Axium™ Detachable Coil System PC-3-8-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL PC-3-8-HELIX V12 · Class 2 · Implant

All models of Axium Detachable Coil System (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000488925), received between Nov 29, 2023 and Nov 29, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K203432); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 29, 2023Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    MEDTRONIC RECEIVED A REPORT THAT AXIUM COIL PREPPED AND HYDRATED VERTICALLY PER FDA INSTRUCTION. DUE TO NON-TAPERED COIL SHEATH OF P C-3-8-HELIX A SMALL GAP EXISTS BETWEEN AXIUM COIL SHEATH AND ECHELON 10 MICROCATHETER ALLOWING THE COIL TO PREMATURELY BREAK CAUSING RESISTANCE ENOUGH FOR THE PHYSICIAN NOT BE ABLE TO ADVANCE COIL( KNOWN ISSUE). SECOND PC-3-8-HELIX PREPPED AND HYDRATED VERTICALLY PER FDA INSTRUCTION AND INSERTED INTO ECHELON 10 MICROCATHETER BUT SAME OCCURRENCE AS FIRST AXIUM DUE TO GAP CAUSED BY NON-TAPERED SHEATH CAUSING RESISTANCE AND UNABLE TO ADVANCE. INSERTED AN AXIUM PRIME 3-8-HX-ES WITH A TAPERED COILING SHEATH AND THE COIL DEPLOYED WITHOUT ISSUE. IT WAS REPORTED THERE WAS COIL RESISTANCE/STUCK IN SHEATH. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SPINAL EMBOLIZATION. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MINIMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18233609 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K203432. One submission often covers many models, so check the product and lot details in each record.

  • Z-1261-2022 · Apr 27, 2022 · Open, Classified
    AXIUM DETACHABLE COIL SYSTEM REF QC-4-12-HELIX; Axium Detachable Coil System REF QC-10-30-3D;