Skip to content

ZINGER® SUPPORT ZNGRS180HJ (UDI 00763000569525)Recalls & reported problems

Medtronic, Inc. · GDWR ZNGRS180HJ ZINGERS 5PK 180 J HY · Class 2

All models of ZINGER SUPPORT (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000569525), received between Aug 19, 2024 and Oct 24, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Deformation Due to Compressive Stress1
  • Detachment of Device or Device Component1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE TWO ZINGER GUIDEWIRES WITH A LEFT VENTRICULAR (LV) LEAD TO NAVIGATE THE CORONARY SINUS. IT WAS REPORTED THAT THE FIRST GUIDEWIRE SHREDDED DURING MANEUVERS OF THE DEVICE. THE ENTIRE GUIDEWIRE WAS SUCCESSFULLY REMOVED. THE SECOND GUIDEWIRE GOT STUCK IN THE LV LEAD AND BOTH PRODUCTS HAD TO BE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS PLACED AS THERE WAS NO CAPTURE IN THE LEFT VENTRICLE, THE LEAD WAS NOT USED. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20520566 ↗
  • MalfunctionAug 19, 2024Entrapment of Device; Deformation Due to Compressive Stress; Material Deformation; Entrapment of Device; Deformation Due to Compressive Stress; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE LV LEAD AND ONE ZINGER GUIDEWIRE TO PLACE A LEAD IN A NON-CALCIFIED NARROW VEIN, WITH MODERATE TORTUOSITY IN THE CORONARY SINUS. THE ZINGER GUIDEWIRE WAS INSPECTED WITH NO ISSUES. THE ZINGER GUIDEWIRE WAS PREPPED PER IFU WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. IT WAS REPORTED THAT THE LEAD COULD NOT BE ADVANCED IN THE NARROW VEIN. AFTER MULTIPLE ATTEMPTS, THE DECISION WAS MADE TO ABORT AND SWITCH TO LEFT BUNDLE BRANCH AREA PACING (LBBAP). WHEN THE LEAD AND WIRE WERE REMOVED FROM THE PATIENT, THE WIRE HAD FRAYED/COME APART AND COULD NOT BE REMOVED FROM THE LEAD. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19999970 ↗

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →