Skip to content

Intuition INTU180HSRecalls & reported problems

Medtronic, Inc. · GDWR INTU180HS INTUITION 5PK 180 HY ST · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000569952), received between May 1, 2024 and May 1, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AN ATTEMPT WAS MADE TO USE ONE RESOLUTE ONYX CORONARY DRUG ELUTING STENT, ONE INTUITION GUIDEWIRE, ONE NC EUPHORA BALLOON AND ONE LAUNCHER GUIDE CATHETER DURING A PROCEDURE. ALL DEVICES WERE INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED ON THE RESOLUTE ONYX AND NC EUPHORA DEVICES WITH NO ISSUES. THE INTUITION AND LAUNCHER DEVICES WERE PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE LAUNCHER DEVICE, AND EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT THE RESOLUTE ONYX STENT DISLODGED DURING DELIVERY TO THE LESION. THE STENT DISLODGED DURING PULL AND PUSH FACTOR OF THE INTUITION GUIDEWIRE, AND THE GUIDEWIRE BROKE IN HALF, POSSIBLY DUE TO CALCIFICATION. THE PATIENT WAS RECOMMENDED FOR SURGERY FOR REMOVAL OF THE GUIDEWIRE AND STENT, WHICH WERE REMOVED BY USING THE SNARE TECHNIQUE. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Insufficient Information

    MAUDE report 19220398 ↗

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page