Skip to content

NA BB11U84R5Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB11U84R5 4PK ADVENTHEALTH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000746032), received between Oct 23, 2024 and Nov 1, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Component Missing” (1 report).

25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing1
  • Crack1
  • Material Puncture/Hole1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 1, 2024Material Separation; Component Missing; Material Separation; Component Missing

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE TUBING ONTO THE CONNECTOR WAS LOOSE AND MISSING THE EXTRA PLASTIC SUPPORT BAND. THE LOOSE/SEPARATED COMPONENT DID NOT FALL INTO THE PATIENT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20585737 ↗
  • MalfunctionOct 23, 2024Crack; Material Puncture/Hole; Crack; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE CONNECTOR ON THE CRYSTALLOID CARDIOPLEGIA SIDE OF THE 4:1 CUSTOM SET CONNECTOR WAS CRACKED AND THE HEAT SHRINK BAND WAS NOT SECURE AROUND THE TUBING THAT THE CONNECTOR FITS INTO. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514320 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.

  • Z-1881-2021 · Apr 29, 2021 · Terminated
    Medtronic CB4W67R20, Custom Pack CB4W67R20 Adult E, Custom Perfusion System, Sterile, Rx.
  • Z-1871-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10Z48R2, Ctr., Custom Pack HY10Z48R2 Adult ECC 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1880-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1878-2021 · Apr 29, 2021 · Terminated
    Medtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
  • Z-1894-2021 · Apr 29, 2021 · Terminated
    Medtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
  • Z-1872-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
  • Z-1874-2021 · Apr 29, 2021 · Terminated
    Medtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
  • Z-1889-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1884-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10W65R1, Custom Pack CB10W65R1 CH Access1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1893-2021 · Apr 29, 2021 · Terminated
    Medtronic BB7D76R13, Custom Pack BB7D76R13 CORR HL, Custom Perfusion System, Sterile, Rx.
  • Z-1875-2021 · Apr 29, 2021 · Terminated
    Medtronic HY7M07R6, Custom Pack HY7M07R6 E 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1888-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
  • Z-1873-2021 · Apr 29, 2021 · Terminated
    Medtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
  • Z-1883-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10J00R12, Custom Pack HY10J00R12 ECC Small CA, Custom Perfusion System,
  • Z-1895-2021 · Apr 29, 2021 · Terminated
    Medtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.

…and 10 more.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page