Skip to content

Custom Perfusion System recallZ-1027-2023 · Medtronic Perfusion Systems · initiated Dec 12, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

FDA root cause category: Process change control

What the company is doing

Medtronic initially notified consignees on 12/12/2022 and followed up with an updated notification on 12/21/2022. Consignees were advised of two options: 1. Continue to use the Affinity Fusion Oxygenator without using the TMA (use other conventional perfusion circuit temperature monitoring methods); or 2. Continue to use the Affinity Fusion Oxygenator and the TMA for arterial temperature monitoring. Ensure there is minimal torque applied to the oxygenator TMA when attaching or detaching the Temperature Probe (Product Number ATP210). Also, minimize manipulation of the TMA-probe connection during the clinical procedure. Consignees were also advised to complete and return the Customer Confirmation Form, and notify personnel or facilities were affected units are located. On about 02/10/2023, Medtronic initiated an expansion and notified customers. Instructions in the customer letter remained the same.

Recall details

Recall number
Z-1027-2023
Status
Open, Classified
Initiated by the company
Dec 12, 2022
Posted by the FDA
Jan 27, 2023
Recalling firm
Medtronic Perfusion Systems
Quantity
63,278 units
Product code
DTZ
FDA submissions
K203111
Distribution
Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Honduras, Hungary, Iceland, Indonesia, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, Turkmenistan, Uganda, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Yemen.

Product description

Custom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF BB811. Used in Cardiopulmonary bypass.

Code information (models, lots, serial numbers)

Show the full code information (26k characters)

