This Medtronic SynchroMed III implanted pump is controlled through a handheld controller and a phone app. Many of the 1,374 reports describe communication problems, such as the app getting stuck when connecting to the pump, and some describe the pump shifting out of place. The reports include 2 deaths, 619 injuries and 753 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,311
- Deaths reported
- 2
- Injuries reported
- 603
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,311 adverse event reports that list this exact device (DI 00763000597030), received between Jan 5, 2024 and Sep 29, 2026 — classified as 2 deaths, 603 injuries, 706 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (560 reports).
40 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Communication or Transmission Problem560
- Adverse Event Without Identified Device or Use Problem333
- Migration or Expulsion of Device139
- Application Program Problem109
- Infusion or Flow Problem104
- Insufficient Device Problem Information95
- Obstruction of Flow79
- Insufficient Flow or Under Infusion77
Patient problems
- No Clinical Signs, Symptoms or Conditions800
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available141
- Unspecified Infection134
- Pain128
- Insufficient Information85
- Muscular Rigidity41
- Seroma41
- Cerebrospinal Fluid Leakage38
Recent adverse event reports
The 25 most recent of 1,311 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Migration or Expulsion of Device; Infusion or Flow Problem; Material Integrity Problem
Information was received from a healthcare provider (HCP) via a manufacturer representative (rep) regarding a patient who was receiving Morphine Drug (1mg/mL at Cannot be obtained; not documented) and Bupivacaine (2mg/mL at Cannot be obtained; not documented) via an implantable pump for non-malignant pain. It was reported that their pump had never worked to relieve pain since the original implant. The patient stated that the pump flipped in the pocket at an unknown date. The physician stated that they did a dye study and unsuccessfully aspirated the catheter. During their revision, it was discovered that the catheter had completed broken at the pump pocket site due to the rotation of the pump. The physician made the medical decision to replace the pump along with the catheter. A new pump was successfully implanted with a new catheter. The tablet was updated by the physician on (b)(6)2026. The patient's pump system was successfully replaced, and the issue was resolved.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 50938042 ↗InjurySep 28, 2026Disconnection; Migration or Expulsion of Device; Communication or Transmission Problem
Information was received from a patient who was receiving Dilaudid (hydromorphone), Fentanyl, and Morphine Drug via an implantable pump for non-malignant pain. It was reported that they were due for a catheter revision surgery because their catheter came undone. The patient. stated the revision was put on hold because the surgeon couldn't get any more suture bags. The agent documented information and transferred the caller to customer services for further assistance. The patient called back and repeated previous documented information. They stated their pump was working great, but the reason why they were calling was about the 'Dacron mesh bag.' The agent reviewed information and transferred the caller to customer service for further assistance. The patient called back and stated they were looking for the sack/pouch for the pump because they were supposed to have revision surgery and needed the pouch. The patient stated the pump was floating around and made it difficult to give a bolus. The patient stated the pump flip sides ways and took about 20 minutes to get a bolus. The agent sent an email to the field for visibility and the manufacturing representative (rep) reported that the
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924058 ↗InjurySep 25, 2026Insufficient Flow or Under Infusion; Adverse Event Without Identified Device or Use Problem
Information was received from the healthcare provider (HCP) via the company representative (rep) regarding a patient receiving intrathecal Morphine 15 mg ml at 8.3 mg day via an implantable pump. It was reported that the HCP said the patient might have had increased pain but was uncertain. No environmental/external/patient factors might have led or contributed to the issue. It was also mentioned that there were discrepancies from the last four refills. Pump was replaced because of under infusion by 5 cc. Patient was doing well with new pump. Old pump was being sent back for analysis. The issue resolved at the time of this report.
Patient problems: Pain
MAUDE report 50902609 ↗InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a healthcare provider via a clinical study regarding a patient receiving an unknown drug via implantable pump. It was reported that on 2026(b)(6) the patient reported severe pump site pain. No actions were taken and the event was ongoing. Additional information received from the healthcare provider via a clinical study indicated that a cluneal nerve block was performed. The patient would like to decrease towards explant as pump was uncomfortable. Additional information received from the healthcare provider via a clinical study indicated that the patient was receiving bupivacaine, hydromorphone, and fentanyl. Additional information received from the healthcare provider via a clinical study indicated that it was reported that the patient was considering pump explant due to pump pocket pain. Cluneal nerve blocks were performed but only provided short-term relief. Additional informati on received from the healthcare provider via a clinical study indicated that the patient was currently decreasing pump towards explant due to ongoing pump pocket pain.
