpable at the skin level, with no evidence of skin penetration. The physician reported that both the lead and INS were potentially impacted. The patient subsequently underwent explantation and replacement of the system. No further patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50695017 ↗InjurySep 8, 2026Fracture
It was reported that the patient with this Tined Lead had it fractured due to the patient fiddling with it. The physician said this patient is a twirler and fiddled with their Neurostimulator (INS) and twisted it so much that is caused the lead to fracture. The patient had surgery to revise their Tined Lead. The patient is doing well.
Patient problems: Twiddlers Syndrome
MAUDE report 50674496 ↗InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient with an SNM Tined lead was experiencing a sharp sensation vaginally and pain at their implant site, regardless of the device being on or off. Troubleshooting was performed by field representatives through attempting multiple programs with different electrode configurations without success. The patient had their device explanted. No further information was provided.
Patient problems: Implant Pain; Inappropriate Device Stimulation of Tissue
MAUDE report 50616908 ↗InjurySep 4, 2026Communication or Transmission Problem; Impedance Problem
It was reported that the patient with the neurostimulator is experiencing urinary incontinence and inadequate stimulation. The patient was met in the office and discovered a red light on their remote and an open impedance. The device was reprogrammed. The error message was cleared, and the patient underwent a lead revision on (b)(6) 2026. There were no additional patient complications.
Patient problems: Inadequate Pain Relief; Urinary Incontinence
MAUDE report 50609682 ↗InjurySep 4, 2026Migration
It was reported that a patient with a Rechargeable SNM IPG (IPG) and SNM Tined Lead was experiencing increased episodes of incontinence. The patient recently felt their IPG rotate and return to a normal position when getting out of their car. When charging the IPG afterwards, the IPG site became hot and causing a burning sensation. Shortly after, the patient used the restroom and experienced painful spasms in their leg and vagina. Therefore, stimulation was turned off; however, the spasms continued. The patient went to the emergency room ER, where X-rays were taken that confirmed slight lead migration. The patient had their device removed. There was no further information provided.
Patient problems: Burning Sensation; Cramp(s) /Muscle Spasm(s); Pain; Urinary Incontinence
MAUDE report 50600240 ↗InjurySep 4, 2026Material Integrity Problem
It was reported that a patient with an SNM Tined Lead stated their device was not working after they fell on their lead and battery site. Reprogramming was done but was unsuccessful. X-rays were taken, and it was discovered that the lead had twisted around itself. A lead revision was done on the patient. There was no additional patient complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50600108 ↗InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient underwent an advanced trial, the patient stated they had good symptom relief for 1-5 days first days of the trial. Then on (b)(6), 2026, the patient started feeling stimulation in the thigh area. Their bladder symptoms returned. The patient elected to revise their lead and add the neurostimulator the next day as they had symptom relief at the beginning of the trial. The surgery took place on (b)(6), 2026.
Patient problems: Inappropriate Device Stimulation of Tissue
MAUDE report 50568954 ↗InjuryAug 31, 2026Communication or Transmission Problem; Impedance Problem
It was reported that a patient with a SNM Tined Lead contacted the representative after observing abnormal indicator lights on the remote control. The patient initially reported a solid red indicator light; however, during a subsequent discussion with the representative, both a solid red and green indicator light were reported. Troubleshooting was performed, including attempts to reprogram around electrode 2, which exhibited an open impedance. Reprogramming was unsuccessful. During follow up evaluation, the patient's remote again displayed a red indicator light, and open impedances were identified on three electrodes. Additional information received indicated that therapy was subsequently turned off, and the patient underwent system removal and replacement procedure. No further patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50515500 ↗InjuryAug 25, 2026Fracture; High impedance; Communication or Transmission Problem; Migration
It was reported that the patient with this Neurostimulator (INS) had an error message and high impedance during a reprogram session. No further troubleshooting could be done. The office ordered an x-ray, and the patient had a follow up appointment. The local care team followed up with the patient to check on the remote control. Additionally, there were no further concerns with the remote control. The patient had their x-ray, and it appeared the lead moved. The patient wanted to have a revision as soon as possible as the device was helping with their urinary symptoms. Additionally, the patient had surgery to revise their Tined Lead and INS. The lead was broken, but the physician was able to remove everything. The patient is doing very well.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50423821 ↗InjuryAug 25, 2026Impedance Problem
It was reported that upon connecting the patient's neurostimulator (INS) to the clinician programmer, an error message appeared and open impedances were noted. The patient stated that their device was not functioning, but the patient later noted that they were fine with no pain or discomfort. The patient underwent an x-ray. A revision was performed and a new lead was placed successfully. There were no patient complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50422514 ↗
FDA clearance and classification
The FDA regulates product code QON (implanted electrical device intended for treatment of fecal incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Common questions about the Axonics 1201
Has the Axonics 1201 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10810005340141), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Axonics 1201?
The FDA's MAUDE database lists 492 adverse event reports for this device, received between Feb 13, 2020 and Sep 29, 2026. In those reports, the most common specific device problems are “Migration” (198), “High impedance” (133) and “Fracture” (44); the most common patient problems are “Failure of Implant” (112), “Incontinence” (100) and “Pain” (53). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Axonics 1201?
None of the reports are classified as deaths, 352 as injuries and 140 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Axonics 1201 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Axonics 1201, and is it still sold?
The FDA's device database lists Axonics Modulation Technologies, Inc. as the company responsible for it (brand Axonics, model 1201). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Implanted Electrical Device Intended For Treatment Of Fecal Incontinence devices
- InterStim™ X 97800Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence5,944 reports
- INTERSTIM® II 3058Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence990 reports
- Axonics 1701Axonics Modulation Technologies, Inc. · Implanted Electrical Device Intended For Treatment Of Fecal Incontinence1 recall31 reports
- InterStim™ SureScan™ 978B133Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence1 recall0 reports
- InterStim™ SureScan™ 978B133Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence1 recall0 reports
- NA TH90QFAMedtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence1 recall0 reports
- InterStim™ SureScan™ 978A141Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence1 recall0 reports
- InterStim™ SureScan™ 978A133Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.