Oxygenator Model BB811; Serial Number Range: 8111483548 - 8113999999; GTIN Number, Tubing Pack Model Number (Lot Numbers): 00643169231948, M390082B (223422891, 223952683, 225651632); 00643169345379, M961233A (223537804); 00643169422025, M445428F (223378929, 223672423, 223909336, 224399508, 224695321, 225303449, 225434680); 00643169439269, M274805B (223714089, 224907239); 00643169449442, BB6U40R11 (223728548, 224155209, 224539886, 224794461); 00643169469495, BB4Q12R15 (223612682, 224026310); 00643169572263, HY8L55R5 (225033983, 225033984, 225045278, 225126456); 00643169583542, M330119D (223547370, 224206448, 225484644); 00643169621473, BB8U95R1 (223488691, 223764562, 225251346); 00643169622715, BB8B70R7 (224281969, 224664696); 00643169632097, M721107D (223538837, 223588897, 224380964, 224567880, 225089620, 225528159); 00643169639027, BB3X42R13 (223558525, 223896827, 224155078, 224664710, 224815782); 00643169645424, BB8M16R3 (224155078); 00643169674486, M722001A (223614712, 224077488, 225161996); 00643169692893, BB8P01R7 (223066047, 223644491, 223983902, 224573049, 225047877); 00643169711945, BB8T77R2 (223283021, 224497848); 00643169717336, BB8Q09R2 (223054336, 223206892, 223409300, 223573934, 223804327, 223896843, 224066169, 224327204, 224664704, 224664705, 224985026, 225170609); 00643169750883, BB8J66R17 (222974930, 222974932, 223206895, 223274762, 224066185, 224282025, 224454020, 224539805, 224752687, 224838236, 224838238); 00643169793699, BB8C33R6 (223234986, 224281982); 00643169793927, BB8M48R5 (223685618, 223833474, 224963605); 00643169840058, M342615P (223630909, 224033583, 224311665, 224567877, 225048104, 225303159, 225484317); 00643169860643, M482903D (223672424, 223969961, 224312691, 224735824, 225207742, 225561384); 00643169865365, BB9D32R2 (223033102, 223127174, 223488375, 223851344, 224233682, 224281957, 224573053, 224794262, 224794263); 00643169872127, BB9L72R1 (223145738, 223631537, 223896364, 224664748, 224923414, 225126448); 00643169876743, BB9B43R3 (223573837, 224794493); 00643169876903, BB8U52R8 (223728547, 224281942, 224473476); 00643169955790, BB3M00R30 (223896474, 224497817, 225047868); 00643169957008, M483002C (223508310, 223815840, 224008812, 224075857, 224992815, 225161643, 225560932); 00643169998643, BB9J85R3 (222974878, 223728616, 224442258, 224963603); 00763000006884, M721902F (223665117, 224069967); 00763000008215, BB8E47R17 (223896458, 224582573); 00763000014001, M059709E (223808314); 00763000014971, HY9E07R9 (223006393, 223764172, 224497852, 225047918); 00763000016944, M447026J (223311071, 224033576, 224525725, 225089619); 00763000021887, M450202J (223378878, 223482251, 223569832, 223695950, 224399265, 224695323, 224948490, 225208459); 00763000024406, BB9N31R3 (223323272, 223603527, 224414960); 00763000025878, BB4B15R14 (223283065, 224066162, 224497806, 225219652); 00763000026240, BB10D95R (224923422); 00763000062613, M233101D (223926251); 00763000062637, M233105F (223926253); 00763000064242, M447708H (223240765, 223909585, 224252242, 224567886, 224736688, 225209646, 225436092, 225651518); 00763000077747, HY0W84R31 (225059180); 00763000099831, BB10B90R6 (223283006, 224277441, 224752773); 00763000112035, M412501B (224115258, 224465727, 224736967, 225303530); 00763000118914, BB8P23R4 (223112645, 224154996, 225104523, 225398699); 00763000126315, BB10H89R4 (223498029, 223896315); 00763000133030, HY9S48R4 (223060795); 00763000135348, BB7J04R11 (223112635, 223603534, 224327223, 224539803, 224811484); 00763000135355, BB8H01R4 (224155218, 224476527, 224923391); 00763000139513, BB8N34R6 (222974686, 223409294, 224019319, 224516511); 00763000139742, M720109G (223378927, 223532940, 224122998, 224652776, 224907237, 225254267); 00763000141363, BB9R53R7 (223235051, 224370597); 00763000155353, BB8C53R12 (222982219, 