Patient problems: Pain
MAUDE report 50899766 ↗InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a healthcare provider (HCP) regarding a patient receiving an unknown drug via an implantable infusion pump for unknown indications for use. It was reported that the HCP wanted to stop the pump due to an 'expected overdose', but they did not have a clinician programmer at hand. It was discussed that the pump could be programmed in temporarily or permanent stopped mode using the clinician app. If they wanted to stop the pump right away, they could place a cardiac magnet on top of the pump, and this would make sure to keep the pump stopped as long as the magnet was placed on top of the pump. The HCP was transferred to the National Answering Service (NAS).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50839465 ↗MalfunctionSep 18, 2026Communication or Transmission Problem
Information was received from the patient receiving intrathecal medication via an implantable pump for non-malignant pain. It was re ported that the pump went well for 'a while' and as it had gotten 'older', the handset takes forever for the bolus to be administered, about 3-6 minutes. Agent reviewed information and assisted patient to clear data on the handset. Patient was able to connect to the pump without issue and patient said it was fast. When asked boluses per day; patient stated they have a max of 3, 4 in their system, and they usually use 2-3. Patient would call back if assistance was needed clearing that data again. Additional information was received from the patient reporting software version of the personal therapy manager (PTM).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50818634 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
On (b)(6)2026, information was received from an Healthcare Provider (HCP), via a Manufacturer Representative (REP), regarding a patient receiving Lioresal (concentration; 500, Dose; 50 mcg) via an implantable pump. It was reported that the patient's pump and pump segment catheter was replaced. The pump was \"apparently\" infected sometime in (b)(6)2026 so the pump and pump segment were explanted. The spinal segment was tied off and remained intact. The pump segment was replaced and connected to a new pump. There were no known reported environmental, external, or patient factors that may have led or contributed to the issue. Diagnostic / troubleshooting actions were \"explant and antibiotics\". The issue was resolved at the time of this report.
Patient problems: Unspecified Infection
MAUDE report 50810698 ↗InjurySep 16, 2026Migration or Expulsion of Device
Information was received from a manufacturer representative (rep) regarding a patient who was receiving Baclofen via an implantable pump for intractable spasticity. It was reported that the pump was in the stomach area, and it was breaking through the skin. The caller stated the doctor determined it was not an infection that caused it. The caller mentioned that the family had asked for a new pump, and they didn't want to wean the patient off the pump and have it explanted. They were replacing the 40cc pump with a 20 cc pump. However, the caller did not know if it was possible to move the pump to the other side of the stomach because of the way the catheter wraps around the patient's back. The agent redirected the caller to the healthcare provider (HCP) about placement location of new pump.
Patient problems: Pocket Erosion
MAUDE report 50794798 ↗MalfunctionSep 11, 2026Communication or Transmission Problem; Application Program Freezes or Fails to Launch
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50736225 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
Information was received from multiple sources (Manufacturer Representative, Friend/Family Member, Healthcare Provider) regarding a patient who was receiving Gablofen (500 mcg/ml at 50mcg/day) via an implantable pump for intractable spasticity. It was reported that the patient became infected. The exact signs and symptoms were unable to be determined. There were no diagnostics or troubleshooting performed. The family stated that the patient was treated with antibiotics, but they could not confirm the antibiotics. The whole system was explanted. The issue was resolved at the time of the report. The healthcare provider (HCP) had no additional information for the manufacturer.
Patient problems: Unspecified Infection
MAUDE report 50724659 ↗InjurySep 9, 2026Insufficient Flow or Under Infusion; Infusion or Flow Problem
Information was received from multiple sources (Manufacturer Representative, Healthcare Provider) regarding a patient who was receiv ing Unknown Baclofen (1000mcs/mg at 230.3 mcg/day) via an implantable pump for intractable spasticity. It was reported that the patient was experiencing withdrawal symptoms and was an inpatient in the hospital after their pump replacement. A dye study was performed and found it was tough to aspirate and the delivery was constricted. The healthcare provider (HCP) observed a fibrin sheath formed around the catheter tip. A catheter revision was performed and the issue was resolved.
