Skip to content

Ascenda® 8782 (UDI 00763000852610)Recalls & reported problems

Medtronic, Inc. · CATH 8782 ASCENDA SPINAL REV KIT EMAN · Class 3

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000852610), received between Jul 23, 2025 and Aug 5, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Difficult to Advance” (3 reports).

40 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance3
  • Material Integrity Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 5, 2025Difficult to Advance; Material Integrity Problem; Difficult to Advance; Material Integrity Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING CATHETERS FOR AN IMPLANTABLE INTRATHECAL PUMP. IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO THREAD THE CATHETER CEPHALAD; THEY KEPT WANTED TO GO DOWN INSTEAD OF UP. AN EXTERNAL FACTOR WAS THAT THE PHYSICIAN WAS ENTERING THE INTRATHECAL SPACE WITHOUT THE USE OF LIVE FLUOROSCOPY AT A 90 DEGREE ANGLE. THE PHYSICIAN USED SPOT CHECKS INTERMITTENTLY INSTEAD OF LIVE FLUOROSCOPY. THE PHYSICIAN SHEARED THE CATHETER. A SPINAL SEGMENT WAS GIVEN TO THE PHYSICIAN AND THE PHYSICIAN SHEARED THE SPINAL SEGMENT, AS WELL AS THE NEXT TWO SPINAL SEGMENTS GIVEN TO THEM. THE PHYSICIAN COMPLAINED THAT THE NEEDLE WAS "CRAPPY" AND BECAME BLUNT/BENT UPON ENTRY AND ASKED FOR A NEW NEEDLE. THE MANUFACTURER REPRESENTATIVE GAVE THE PHYSICIAN THEIR LAST CATHETER AND THE SAME ISSUE RECURRED WITH THE INABILITY TO THREAD THE CATHETER CEPHALAD INSTEAD OF DOWN, AND THE LAST CATHETER WAS SHEARED. THE CASE WAS ABORTED DUE TO THE INABILITY TO PROPERLY PLACE THE CATHETER AND RESCHEDULED FOR (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22717627 ↗
  • InjuryJul 23, 2025Difficult to Advance; Material Integrity Problem; Difficult to Advance; Material Integrity Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING CATHETERS FOR AN IMPLANTABLE INTRATHECAL PUMP. IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO THREAD THE CATHETER CEPHALAD; THEY KEPT WANTED TO GO DOWN INSTEAD OF UP. AN EXTERNAL FACTOR WAS THAT THE PHYSICIAN WAS ENTERING THE INTRATHECAL SPACE WITHOUT THE USE OF LIVE FLUOROSCOPY AT A 90 DEGREE ANGLE. THE PHYSICIAN USED SPOT CHECKS INTERMITTENTLY INSTEAD OF LIVE FLUOROSCOPY. THE PHYSICIAN SHEARED THE CATHETER. A SPINAL SEGMENT WAS GIVEN TO THE PHYSICIAN AND THE PHYSICIAN SHEARED THE SPINAL SEGMENT, AS WELL AS THE NEXT TWO SPINAL SEGMENTS GIVEN TO THEM. THE PHYSICIAN COMPLAINED THAT THE NEEDLE WAS "CRAPPY" AND BECAME BLUNT/BENT UPON ENTRY AND ASKED FOR A NEW NEEDLE. THE MANUFACTURER REPRESENTATIVE GAVE THE PHYSICIAN THEIR LAST CATHETER AND THE SAME ISSUE RECURRED WITH THE INABILITY TO THREAD THE CATHETER CEPHALAD INSTEAD OF DOWN, AND THE LAST CATHETER WAS SHEARED. THE CASE WAS ABORTED DUE TO THE INABILITY TO PROPERLY PLACE THE CATHETER AND RESCHEDULED FOR (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22583416 ↗
  • InjuryJul 23, 2025Difficult to Advance; Material Integrity Problem; Difficult to Advance; Material Integrity Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING CATHETERS FOR AN IMPLANTABLE INTRATHECAL PUMP. IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO THREAD THE CATHETER CEPHALAD; THEY KEPT WANTED TO GO DOWN INSTEAD OF UP. AN EXTERNAL FACTOR WAS THAT THE PHYSICIAN WAS ENTERING THE INTRATHECAL SPACE WITHOUT THE USE OF LIVE FLUOROSCOPY AT A 90 DEGREE ANGLE. THE PHYSICIAN USED SPOT CHECKS INTERMITTENTLY INSTEAD OF LIVE FLUOROSCOPY. THE PHYSICIAN SHEARED THE CATHETER. A SPINAL SEGMENT WAS GIVEN TO THE PHYSICIAN AND THE PHYSICIAN SHEARED THE SPINAL SEGMENT, AS WELL AS THE NEXT TWO SPINAL SEGMENTS GIVEN TO THEM.  THE PHYSICIAN COMPLAINED THAT THE NEEDLE WAS "CRAPPY" AND BECAME BLUNT/BENT UPON ENTRY AND ASKED FOR A NEW NEEDLE. THE MANUFACTURER REPRESENTATIVE GAVE THE PHYSICIAN THEIR LAST CATHETER AND THE SAME ISSUE RECURRED WITH THE INABILITY TO THREAD THE CATHETER CEPHALAD INSTEAD OF DOWN, AND THE LAST CATHETER WAS SHEARED. THE CASE WAS ABORTED DUE TO THE INABILITY TO PROPERLY PLACE THE CATHETER AND RESCHEDULED FOR (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22583414 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.

  • Z-2201-2026 · Apr 8, 2026 · Open, Classified
    Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
  • Z-1506-2025 · Jan 9, 2025 · Open, Classified
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
  • Z-2352-2024 · May 10, 2024 · Open, Classified
    Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
  • Z-0615-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-40, Programmable pump
  • Z-0614-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-20, Programmable pump
  • Z-0522-2024 · Jul 13, 2020 · Completed
    Medtronic SynchroMed, Model A10
  • Z-0508-2020 · Oct 11, 2019 · Terminated
    Synchromed II programmable pump, Model numbers 8637-20 and 8637-40
  • Z-0537-2018 · Sep 21, 2017 · Terminated
    Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
  • Z-0896-2018 · Aug 9, 2017 · Open, Classified
    Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
  • Z-0788-2017 · Oct 3, 2016 · Terminated
    Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
  • Z-1271-2016 · Feb 9, 2016 · Terminated
    Medtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
  • Z-1681-2015 · Apr 10, 2015 · Terminated
    Medtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
  • Z-1062-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
  • Z-1061-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
  • Z-1060-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow

…and 25 more.

Other Pump, infusion, implanted, programmable devices

All Pump, infusion, implanted, programmable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.