223558532, 223728602); 00763000159504, M273508D (223994900, 225049328); 00763000162245, M371203C (223540014, 223953001, 224865428, 225346409); 00763000175726, HY10G12R7 (223558545); 00763000178048, BB7Y33R8 (223014654, 223196558, 225047919); 00763000186180, M410209L (223422890, 223969951, 224337397, 224993580); 00763000188276, BB8G43R7 (223342946, 223922491, 224497816, 224923386); 00763000201159, BB8P06R10 (223282994, 223558454, 224281953, 224582631); 00763000201456, BB10S84R (223014671, 223165007, 223206893, 223896868, 224539690); 00763000203405, M391414L (223439639, 224008699, 225065951, 225254280); 00763000211271, HY9R12R13 (224112059); 00763000214531, M331107A (225049339); 00763000224349, BB0E27R32 (223165010); 00763000231538, BB2493R44 (222918292, 223169854, 223558505); 00763000233259, HY5V41R21 (225008158); 00763000234911, BB10P38R2 (222980256, 223728607, 224281988); 00763000236588, HY7E57R17 (224923409); 00763000245351, M495523C (224993583); 00763000246914, M483804B (223508320, 224123002, 225228308); 00763000252663, M721333F (223385805, 223816583, 225331628); 00763000255220, BB8Q56R7 (223044144, 223498060, 223939254, 223967258, 224327237, 224582569, 225086642); 00763000256739, BB8D44R6 (223014665, 223366534, 223896431, 224282016, 224664764, 225270451); 00763000258535, BB10X94R1 (223014672, 223939240, 225170565); 00763000260217, BB8L80R5 (223804388, 224282106, 224752767, 224827938); 00763000262372, BB8X02R9 (223728556, 224282007, 225047878); 00763000263799, BB10S31R3 (223748240, 224196599); 00763000273026, BB8E21R10 223196716); 00763000274610, BB8M09R1 (223645439, 223764064); 00763000275259, BB6M22R15 (224752717); 00763000275341, BB4Z62R19 (223113114, 223728590, 224155206, 224196598, 224370575, 224370576, 224838243, 225008581, 225126482); 00763000276010, BB10V32R (223728486, 223833388, 224758658, 224826799); 00763000280277, M344803C (223648108); 00763000286897, BB10Y67R3 (223234951, 223454165, 224155184, 224539899); 00763000287948, HY9G02R4 (224963556); 00763000288266, HY7W91R22 (225047922); 00763000291839, M337709C (223753451, 224252233, 224948527, 225484643); 00763000293796, BB7H73R17 (223066046, 223334261, 223983917); 00763000299064, BB10S01R2 (225008518); 00763000299859, HY10W55R4 (223060796, 223742623, 223896428); 00763000300692, BB8B96R14 (223095081, 223308375, 223851415, 224327302, 224727367, 224727381); 00763000300937, P490302G (223952674); 00763000316976, HY7V82R20 (225126488, 225126489, 225126490); 00763000319984, HY10S39R3 (223983866); 00763000324773, BB8C09R8 (223409262, 224539793); 00763000325190, M075502A (224419078); 00763000334079, BB10E61R4 (223196718, 223558523, 224281959, 224681554, 225047867); 00763000334710, M612704H (223589853, 224048919, 224399524, 224644262, 225528147); 00763000339319, BB10X57R2 (222974895, 223283046, 223685655, 224281979, 224623803, 225170602); 00763000339319, BB10X57R2 (224794435); 00763000339982, M490645H (223393275, 223713005, 223862637, 224033582, 224337395); 00763000341053, M343105N (223241295, 224188635, 225651514); 00763000343088, M390728D (223712998, 224206461, 225120479, 225303518); 00763000343101, M333715H (223508318); 00763000343125, M391714B (223245021, 223521591, 224096698); 00763000344467, HY2U43R17 (222918489, 223366491, 224019352, 224582540); 00763000345082, BB11D80R (223366501, 223922494, 224582570); 00763000345341, M341506M (223752205, 224493232, 224865426, 225303515); 00763000346478, BB10P64R2 (223054331, 223054338, 223939276, 224454048, 224879178); 00763000346508, BB10N96R9 (223039591, 223409295, 223851363, 224282011, 224664692); 00763000347468, M393929B (224008686); 00763000351632, M322226B (223311076, 223837958, 224652840); 00763000353988, BB8B99R6 (223014674, 223054435, 223283257, 223728610, 224066167, 224281974, 224454133, 224582611, 224784931, 224827935); 00763000361631, BB10V51R6 (223644493, 224026309, 224327297, 224582608, 225126484); 00763000362119, BB11A97R2 (223100965, 223232472, 223323298, 223366554, 223558497, 223558498, 223558499, 224112033, 224281950, 224370559, 224454051, 224454052, 224539876, 224794434, 224963551, 224963552, 225270478); 00763000365837, BB11C80R1 (223234952, 223728861, 224281954, 224827933, 224908927, 225398632); 00763000367671, HY10R64R10 (224794438); 00763000369798, BB10R29R9 (223283067, 223764143, 224155201, 224281965, 224438188, 224807342); 00763000371456, BB1A19R53 (223127717, 223558636, 223741575, 224005631, 224370568, 224370569, 224623765, 225086681); 00763000373955, BB8G55R15 (225170562); 00763000378493, BB11E97R1 (223939212); 00763000378707, BB8D19R9 (223851379, 224752796, 225443660); 00763000400071, BB5N63R5 (224111962, 224497871, 225126503, 225219702); 00763000400576, BB10Y59R9 (223644492, 224497855, 225408501); 00763000401979, M632007A (223552152, 223552153, 224033798); 00763000402501, M483507A (223422023, 223862847, 224224164, 224297726, 224907140, 225090699, 225346302, 225651635); 00763000404062, M075204A (223482693); 00763000404239, M494833G (223466603, 223612884, 223695960, 224865316, 225389397); 00763000408343, HY8T43R9 (222991247, 222991247, 223093372, 223093372, 223728599, 223728599, 223851364, 223896429, 223896429, 224019182, 224019182, 224019183, 224019183, 224019184, 224019184, 224370593, 224370593, 224370594, 224370594, 224497860, 224497860); 00763000408350, BB7V97R15 (223033103, 223196545, 223498044, 223603535, 223983880, 224232482, 224232482, 224281943, 224327276, 224370542, 224370543, 224497755, 224497756, 224497756, 224794357, 224794357, 224794358, 224794358, 225126493); 00763000408824, M964503C (223484468, 224711012); 00763000412128, M334211F (225651640); 00763000412838, BB10L76R4 (223323277, 224155000, 224794449); 00763000412883, BB8W14R8 (223093369, 223206888, 223488706, 223764171, 223931082, 224243204, 224529301, 224529301); 00763000415518, M442233G (223521613, 223882114, 224225700, 224610491); 00763000416355, BB10Q99R11 (223026452, 223206896, 223366549, 223685610, 223685611, 224112042, 224282010, 224454017, 224539762, 224565388, 224879086, 225086646, 225311198); 00763000416980, HY11G55R1 (223558536, 223983966, 224370546, 224370547, 224963529); 00763000417642, M720727L (223672425, 223881861, 224652780, 225161916); 00763000417666, M720413J (225049343); 00763000418649, BB8E17R7 (224019348, 224019348, 224170585, 224327212, 224539786, 224539786, 224879115, 224879115); 00763000418748, M333816C (224206455); 00763000425760, BB10R49R6 (223054433, 223103737, 223764138, 224170582, 224281990, 224539789, 224794453, 225219686); 00763000426002, HY11E26R3 (222724087); 00763000426354, BB5Z97R18 (222998309, 223172628, 223804414, 224281949, 224370558, 224923413); 00763000428488, HY10H88R12 (223343008, 223764185, 224327202, 224794432); 00763000428716, BB3V10R34 (224282013); 00763000429058, M445885K (223352900, 223466329, 223815842, 224138282, 224652778, 225041058, 225436095); 00763000431570, HY10L69R7 (223173597, 223685788, 223685788); 00763000431822, M450311U (223303584, 223837444, 223908982); 00763000431839, M342507G (224865311, 225651642); 00763000431860, M422114D (224652877, 225273957); 00763000431877, M340615R (223393278, 223714094, 224096966, 224337418, 225090362, 225484631); 00763000432638, BB8S31R13 (223983970, 224923408); 00763000432669, HY11L82R1 (223366499, 223896483); 00763000432959, M330416E (223881865, 224048911, 224397994, 224865226, 225254273, 225447057); 00763000436612, M332410K (223864228); 00763000439828, BB5U30R11 (223366471, 224539686); 00763000439835, BB8U85R3 (223112591, 224539693); 00763000442613, P393001E (223260994, 223775521, 224206457, 224736692, 225346296); 00763000442750, M346203E (223508309, 223881863, 224047764, 225040700, 225434676, 225651643); 00763000445904, BB10J28R7 (223558533, 224327244, 225086634); 00763000455149, M399004Q (223714088, 224695402, 225208455); 00763000458072, HY10N63R6 (222755419, 224388050); 00763000458164, BB11A61R3 (223558486, 224112029, 224497813, 225008562); 00763000467074, BB8B22R14 (223002552); 00763000467487, M482406A (223532934, 223837990, 224252237, 225161991, 225651629); 00763000474058, BB9G98R8 (224196587, 224578585, 225170623); 00763000479428, M332831G (223303585, 223753452, 223993384, 224225698, 224865200); 00763000479626, BB11G35R3 (224826824); 00763000486723, M336213Q (223881862, 225120380, 225651512); 00763000486730, M336214P (223862062, 224077482, 224736686, 225120379, 225303457); 00763000487829, M331404M (223538843, 224271884, 225303512, 225534098); 00763000496548, P391802L (223508314, 223656860, 224271882, 224695307, 224865236, 225346288); 00763000497781, M342122D (223393281, 224297748, 224865238, 225434678); 00763000498443, HY10V75R8 (224028722, 224282020); 00763000498573, BB0Y86R39 (223558527, 224196574); 00763000499129, BB7U03R21 (223196579, 223685659, 223922497, 224327326, 224453934, 224923398); 00763000500962, BB10R11R5 (223095082, 223128090, 223323299, 223728558, 223728560, 223728561, 224112093, 224370601, 224370602, 224714330, 224794452, 225008565, 225008566, 225311121); 00763000501068, HY3P08R39 (224838255); 00763000501228, M337435C (223352899, 223713000, 223969950, 224652842, 224948504, 225389455); 00763000504137, M390361E (225331250); 00763000505417, HY2J36R23 (224923390); 00763000513214, BB10B05R15 (223305124, 224066122, 224539897, 225047883); 00763000516727, HY9D67R2 (223031034, 223093366); 00763000517854, HY9G55R5 (224743163, 225021492); 00763000518806, M411612T (223753453, 224033606); 00763000519254, BB8A70R13 (223127551, 223939277, 224664769); 00763000519919, P391805A (224456637); 00763000520205, M336111P (223735980); 00763000522810, M447215M (224096703, 224568266, 224948492); 00763000523176, HY10W75R5 (222899602, 223173593, 223409284); 00763000523213, M330211J (223789662, 224252239); 00763000524104, M330523B (223588904, 223735983); 00763000530419, BB8D56R15 (223127555, 223410326, 223728600, 224066161, 224282110, 224521225, 224743169, 224827932, 224908617, 224908618); 00763000532581, M331247C (223465272, 223789253, 224122616, 224225693, 224967449, 225354449); 00763000532680, BB9L88R22 (223054337, 224155210, 224327335, 224539788, 224879149); 00763000534561, HY6V44R36 (224699965, 224794363); 00763000535001, HY9L07R6 (224664715, 224996019); 00763000535018, HY9L80R6 (224453920, 224743173, 225008586); 00763000535070, HY8C67R16 (223112636, 223707560, 224196557, 224582609, 225086663); 00763000535407, BB8L56R9 (223138510, 224730556, 224732557); 00763000535841, HY1W12R36 (224963549); 00763000536206, HY10U52R7 (223196721, 223196722, 223196723, 223326043, 223326045, 223746645, 223747906, 223747907); 00763000539702, M300014B (223507398, 223862852, 224907229); 00763000541378, M337503Q (223612883, 223862855); 00763000541392, M336912J (224381033); 00763000542139, HY10U10R4 (223366492, 223804331, 224539881, 225008574); 00763000542146, HY10Q87R8 (223012945, 223650970, 224794440, 225219608); 00763000542153, HY10E56R13 (223366510, 223830711, 224281963, 224743174); 00763000542269, BB9H96R3 (224327205, 224838209); 00763000542566, BB11P45R5 (223189916); 00763000543969, HY11A91R9 (223155619, 223409261); 00763000544263, HY10B82R16 (223830712, 224327221, 224699948, 224811567); 00763000544379, HY7F85R17 (223112638, 223603565, 224281946, 224454057, 224879063); 00763000544546, M338803F (224096967, 224312701, 224735821); 00763000544607, M964909D (223547368, 224123953); 00763000544652, M332025J (223439636, 224138275, 224399257, 225049336, 225208457, 225651630); 00763000545888, BB9N38R9 (223173688, 223173689, 223283246, 223283247); 00763000546076, HY9Q42R9 (223196621, 224019281, 224327281, 224752724); 00763000546199, HY10U97R10 (223014745, 223685653, 223870495, 224066196, 224414982, 224830253, 224861214); 00763000549374, BB8X22R4 (223939262); 00763000550813, M343810D (223438639, 223589851, 224068598, 225303523); 00763000550837, M338014C (223422894, 223995012, 224948502, 225494182); 00763000550844, M335007M (223507401, 223588899, 223862632, 224007264, 224122615, 224312692, 224695542, 225161923); 00763000552015, BB1P07R23 (223830708, 224154987, 224794356); 00763000552060, M333715J (223745705, 224206458, 224780960, 225209647, 225436097); 00763000553890, M395912L (223508306, 224033577, 224736704, 225207739, 225561606); 00763000554095, HY7Z58R14 (223014656); 00763000554248, P392015A (223630915, 223918199); 00763000554286, BB2Q31R9 (224281952, 224923412); 00763000557171, BB1F66R24 (223283000, 224539887, 224963554); 00763000557188, HY4953R29 (224752787); 00763000557614, HY10X07R7 (223173591, 224135886); 00763000557621, HY10Z53R7 (223601660, 224415005); 00763000558413, BB8L58R13 (223066090, 223728552, 223896437); 00763000559687, M334908J (223672420, 223815906, 224440775, 224568270); 00763000564148, BB9H23R19 (223113115, 223644476, 224112056); 00763000564377, BB3C53R17 (223764130); 00763000569129, BB8W70R13 (222974838, 223663194, 223685567); 00763000573201, BB8S37R32 (222941495, 223558463, 224370599, 224752788); 00763000573836, BB0C80R37 (223112532); 00763000574079, HY10V50R1 (224734918); 00763000574666, BB8P76R4 (222941500, 223366541, 224066146, 224414946); 00763000576684, BB8P98R5 (223323355, 223983884, 224573409, 225170604); 00763000576752, HY11R94R1 (225270471); 00763000578329, HY11R00R6 (224784927, 224963587, 225270476, 225270477); 00763000578466, HY8N78R5 (224794445, 225398668); 00763000580698, BB10Q73R5 (223323133, 223873281); 00763000580704, BB11R26R (223323134, 223872619); 00763000582913, BB1D41R30 (223366508, 224155004, 224582627, 225086671); 00763000583002, BB6H26R40 (222916985, 223341605, 223804397, 224539866, 224796510, 224912392, 225086643); 00763000586645, M494516D (224865234); 00763000586669, HY9N98R14 (222974890, 223764201, 224281967, 224497786); 00763000589738, BB10J17R10 (223145657, 223851413, 224066181, 225337326); 00763000595319, M449117G (223790476, 224188645, 224865227, 225254269); 00763000596644, M333924W (223547374, 224168530, 224380962, 224695528, 224947972, 225090704, 225254278); 00763000597801, M411615A (223423119, 223745948); 00763000597818, M310892A (224907235; 00763000597870, HY10Q64R8 (223026510); 00763000600174, HY11P17R5 (223558538, 224453944, 224923387, 225219637); 00763000602666, HY9D67R3 (223254163, 223682449, 223682450, 224454019, 224539833, 224539834, 225270458); 00763000603021, M962803D (223393270, 224048912, 224711010); 00763000605193, HY7Z58R16 (223173685, 223196574); 00763000605292, BB11G87R3 (223130151, 223130152, 223558578, 224155216, 224282109, 224539801); 00763000606190, BB11S82R (223323219); 00763000608828, BB8B22R15 (223742624, 223744002, 223905178, 223983858, 223983859, 224066123, 224539888, 224664687, 224794427, 225008580, 225219626); 00763000608859, HY11Q67R2 (223213010, 224196590, 224730552); 00763000608958, M330350Q (223508315); 00763000611149, M337503R (224188641, 224651362); 00763000611170, M495422B (224012294); 00763000611293, BB11F25R2 (223939299, 224154913, 224908625); 00763000611644, M348104C (224679126, 225074770); 00763000611712, BB11A90R2 (223282899); 00763000612856, BB1845R50 (223282947, 223983968, 224281989, 224752778, 225219654); 00763000612931, M333519H (223837952); 00763000615284, HY9M47R19 (224854723, 225035305, 225035368, 225035675); 00763000616199, M382708B (223544697); 00763000616212, M310640B (223423007, 223631052); 00763000618735, BB11U53R (223341589, 