FlexCath Contour™ 10FCC20Recalls & reported problems
Medtronic, Inc. · SHEATH 10FCC20 FLEXCATH CONTOUR 10F US · Class 2
All models of FlexCath Contour (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00763000659097), received between Mar 7, 2024 and Nov 7, 2025 — classified as 2 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent10
- Gas/Air Leak5
- Adverse Event Without Identified Device or Use Problem2
- Fluid/Blood Leak2
- Leak/Splash2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Hematoma1
- Non specific EKG/ECG Changes1
- Pericardial Effusion1
- Pericarditis1
- Tachycardia1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS LATER REPORTED THAT THE PATIENT UNDERWENT EXTENDED HOSPITALIZATION AFTER THE PROCEDURE.
Patient problems: Hematoma; Non specific EKG/ECG Changes; Pericarditis
MAUDE report 23498445 ↗MalfunctionMar 31, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, WHEN ATTEMPTING TO DEFLECT INTO THE LEFT INFERIOR PULMONARY VEIN (LIPV), AN AUDIBLE SOUND WAS HEARD THAT WAS SURMISED TO BE A PULL WIRE SNAPPING IN THE SHEATH. AFTER THAT, THE SHEATH WAS UNABLE TO DEFLECT. THE SHEATH WAS REPLACED WHICH RESOLVED THE ISSUE. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21731842 ↗MalfunctionJan 2, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, THERE WAS AN ISSUE WITH THE SHEATH SIDE PORT TUBING, WHICH KINKED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT SYMPTOMS OR COMPLICATIONS RELATED TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21057630 ↗MalfunctionDec 6, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, A KINK WAS OBSERVED AT THE SIDE PORT OF THE SHEATH. THE HEALTHCARE PROFESSIONAL EXPRESSED CONCERN THAT THE KINK COULD POTENTIALLY LEAD TO COMPLICATIONS; HOWEVER, THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT SYMPTOMS OR COMPLICATIONS. NO MEDICAL OR SURGICAL INTERVENTION WAS NEEDED TO PREVENT PERMANENT IMPAIRMENT, AND THE EVENT DID NOT LEAD TO OR EXTEND PATIENT HOSPITALIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20860510 ↗MalfunctionDec 4, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE USING THE PULSE FIELD ABLATION SYSTEM, THE SIDE PORT OF THE SHEATH TUBING KINKED. THE PROCEDURE WAS COMPLETED WITH PFA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20845201 ↗MalfunctionOct 29, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT TUBING ON THE SHEATH HANDLE KINKED. IT WAS SURMISED VIA CUSTOMER FEEDBACK THAT THIS MALFUNCTION COULD LEAD TO CLOTTING IN THE DEVICE HANDLE. DESPITE THE ISSUE, THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20561102 ↗InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT'S HEART RATE WAS HIGHER THAN EXPECTED. IMAGING WAS CARRIED OUT AND A PERICARDIAL EFFUSION WAS OBSERVED. A PERICARDIAL TAP WAS THEN CARRIED OUT ON THE PATIENT. THE CASE WAS COMPLETED WITH CRYO. THE PATIENTS HOSPITALIZATION WAS EXTENDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Pericardial Effusion; Tachycardia
MAUDE report 19943319 ↗MalfunctionMay 20, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN RADIOFREQUENCY (RF) PROCEDURE, THE SIDEARM WAS KINKED. IT WAS FURTHER REPORTED THAT AIR BUBBLES WERE OBSERVED BY THE PHYSICIAN. THE CASE WAS COMPLETED WITH RADIOFREQUENCY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19351041 ↗MalfunctionMay 13, 2024Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION (PFA) PROCEDURE, AFTER THE CATHETER WAS REPLACED, THE FLUSH LINE OF THE SHEATH COULD NOT BE ASPIRATED. IT WAS NOTED THAT THE SIDEARM WAS KINKED. THE SIDEARM WAS UNKINKED TO RESOLVE THE ISSUE. IT WAS ALSO REPORTED THAT MICROBUBBLES WERE SEEN DURING THE PROCEDURE. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION (PFA). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19305514 ↗MalfunctionApr 5, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE INTEGRATED DILATOR/NEEDLE WOULD NOT LOCK FIRMLY INTO THE SHEATH DESPITE MULTIPLE ATTEMPTS. IT WAS ALSO REPORTED THAT LARGE BUBBLES WERE OBSERVED IN THE SIDE PORT OF THE SHEATH. THE CASE WAS COMPLETED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19054409 ↗MalfunctionApr 2, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION, AIR BUBBLES WERE NOTED IN SIDE ARM DURING CATHETER REMOVAL AND INSERTION AND BUBBLES SEEN DURING ASPIRATION OF SIDE ARM WITH CATHETER IN THE SHEATH. THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19017683 ↗MalfunctionMar 29, 2024Fluid/Blood Leak; Material Twisted/Bent; Fluid/Blood Leak; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE VALVE ON THE SHEATH WAS LEAKY. THE VALVE CONTINUED TO LEAK WHEN THE CATHETER WAS REMOVED FROM THE SHEATH. DESPITE THE ISSUE, THE CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19005452 ↗MalfunctionMar 25, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT OF THE SHEATH HAD A KINK AND MICROBUBBLES WERE OBSERVED IN THE SIDE PORT. DESPITE THE ISSUES, CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18967006 ↗MalfunctionMar 19, 2024Gas/Air Leak; Material Twisted/Bent; Gas/Air Leak; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PULSED FIELD ABLATION PROCEDURE, THE SIDE PORT OF THE SHEATH HAD A KINK IN IT. THE SIDEARM OF THE SHEATH WAS AGAINST THE PATIENT, AND WHILE ASPIRATION WAS BEING PERFORMED, MICROBUBBLES WERE OBSERVED IN THE SIDEARM. DESPITE THE ISSUES, CASE WAS COMPLETED WITH PULSED FIELD ABLATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18935311 ↗MalfunctionMar 7, 2024Leak/Splash; Gas/Air Leak; Leak/Splash; Gas/Air Leak
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, MICROBUBBLES WERE OBSERVED IN THE SIDEARM OF THE SHEATH. THE BUBBLES SEEMED TO BE MORE PRONOUNCED DURING ASPIRATION OF THE SHEATH UPON CATHETER REMOVAL. DESPITE THE ISSUE, THE CASE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18861744 ↗
FDA clearance and classification
Cleared through 510(k) K232321 on Oct 31, 2023 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “FlexCath Contour Steerable Sheath”. The FDA review took 89 days. FDA record ↗
The FDA regulates product code DRA (catheter, steerable) under 21 CFR 870.1280 as a Class II device. 21 CFR 870.1280 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the FlexCath Contour™ 10FCC20
Has the FlexCath Contour™ 10FCC20 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000659097), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the FlexCath Contour™ 10FCC20?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Mar 7, 2024 and Nov 7, 2025. In those reports, the most common specific device problems are “Material Twisted/Bent” (10), “Gas/Air Leak” (5) and “Fluid/Blood Leak” (2); the most common patient problems are “Hematoma” (1), “Non specific EKG/ECG Changes” (1) and “Pericardial Effusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FlexCath Contour™ 10FCC20?
None of the reports are classified as deaths, 2 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the FlexCath Contour™ 10FCC20 FDA approved?
It was cleared through 510(k) K232321 on Oct 31, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the FlexCath Contour™ 10FCC20, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand FlexCath Contour™, model 10FCC20). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, STEERABLE devices
- FARADRIVE™ M004PF21M402Boston Scientific Corporation · CATHETER, STEERABLE1 recall16 deaths2,303 reports
- MITRACLIP SGC0701Abbott Vascular Inc. · CATHETER, STEERABLE17 deaths962 reports
- STEERABLE GUIDE CATHETER SGC0301Abbott Vascular Inc. · CATHETER, STEERABLE5 deaths390 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE6 deaths167 reports
- FlexCath Contour™ 10FCC13Medtronic, Inc. · CATHETER, STEERABLE2 deaths115 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE4 deaths61 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE4 deaths60 reports
- FlexCath Advance™ 4FC12Medtronic, Inc. · CATHETER, STEERABLE59 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page