Patient problems: Insufficient Information
MAUDE report 50696601 ↗InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem
Information was received from multiple sources (Manufacturer Representative, Healthcare Provider) regarding a patient who was receiv ing Compounded Baclofen (83.3 mcg/ml at 20.8 mcg/day) and Dilaudid (hydromorphone) (1.5mg/ml at .3746 mg) via an implantable pump for non-malignant pain. It was reported that pocket was opening due to infection. The pump was explanted and pocket was flushed and closed.
Patient problems: Unspecified Infection; Wound Dehiscence
MAUDE report 50594396 ↗InjurySep 1, 2026Obstruction of Flow; Infusion or Flow Problem
Information was received from a healthcare provider (HCP) and a patient's family member via a manufacturer's representative (Rep) re garding a patient receiving Baclofen (1243mcg/mL at 250mcg/day) via an implantable infusion pump for unknown indications for use. It was reported that the patient had baclofen withdrawal symptoms due to a potential compromised catheter. Patient factors noted were that a family member mentioned the patient fallen a short distance off a chair (b)(6) 2026, but also noted that a physical therapist reported an increase in stiffness and muscle tone changes (b)(6) 2026. (b)(6) 2026 symptoms had worsened, and medical care was sought. According to a family member, the HCP ruled out a urinary tract infection and then conducted a catheter access port (CAP) study, or a dye study, on the pump which was unsuccessful and a revision surgery was then scheduled. The dye study was performed and indicated an occlusion or kink in the catheter. The revision surgery was completed (b)(6) 2026. After opening the pocket and removing the pump and catheter, no visual kinks were identified. An attempt was made to aspirate the catheter at that time, but was unsuccessful. The HCP t
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information; Muscular Rigidity
MAUDE report 50549222 ↗InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a health care provider (hcp) regarding a patient with an implantable pump. It was reported that the patient was admitted to the emergency room (ER) with suspected overdose. The hcp wanted to stop the pump. Technical services explained that it was not possible to stop the pump without a clinician programmer. The hcp was instructed to put a magnet on the pump to induce a motor stall, and explained that a two tone alarm would go off once the motor was stalled. The call was transferred to the national answering service (NAS) to page a rep to interrogate the pump.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50541842 ↗InjuryAug 31, 2026Obstruction of Flow
Information was received from multiple sources (Manufacturer Representative, Healthcare Provider) regarding a patient who was receiv ing Clonidine (250 mcg/ml at 3.229) via an implantable pump for malignant pain. It was reported that the medication wasn¿t flowing correctly since the catheter was twisted. The doctor did not like how the medication was flowing in the catheter when he tried to aspirate it. The doctor did not state if the patient had any symptoms. Instead of doing a revision, the patient and the doctor both decided on just giving him a new catheter. The replacement resolved the issue.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50523687 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a healthcare provider (HCP) regarding a patient receiving hydromorphone and bupivacaine via an implantable infusion pump for spinal pain indications and degenerative disc disease/herniated disc pain. It was reported that the patient had their pump explanted on (b)(6)2026 because they thought it was infected. The HCP stated that on (b)(6)2026 the patient called to report that they noticed a small pinhole in the abdominal incision and sometimes fluid came out of it.
