Medtronic, Inc. · SHEATH 10FCC20 FLEXCATH CONTOUR 10F US · Class 2

Adverse event reports
15
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00763000659097), received between Mar 7, 2024 and Nov 7, 2025 — classified as 2 injuries, 13 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent10
  • Gas/Air Leak5
  • Adverse Event Without Identified Device or Use Problem2
  • Fluid/Blood Leak2
  • Leak/Splash2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Hematoma1
  • Non specific EKG/ECG Changes1
  • Pericardial Effusion1
  • Pericarditis1
  • Tachycardia1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS LATER REPORTED THAT THE PATIENT UNDERWENT EXTENDED HOSPITALIZATION AFTER THE PROCEDURE.

    Patient problems: Hematoma; Non specific EKG/ECG Changes; Pericarditis

    MAUDE report 23498445 ↗
  • MalfunctionMar 31, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, WHEN ATTEMPTING TO DEFLECT INTO THE LEFT INFERIOR PULMONARY VEIN (LIPV), AN AUDIBLE SOUND WAS HEARD THAT WAS SURMISED TO BE A PULL WIRE SNAPPING IN THE SHEATH. AFTER THAT, THE SHEATH WAS UNABLE TO DEFLECT. THE SHEATH WAS REPLACED WHICH RESOLVED THE ISSUE. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21731842 ↗
  • MalfunctionJan 2, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, THERE WAS AN ISSUE WITH THE SHEATH SIDE PORT TUBING, WHICH KINKED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT SYMPTOMS OR COMPLICATIONS RELATED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21057630 ↗
  • MalfunctionDec 6, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, A KINK WAS OBSERVED AT THE SIDE PORT OF THE SHEATH. THE HEALTHCARE PROFESSIONAL EXPRESSED CONCERN THAT THE KINK COULD POTENTIALLY LEAD TO COMPLICATIONS; HOWEVER, THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT SYMPTOMS OR COMPLICATIONS. NO MEDICAL OR SURGICAL INTERVENTION WAS NEEDED TO PREVENT PERMANENT IMPAIRMENT, AND THE EVENT DID NOT LEAD TO OR EXTEND PATIENT HOSPITALIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20860510 ↗
  • MalfunctionDec 4, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, THE SIDE PORT OF THE SHEATH TUBING KINKED. THE PROCEDURE WAS COMPLETED WITH PFA.  NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20845201 ↗
  • MalfunctionOct 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT TUBING ON THE SHEATH HANDLE KINKED. IT WAS SURMISED VIA CUSTOMER FEEDBACK THAT THIS MALFUNCTION COULD LEAD TO CLOTTING IN THE DEVICE HANDLE. DESPITE THE ISSUE, THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20561102 ↗
  • InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT'S HEART RATE WAS HIGHER THAN EXPECTED. IMAGING WAS CARRIED OUT AND A PERICARDIAL EFFUSION WAS OBSERVED. A PERICARDIAL TAP WAS THEN CARRIED OUT ON THE PATIENT. THE CASE WAS COMPLETED WITH CRYO. THE PATIENTS HOSPITALIZATION WAS EXTENDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pericardial Effusion; Tachycardia

    MAUDE report 19943319 ↗
  • MalfunctionMay 20, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN RADIOFREQUENCY (RF) PROCEDURE, THE SIDEARM WAS KINKED. IT WAS FURTHER REPORTED THAT AIR BUBBLES WERE OBSERVED BY THE PHYSICIAN. THE CASE WAS COMPLETED WITH RADIOFREQUENCY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351041 ↗
  • MalfunctionMay 13, 2024Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION (PFA) PROCEDURE, AFTER THE CATHETER WAS REPLACED, THE FLUSH LINE OF THE SHEATH COULD NOT BE ASPIRATED. IT WAS NOTED THAT THE SIDEARM WAS KINKED. THE SIDEARM WAS UNKINKED TO RESOLVE THE ISSUE. IT WAS ALSO REPORTED THAT MICROBUBBLES WERE SEEN DURING THE PROCEDURE. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION (PFA). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19305514 ↗
  • MalfunctionApr 5, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE INTEGRATED DILATOR/NEEDLE WOULD NOT LOCK FIRMLY INTO THE SHEATH DESPITE MULTIPLE ATTEMPTS. IT WAS ALSO REPORTED THAT LARGE BUBBLES WERE OBSERVED IN THE SIDE PORT OF THE SHEATH. THE CASE WAS COMPLETED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19054409 ↗
  • MalfunctionApr 2, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION, AIR BUBBLES WERE NOTED IN SIDE ARM DURING CATHETER REMOVAL AND INSERTION AND BUBBLES SEEN DURING ASPIRATION OF SIDE ARM WITH CATHETER IN THE SHEATH. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19017683 ↗
  • MalfunctionMar 29, 2024Fluid/Blood Leak; Material Twisted/Bent; Fluid/Blood Leak; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE VALVE ON THE SHEATH WAS LEAKY. THE VALVE CONTINUED TO LEAK WHEN THE CATHETER WAS REMOVED FROM THE SHEATH. DESPITE THE ISSUE, THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19005452 ↗
  • MalfunctionMar 25, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT OF THE SHEATH HAD A KINK AND MICROBUBBLES WERE OBSERVED IN THE SIDE PORT. DESPITE THE ISSUES, CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18967006 ↗
  • MalfunctionMar 19, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT OF THE SHEATH HAD A KINK IN IT. THE SIDEARM OF THE SHEATH WAS AGAINST THE PATIENT, AND WHILE ASPIRATION WAS BEING PERFORMED, MICROBUBBLES WERE OBSERVED IN THE SIDEARM. DESPITE THE ISSUES, CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18935311 ↗
  • MalfunctionMar 7, 2024Leak/Splash; Gas/Air Leak; Leak/Splash; Gas/Air Leak

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, MICROBUBBLES WERE OBSERVED IN THE SIDEARM OF THE SHEATH. THE BUBBLES SEEMED TO BE MORE PRONOUNCED DURING ASPIRATION OF THE SHEATH UPON CATHETER REMOVAL. DESPITE THE ISSUE, THE CASE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18861744 ↗

Other CATHETER, STEERABLE devices

All CATHETER, STEERABLE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.