Skip to content

VISUALASE 9735542 (UDI 00763000521370)Recalls & reported problems

Medtronic Navigation, Inc. · SYSTEM 9735542 15W VISUALASE · Class 2

All models of Visualase (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000521370), received between Feb 13, 2025 and Feb 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K211269); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026Infusion or Flow Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A THERMAL THERAPY SYSTEM BEING USED DURING A SOFT TISSUE ABLATION (NEURO) PROCEDURE. IT WAS REPORTED THAT THE SYSTEM'S COOLING FLUID PUMP MALFUNCTIONED IN A CASE. THE PROCEDURE WAS APPROXIMATELY 80% COMPLETE, AT WHICH TIME THE PUMP STOPPED MOVING AND BEGAN MAKING A "SQUEALING" NOISE AS IF A GEAR WAS JAMMED INSIDE. THE MANUFACTURER REPRESENTATIVE (REP) TRIED REVERSING DIRECTIONS AS WELL AS DIFFERENT POWER OUTLETS, BUT THE ISSUE PERSISTED. THE REP WAS ABLE TO MANUALLY MOVE A BEARING IN THE PUMP AND GET IT WORKING AGAIN FOR APPROXIMATELY 30 SECONDS, BUT THEN IT STOPPED MOVING AGAIN. THE REP WAS UNABLE TO GET IT WORKING AT THAT POINT. THE SURGEON THEN OPTED TO ABANDON FURTHER USE OF THE SYSTEM. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316272 ↗
  • MalfunctionFeb 13, 2025Infusion or Flow Problem; Infusion or Flow Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A THERMAL THERAPY SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHEN INSTALLING THE SYSTEM, THE WATER PUMP DID NOT WORK. IT HAD POWER, BUT WHEN THE LEVER WAS PULLED DOWN, NOTHING HAPPENED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21377223 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K211269. One submission often covers many models, so check the product and lot details in each record.

  • Z-0244-2023 · Oct 14, 2022 · Open, Classified
    Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulate soft tissue by physicians in minimally invasive surgical

Other Neurosurgical laser with MR thermography devices

All Neurosurgical laser with MR thermography devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page