Patient problems: Fluid Discharge; Unspecified Infection; Wound Dehiscence
MAUDE report 50463184 ↗InjuryAug 24, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a patient via a manufacturing representative who was receiving morphine ( 2 mg/mL at 0.8446 mg/day) v ia an implantable pump for unknown indication. It was reported that that the patient reported that pump pocket had become more superficial and uncomfortable due to unintentional weight loss of approximately 80 pounds secondary to a recently diagnosed gastrointes tinal (G.I). disorder that resulted in a colostomy bag and tube feedings. The patient was taken to the Operating Room (OR) today (b)(6) 2026 for pocket revision. The physician open up the pump pocket and dissected down to the existing pump and proximal segment. Once he disconnected the pump from the catheter, he noted freely dripping Cerebrospinal fluid (CSF) from the catheter. He then moved the pocket move inferior and medial. The pump was then reconnected to the catheter and placed back into the pump pocket. The physician opted not to aspirate the catheter because he noted the CSF freely dripping prior to reconnecting it to the pump. Once back in the pocket, incisions were then irrigated and closed. The issue was resolved at the time of the report. The healthcare provider had no further infor
Patient problems: Discomfort; Insufficient Information
MAUDE report 50418856 ↗InjuryAug 23, 2026Insufficient Flow or Under Infusion; Insufficient Device Problem Information
Information was received from the healthcare provider (HCP) via the company representative (rep) regarding a patient receiving intrathecal medication via an implantable pump. It was reported that the patient's pump was replaced.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50397275 ↗InjuryAug 19, 2026Infusion or Flow Problem
Information was received from multiple sources (Patient, Manufacturer Representative, Healthcare Provider) regarding a patient who was receiving Bupivacaine (5000 mcg/ml at 2000 mcg/day) and Fentanyl (200 mcg/ml at 80 mcg/day) via an implantable pump. It was reported that the patient had a return of symptoms. A dye study attempted was today, but the healthcare provider was unable to aspirate the catheter. No external factors were involved. On (b)(6)2026 a revision occurred and part of the old catheter was explanted with part left implanted but not in use. A new catheter was placed.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information
MAUDE report 50328998 ↗InjuryAug 18, 2026Fluid/Blood Leak; Obstruction of Flow; Material Integrity Problem
Information was received from multiple sources (Manufacturer Representative (rep), Healthcare Provider (HCP) ) regarding a patient w ho was receiving Baclofen (500mcg/ml at 100 mcg/day) via an implantable pump for other and intractable spasticity. It was reported that the pump section of the catheter was replaced and part of the spinal segment was removed due to kinking. The patient was experiencing loss of medication effectiveness. The patient had fluid accumulation at the spinal pocket. Cerebrospinal fluid (CSF) was found to be leaking around the catheter. There were no known contributing factors. The HCP found leaking around the catheter. They performed a purse string suture to close up the leak. The kinking at the catheter was removed by cutting some of the spinal segment out and then replacing the pump segment with a new catheter. The issue was resolved at the time of the report. Additional information received from the rep on 2026-Aug-05 indicated that the catheter had to be revised and cut numerous times before it was completely replaced.
Patient problems: Cerebrospinal Fluid Leakage
MAUDE report 50306457 ↗InjuryAug 14, 2026Insufficient Flow or Under Infusion; Obstruction of Flow
Information was received from a healthcare provider via company representative regarding a patient with an implantable pump containing Gablofen (500 mcg/ml at 175 mcg/day). It was reported that the patient had experienced a return of symptoms with high residual volumes at refills. It was unknown if there were any environmental/external/patient factors that may have led or contributed. The diagnostics/troubleshooting steps that were completed was a dye study and an attempt to aspirate in the operating room. Per the healthcare provider, the results of the dye study done on (b)(6) 2026 came back as inconclusive. The physician was also unable to aspirate in the operating room. Surgical interventions was performed by the physician where it was observed that the existing pump segment was kinked. The physician attached a new pump segment to the existing tip segment and then was able to successfully aspirate the entire catheter resolving the issue.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50261035 ↗InjuryAug 10, 2026Obstruction of Flow; Infusion or Flow Problem; Insufficient Device Problem Information
Information was received from a company representative regarding a patient with an implantable pump containing Fentanyl for spinal pain indications and Other. It was reported that the patient was having a catheter revision. The representative reported the patient lost efficacy with treatment a couple of months ago so the physician was replacing part of the catheter today.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50195265 ↗InjuryAug 7, 2026Infusion or Flow Problem
Information was received from a healthcare provider (hcp) and a manufacturing representative (rep) regarding a patient who was receiving Fentanyl (500mcg/mL at 250mcg/day ) via an implantable pump for non-malignant pain. It was reported that the reason for call was caller (rep) stated that a doctor called him today ((b)(6)2026) about a patient where they had trouble aspirating the catheter. Caller stated that the doctor was going to do a dye study but couldn't aspirate. Caller had no additional details on the issue but stated that the patient being referred to physician to investigate the catheter. Additional information was received from a healthcare provider via a manufacturing representative. It was reported that the patient experienced withdrawal symptoms, and had a dye study done on (b)(6)2026. The physician was unable to aspirate catheter and referred her to have a catheter revision. She was admitted to the hospital since that time due to increased withdrawal symptoms. She did not remember what specific day. The catheter was successfully replaced by the physician on (b)(6)2026. There were no environmental/external/patient factors that might had led or contributed to the issue
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50177338 ↗InjuryAug 4, 2026Excess Flow or Over-Infusion; Migration or Expulsion of Device; Insufficient Device Problem Information
Information was received from the healthcare provider (HCP) regarding a patient receiving intrathecal medication via an implantable pump. It was reported that the patient's pump was explanted and replaced.
