Skip to content

Avigo 103-0606-200Recalls & reported problems

Micro Therapeutics, Inc. · GUIDEWIRE 103-0606-200 V05 AVIGO MDR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000518905), received between May 9, 2025 and May 9, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2025Break; Material Separation; Break; Material Separation

    MEDTRONIC RECEIVED A REPORT THAT AN AVIGO GUIDEWIRE WAS BROKEN. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF BASILAR ARTERY THROMBOSIS. IT WAS REPORTED THAT THE SHEATH WAS INSERTED FROM THE FEMORAL ARTERY. THE 8FGUIDING WAS INSERTED INTO THE TON-BRIDGE MEDICAL 6F115 I NTERMEDIATE CATHETER TO REACH THE BASILAR ARTERY. THE AVIGO GUIDEWIRE PASSED THROUGH THE THROMBUS AND CARRIED THE MICROCATHETER REBAR18 COAXIALLY THROUGH THE THROMBUS. WHEN THE AVIGO GUIDEWIRE WAS WITHDRAWN, THE DEVELOPING SEGMENT AT THE TIP WAS NOT WITHDRAWN. THE OPERATOR PUSHED FORWARD AND BACKWARD AND FOUND THAT IT WAS BROKEN. THEY THEN PUSHED THE TON-BRIDGE MEDICAL 6F115 INTERMEDIATE CATHETER TO THE DISTAL END TO COVER THE MICROCATHETER. USED AN INTRACARDIAC 2X15 BALLOON AND 15 PRESSURES TO INFLATE AND SQUEEZE THE RESIDUAL GUIDEWIRE IN THE INTERMEDIATE CATHETER, THEN WITHDREW IT, AND SUCCESSFULLY REMOVED THE INTERMEDIATE CATHETER GUIDEWIRE AND THE RESIDUAL MICROCATHETER OUT OF THE BODY. AFTER REMOVAL FROM THE BODY, THE BROKEN PART WAS ABOUT 10 CM. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21993013 ↗

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page