NA BB11X80R3Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB11X80R3 BB PUMP · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000763985), received between Aug 3, 2023 and Aug 22, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
- Material Separation2
- Air/Gas in Device1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hemorrhage/Blood Loss/Bleeding2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE CONNECTORS IN THE AV LOOP STARTED LEAKING WHEN THE PRIMING BEGAN. THE CUSTOMER OBSERVED THERE WAS NO SEAL AT THE TUBING AND CONNECTOR. THE CUSTOMER TRIED ADDING AN EXTRA TIE BAND BUT THE LEAK CONTINUED SO THE CUSTOMER CUT OUT AND REPLACED THE CONNECTOR. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17594433 ↗MalfunctionAug 8, 2023Fluid/Blood Leak; Material Separation; Air/Gas in Device; Fluid/Blood Leak; Material Separation; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM PACK, IT WAS REPORTED THAT WHILE PRIMING, THE PERFUSIONIST NOTICED FLUID LEAKING FROM A CONNECTOR IN THE ROLLER TUBING LINE. THE USE OF THE DEVICE IS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. THE CUSTOMER WAS ABLE TO REMOVE TIE BAND ON THE CONNECTOR AND REPLACE IT WITH TWO NEW TIE BANDS WHICH ALLEVIATED THE LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17480668 ↗MalfunctionAug 4, 2023Material Puncture/Hole; Material Puncture/Hole
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE CUSTOMER NOTICED AN ABNORMALITY AT THE TUBING/CONNECTOR POINT IN THE ROLLER PUMP LINE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17456934 ↗MalfunctionAug 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED ONCE ON PUMP, THE CUSTOMER OBSERVED A SLOW BLOOD LEAK AT ONE OF THE CONNECTIONS IN THE ROLLER BOOT WHERE THE TUBING TRANSITIONS FROM 3/8 TO 1/2. IT WAS A SLOW LEAK AT THE CONNECTION SO THEY DID NOT CHANGE IT OUT. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 17448466 ↗MalfunctionAug 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT ONCE ON PUMP, THE CUSTOMER OBSERVED A LEAK IN THE ROLLER BOOT CONNECTION WHERE THE TUBING TRANSITIONS FROM 1/2 TO 3/8. IT WAS A VERY SLOW LEAK SO THEY DID NOT CHANGE OUT THE TUBING. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 17448375 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.
- Z-1890-2021 · Apr 29, 2021 · TerminatedMedtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
- Z-1880-2021 · Apr 29, 2021 · TerminatedMedtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
- Z-1886-2021 · Apr 29, 2021 · TerminatedMedtronic HY10D90R2, Custom Pack HY10D90R2 3/8 ACC, Custom Perfusion System, Sterile, Rx.
- Z-1879-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W63R1, Custom Pack CB10W63R1 HL 20 1/4, Custom Perfusion System, Sterile, Rx.
- Z-1873-2021 · Apr 29, 2021 · TerminatedMedtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
- Z-1894-2021 · Apr 29, 2021 · TerminatedMedtronic BB9G78R8, Custom Pack BB9G78R8 ECC Supp, Custom Perfusion System, Sterile, Rx.
- Z-1887-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R79R6, Custom Pack CB10R79R6 1/4X1/4 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1874-2021 · Apr 29, 2021 · TerminatedMedtronic BB11J14R2, Custom Pack BB11J14R2 Bridge, Custom Perfusion System, Sterile, Rx.
- Z-1892-2021 · Apr 29, 2021 · TerminatedMedtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.
- Z-1888-2021 · Apr 29, 2021 · TerminatedMedtronic CB10R55R5, Custom Pack CB10R55R5 3/8X3/8 Loop, Custom Perfusion System, Sterile, Rx.
- Z-1877-2021 · Apr 29, 2021 · TerminatedMedtronic HY11B40R1, Custom Pack HY11B40R1 1/4 Respirato, Custom Perfusion System, Sterile, Rx.
- Z-1883-2021 · Apr 29, 2021 · TerminatedMedtronic HY10J00R12, Custom Pack HY10J00R12 ECC Small CA, Custom Perfusion System,
- Z-1891-2021 · Apr 29, 2021 · TerminatedMedtronic CB10Y89R10, Custom Pack CB10Y89R10 ECC 3/8 A/V, Custom Perfusion System, Sterile, Rx.
- Z-1882-2021 · Apr 29, 2021 · TerminatedMedtronic CB10W64R1, Custom Pack CB10W64R1 CH Access 3/8, Custom Perfusion System, Sterile, Rx.
- Z-1881-2021 · Apr 29, 2021 · TerminatedMedtronic CB4W67R20, Custom Pack CB4W67R20 Adult E, Custom Perfusion System, Sterile, Rx.
…and 10 more.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page