Patient problems: Anxiety; No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 50134962 ↗InjuryAug 4, 2026Migration or Expulsion of Device; Material Integrity Problem
Information was received from a patient (con) and a healthcare provider (hcp) and a manufacturing representative (rep) regarding a patient who was receiving an unknown drug via an implantable pump for unknown indication. It was reported that the pump was flipped around in the pump pocket and caused the catheter to flip around as well. The cause of the event was not determined. A segment of the catheter was replaced by the provider. The issue was resolved at the time of the report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50128522 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P860004 on Mar 14, 1988. Submitted by Medtronic, Inc. as “Medtronic(R) Synchromed(Tm) Pump & Infusion System”. The FDA review took 770 days. The FDA has since approved 445 supplements (changes to the device or its labeling), most recently on Oct 30, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Recalls under the same FDA submission
These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.
- Z-2201-2026 · Apr 8, 2026 · Open, ClassifiedMedtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
- Z-1506-2025 · Jan 9, 2025 · Open, ClassifiedA820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
- Z-2352-2024 · May 10, 2024 · Open, ClassifiedMedtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
- Z-0615-2024 · Nov 13, 2023 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-40, Programmable pump
- Z-0614-2024 · Nov 13, 2023 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-20, Programmable pump
- Z-0522-2024 · Jul 13, 2020 · CompletedMedtronic SynchroMed, Model A10
- Z-0508-2020 · Oct 11, 2019 · TerminatedSynchromed II programmable pump, Model numbers 8637-20 and 8637-40
- Z-0537-2018 · Sep 21, 2017 · TerminatedMedtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
- Z-0896-2018 · Aug 9, 2017 · Open, ClassifiedMedtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
- Z-0788-2017 · Oct 3, 2016 · TerminatedMedtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
- Z-1271-2016 · Feb 9, 2016 · TerminatedMedtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
- Z-1681-2015 · Apr 10, 2015 · TerminatedMedtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
- Z-1062-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
- Z-1061-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
- Z-1060-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
…and 25 more.
Common questions about the SYNCHROMED™ III 8667-40
Has the SYNCHROMED™ III 8667-40 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000597030), based on the FDA recall file dated Nov 27, 2026. However, 40 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the SYNCHROMED™ III 8667-40?
The FDA's MAUDE database lists 1,311 adverse event reports for this device, received between Jan 5, 2024 and Sep 29, 2026. In those reports, the most common specific device problems are “Communication or Transmission Problem” (560), “Migration or Expulsion of Device” (139) and “Application Program Problem” (109); the most common patient problems are “Unspecified Infection” (134), “Pain” (128) and “Muscular Rigidity” (41). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SYNCHROMED™ III 8667-40?
2 reports are classified as deaths, 603 as injuries and 706 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SYNCHROMED™ III 8667-40 FDA approved?
It was approved through premarket approval (PMA) P860004 on Mar 14, 1988.
Is the SYNCHROMED™ III 8667-40 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the SYNCHROMED™ III 8667-40, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand SYNCHROMED™ III, model 8667-40). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, infusion, implanted, programmable devices
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,876 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,270 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,118 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,052 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,083 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,587 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,289 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,189 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page