223341591, 223341592, 223409251, 223498039, 223498041, 224664657, 224664658, 225169142, 225169143, 225169144, 225377168, 225377169, 225377170); 00763000618803, BB11T08R1 (223645659, 223804310, 224066091); 00763000619015, BB9H27R7 (223441265, 223833392, 223983162); 00763000620769, M311244B (223630911, 224048946, 224380961, 224960126, 225463283); 00763000620813, BB11U58R (223397188, 223558408, 223558409, 225170560, 225443635); 00763000625078, M331712E (224033569, 224354073, 224492614, 225651506); 00763000625092, M330521Q (223457947, 223968625, 224048947, 224206462, 224651361, 224907236, 225303452); 00763000625122, BB0E27R33 (223454120, 223939279, 224370583, 224730553, 224732556, 224908623); 00763000629878, HY11U75R2 (223463274, 223463275, 223463276); 00763000630225, BB11S99R2 (223462993, 223993306, 224277445, 224722407); 00763000632847, HY11V34R (224854721); 00763000633479, BB11V07R1 (223513739); 00763000633875, HY10W75R6 (223463269); 00763000633882, HY11A91R10 (223610879, 223610881, 223611017, 223613147, 223685787, 223896449, 223896450, 224066148, 224066149, 224066150, 224066151, 224344276, 224414993, 224414994); 00763000633899, M720623G (224047763, 224297736, 224610674); 00763000635022, M391908L (223575405, 223753454, 223862627, 224419073); 00763000637866, BB11V07R2 (223764559, 224539870, 224752793); 00763000637927, BB11L42R1 (224141302, 225270452); 00763000638924, BB7U21R21 (223764133, 224155079, 224794467); 00763000639181, M370519D (224330674); 00763000639815, M340208E (224217489); 00763000639846, BB11V67R1 (223666340, 223666343); 00763000640071, BB9H69R8 (223896478, 224327288, 224963527); 00763000640125, M382708C (223783763, 223908975, 224312702, 224865429, 225254287); 00763000640507, BB11V28R1 (223701901, 224019311, 224282009, 224752804); 00763000641962, M485003D (223784075); 00763000643768, BB11L73R5 (224674341); 00763000643775, HY10V01R7 (223905173); 00763000643782, HY11C28R10 (223858951, 224028723, 224344272, 224414996); 00763000643799, HY11R97R2 (223905172); 00763000644284, BB10Y64R5 (223804308, 223939219); 00763000645250, BB11V20R2 (223856935, 225398671); 00763000646721, M336310H (223995011, 224188639, 224390526); 00763000646981, BB10Q03R3 (224066077, 224146340, 225251243); 00763000647117, M450311W (224206729, 224864023, 225048105, 225253865, 225484319); 00763000647162, P392021A (224151381, 225049326); 00763000648350, HY9E87R10 (225086647); 00763000648428, BB11U86R1 (223911691, 224623750); 00763000648602, BB2493R47 (223931081, 224752712); 00763000648930, HY11U21R3 (225008555); 00763000650599, HY10U52R8 (224066188, 224066189, 224155193, 224387583, 224539898, 224582637, 224862882, 224862883, 224862884, 224862885, 225141913, 225141914, 225141915); 00763000651183, HY7Z58R17 (224277397, 224282101, 224861612); 00763000654740, M614023E (224396995, 225032739, 225338014); 00763000655174, BB8E21R11 (224071038); 00763000656072, BB11V68R2 (224134370, 224281886); 00763000657741, BB5W51R9 (224036258, 224155174, 224370604, 224827939); 00763000657765, BB8X22R9 (224879057); 00763000658588, BB11P45R9 (224061837, 224277423, 224277424, 224573050, 224752680, 224963533, 224963534); 00763000658601, P393216D (224115431);00763000658892, BB11W98R1 (224080416, 224539746); 00763000659585, BB11X15R1 (224082690, 224082693, 225104524); 00763000661083, M310640D (225090773); 00763000661090, M332410L (225494174); 00763000661106, M333519J (224297740); 00763000661489, BB11W97R2 (224277360, 224277361, 224838240); 00763000661496, BB8C31R5 (224162781, 224282090, 225008585, 225170614); 00763000661519, M330350R (224399249, 224610322, 224864022, 225161642, 225398438); 00763000663766, M411612U (224297712, 225041048, 225561517); 00763000665982, BB11V67R3 (224277390, 224539906); 00763000668211, P490607L (225090698); 00763000687632, BB11W30R5 (224277402, 