Axium™ Detachable Coil System QC-4-10-3DRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-4-10-3D V14 · Class 2 · Implant

All models of Axium Detachable Coil System (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000845872), received between Jun 30, 2026 and Jun 30, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2026Separation Failure; Material Deformation

    MEDTRONIC RECEIVED A REPORT THAT THERE WAS AXIUM COIL NON-DETACHMENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, RUPTURED INTERNAL CAROTID ARTERY (ICA) BIFURCATION BRANCH ANEURYSM WITH A MAX DIAMETER OF 6 MM AND A 3.5 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS SEVERE. IT WAS REPORTED THAT THE PHYSICIAN ATTEMPTED TO DETACH BOTH COILS USING ID-1-5 INSTANT DETACHER MULTIPLE TIMES. COIL DID NOT DETACH. PHYSICIAN USED THE BACK UP METHOD AND BROKE THE WIRE AND PULLED THE SUTURE AND BOTH COILS STILL DID NOT DETACH. THE SUTURE STRETCHED IN BOTH CASES. THE PHYSICIAN THEN PULLED HARDER AND MORE AGGRESSIVELY AND THE COILS FINALLY RELEASED. THERE WAS COIL NON-DETACHMENT. THE PHYSICIAN ATTEMPTED TO DETACH THE COIL WITH THE INSTANT DETACHER 3 TIMES AND AN ADDITIONAL 3 TIMES WITH A SECOND INSTANT DETACHER. THERE WERE TWO MANUAL ATTEMPTS AT DETACHMENT VIA HYPOTUBE. THERE WAS NO ISSUE WITH THE INSTANT DETACHER. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. ANCILLARY DEVICES INCLUDE A BENCHMARK BMK6F105, SL10 M00316890, A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25673369 ↗

Other Device, Vascular, for Promoting Embolization devices