Ascenda® 8784 (UDI 00763000916671)Recalls & reported problems
Medtronic, Inc. · CATH 8784 ASCENDA PUMP SEG REV KIT EMAN · Class 3
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000916671), received between Jan 22, 2025 and Sep 28, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
40 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Connection Problem1
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Disconnection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50924257 ↗MalfunctionDec 4, 2025Connection Problem; Connection Problem
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA COMPANY REPRESENTATIVE REGARDING AN IMPLANTABLE CATHETER. IT WAS REPORTED THAT THE COLLET CLAMPED BEFORE IMPLANTING. IT SEEMED TO OCCUR WITH MUCH FORCE THAN USUALLY REQUIRED. THE PRODUCT WAS NEVER IMPLANTED. IT WAS OPENED BUT NOT USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23720634 ↗MalfunctionOct 17, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS RECEIV ING MORPHINE (4MG/ML AT 0.25 MG/DAY) VIA AN IMPLANTABLE PUMP. IT WAS REPORTED THAT THE COLLET IN KIT WAS NON FUNCTIONING AS THE PLASTIC BROKE TO THE DISTAL TIP. THEY OPENED THE PRODUCT ON THE FIELD AND THERE WERE NO ISSUES BEFORE OPENING THE PACKAGING. THEY WERE UNABLE TO TROUBLESHOOT AS THE PRODUCT WAS NOT FUNCTIONING. THEY OPENED A NEW 8785 FOR AN ADDITIONAL COLLET TO COMPLETE THE CASE. THE ISSUE WAS RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23324985 ↗MalfunctionJan 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ON 2025-JAN-16, INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP), REGARDING A PATIENT RECEIVING MORPHINE (10 MG/ML, 1.299 MG) AND BUPIVACAINE (5 MG/ML, 0.6497) VIA AN IMPLANTABLE PUMP. IT WAS REPORTED A PUMP PROCEDURE (SEE RELATED EVENT), WHILE USING A 8784 TO TUNNEL A NEW PUMP SEGMENT TO NEW POCKET, THE COLLET OR "MUSHROOM PART THAT CLAPS THE CATHETER" FELL / BROKE OFF WHEN THE HEALTHCARE PROFESSIONAL (HCP) WAS THREADING THE CATHETER INTO THE COLLET. THE REP REPORTED THERE WAS NO "ROUGH PLAY" AS THEY WERE "WATCHING THE WHOLE TIME". A NEW 8784 WAS OPENED TO GET A NEW COLLET. THE REPORTED ACTIONS / INTERVENTIONS THAT TOOK PLACE TO RESOLVE THE EVENT WERE "HAD TO UTILIZE 2 DIFFERENT 8784 IN ORDER TO CONNECT NEW PUMP SEGMENT TO ALREADY PLACED SPINAL SEGMENT. THE CATHETER WAS ATTACHED AND THE PUMP SEGMENT WAS ATTACHED TO THE PUMP WITH EASY ASPIRATION OF CEREBROSPINAL FLUID (CSF). THE POCKET WAS CLOSED WITH EVERYTHING RUNNING AS IT SHOULD. THE ISSUE WAS RESOLVED AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21211866 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.
- Z-2201-2026 · Apr 8, 2026 · Open, ClassifiedMedtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
- Z-1506-2025 · Jan 9, 2025 · Open, ClassifiedA820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
- Z-2352-2024 · May 10, 2024 · Open, ClassifiedMedtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
- Z-0615-2024 · Nov 13, 2023 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-40, Programmable pump
- Z-0614-2024 · Nov 13, 2023 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-20, Programmable pump
- Z-0522-2024 · Jul 13, 2020 · CompletedMedtronic SynchroMed, Model A10
- Z-0508-2020 · Oct 11, 2019 · TerminatedSynchromed II programmable pump, Model numbers 8637-20 and 8637-40
- Z-0537-2018 · Sep 21, 2017 · TerminatedMedtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
- Z-0896-2018 · Aug 9, 2017 · Open, ClassifiedMedtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
- Z-0788-2017 · Oct 3, 2016 · TerminatedMedtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
- Z-1271-2016 · Feb 9, 2016 · TerminatedMedtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
- Z-1681-2015 · Apr 10, 2015 · TerminatedMedtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
- Z-1061-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
- Z-1062-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
- Z-1060-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
…and 25 more.
Other Pump, infusion, implanted, programmable devices
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.