224277403, 224664754, 225008582); 00763000687687, BB11T08R2 (224281905, 224327484, 224327485, 224582574, 224743165); 00763000687915, HY11N01R1 (224387577); 00763000687953, M331709P (224297524, 224381034, 224679487, 225463282); 00763000688158, HY2S09R20 (225126455); 00763000688844, HY11B96R3 (224453919, 224838220); 00763000689070, HY11S98R1 (224415022); 00763000689094, M448913A (224544065); 00763000689100, M484610A (224865433, 225389400); 00763000689148, BB11W92R1 (224313925, 224539703, 224752706); 00763000689377, BB5C09R10 (224327240, 224391692, 224453928, 224453930, 224539841); 00763000689445, BB11V88R2 (224327483, 224963599, 225492104); 00763000689919, BB11X99R2 (224303174, 225008572); 00763000699697, BB11X05R1 (224539844); 00763000699703, BB10J83R6 (225398707); 00763000700546, BB9N38R12 (224334590, 224414911, 224414913, 225126467, 225353756); 00763000701321, M310643G (225220421); 00763000701963, M614026B (224644207); 00763000702014, M490645J (224567874, 224865202, 225303451); 00763000702021, M470734A (224948453); 00763000702038, M460236G (224993570); 00763000702052, M252003C (224417843); 00763000703523, M410714C (224448472); 00763000704841, HY11W65R1 (224794464); 00763000704858, BB11W56R3 (224453916, 224453917); 00763000704865, BB11U19R4 (225425032); 00763000708191, BB8E21R13 (224463906, 224838241); 00763000712648, M336310J (224594409, 224594410); 00763000712716, HY10P39R5 (224794455); 00763000715366, M470879A (224948493); 00763000715748, BB11W98R2 (225047920); 00763000716196, HY9N98R16 (224875405); 00763000716684, HY11T97R6 (224752585); 00763000717629, HY2G71R7 (224664460); 00763000718862, BB0Y86R40 (224963528); 00763000721275, BB10S31R5 (224752674, 225270475); 00763000721374, BB10Y07R9 (225170628); 00763000723675, BB3C53R18 (224794252); 00763000723989, BB11S99R10 (225270581); 00763000724597, BB11V07R3 (225311189); 00763000726317, BB11U86R3 (224845842, 224980052); 00763000727307, HY10V01R8 (224734919, 224734920, 224734921); 00763000734251, BB11Z74R (225270369); 00763000736408, M484904D (225050406, 225210946); 00763000740214, M233110B (225227873); 00763000740320, BB11U86R5 (225008508); 00763000740900, M310566F (225330362); 00763000743406, M449119C (225089625); 00763000743413, M442233J (225081530, 225208463); 00763000743444, M981001F (225090934); 00763000746902, BB4Q12R19 (225219615); 00763000747435, M334908K (225331456); 00763000748159, BB11A90R5 (225251333); 00763000752880, M450202U (225528160); 00763000754198, BB12A07R (225259471); 00763000760113, M337503T (225651906); 00763000760229, BB11Y04R2 (225360620); 00763000761820, M333519K (225389332); 00763000761868, BB7Q27R10 (225408607); 00763000762339, HY11V34R1 (225424991); 00763000763831, M332520A (225668319); 00763000770877, M332521A (225633224); 20643169863334, M332111B (223482422, 223815834, 224008704, 224526153, 225162263, 225651639); 20763000129440, M321634E (223631050); 20763000300085, M495227J (224865319); 20763000338477, M580440B (223422989, 223631056, 223815833, 223995015, 224225696, 224354190, 224736969, 225120477, 225528158); 20763000417677, M497549P (225049373); 20763000544526, M332710N (223538838, 223968623, 224122996, 224312703, 224526150, 225254274, 225389401); 20763000544571, M493718D (224695543); 20763000550824, M321020G (223656861); 20763000573861, M332611H (223790479, 224311663, 224567883, 225228307, 225651511); 20763000611105, M311041M (223837963, 224399255, 225162262); 20763000645599, M321634F (223952993, 224123000, 224652826, 224907241, 225209651, 225369616)

Devices named in this recall

211 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.

Product lines

Other devices under the same FDA submission

6 devices cleared under K203111 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91343).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page