The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000916671), received between Jan 22, 2025 and Sep 28, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

40 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Connection Problem1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Disconnection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924257 ↗
  • MalfunctionDec 4, 2025Connection Problem; Connection Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA COMPANY REPRESENTATIVE REGARDING AN IMPLANTABLE CATHETER. IT WAS REPORTED THAT THE COLLET CLAMPED BEFORE IMPLANTING. IT SEEMED TO OCCUR WITH MUCH FORCE THAN USUALLY REQUIRED. THE PRODUCT WAS NEVER IMPLANTED. IT WAS OPENED BUT NOT USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23720634 ↗
  • MalfunctionOct 17, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS RECEIV ING MORPHINE (4MG/ML AT 0.25 MG/DAY) VIA AN IMPLANTABLE PUMP. IT WAS REPORTED THAT THE COLLET IN KIT WAS NON FUNCTIONING AS THE PLASTIC BROKE TO THE DISTAL TIP. THEY OPENED THE PRODUCT ON THE FIELD AND THERE WERE NO ISSUES BEFORE OPENING THE PACKAGING. THEY WERE UNABLE TO TROUBLESHOOT AS THE PRODUCT WAS NOT FUNCTIONING. THEY OPENED A NEW 8785 FOR AN ADDITIONAL COLLET TO COMPLETE THE CASE. THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23324985 ↗
  • MalfunctionJan 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    ON 2025-JAN-16, INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP), REGARDING A PATIENT RECEIVING MORPHINE (10 MG/ML, 1.299 MG) AND BUPIVACAINE (5 MG/ML, 0.6497) VIA AN IMPLANTABLE PUMP. IT WAS REPORTED A PUMP PROCEDURE (SEE RELATED EVENT), WHILE USING A 8784 TO TUNNEL A NEW PUMP SEGMENT TO NEW POCKET, THE COLLET OR "MUSHROOM PART THAT CLAPS THE CATHETER" FELL / BROKE OFF WHEN THE HEALTHCARE PROFESSIONAL (HCP) WAS THREADING THE CATHETER INTO THE COLLET. THE REP REPORTED THERE WAS NO "ROUGH PLAY" AS THEY WERE "WATCHING THE WHOLE TIME". A NEW 8784 WAS OPENED TO GET A NEW COLLET. THE REPORTED ACTIONS / INTERVENTIONS THAT TOOK PLACE TO RESOLVE THE EVENT WERE "HAD TO UTILIZE 2 DIFFERENT 8784 IN ORDER TO CONNECT NEW PUMP SEGMENT TO ALREADY PLACED SPINAL SEGMENT. THE CATHETER WAS ATTACHED AND THE PUMP SEGMENT WAS ATTACHED TO THE PUMP WITH EASY ASPIRATION OF CEREBROSPINAL FLUID (CSF). THE POCKET WAS CLOSED WITH EVERYTHING RUNNING AS IT SHOULD. THE ISSUE WAS RESOLVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21211866 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.

  • Z-2201-2026 · Apr 8, 2026 · Open, Classified
    Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
  • Z-1506-2025 · Jan 9, 2025 · Open, Classified
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
  • Z-2352-2024 · May 10, 2024 · Open, Classified
    Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
  • Z-0615-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-40, Programmable pump
  • Z-0614-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-20, Programmable pump
  • Z-0522-2024 · Jul 13, 2020 · Completed
    Medtronic SynchroMed, Model A10
  • Z-0508-2020 · Oct 11, 2019 · Terminated
    Synchromed II programmable pump, Model numbers 8637-20 and 8637-40
  • Z-0537-2018 · Sep 21, 2017 · Terminated
    Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
  • Z-0896-2018 · Aug 9, 2017 · Open, Classified
    Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
  • Z-0788-2017 · Oct 3, 2016 · Terminated
    Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
  • Z-1271-2016 · Feb 9, 2016 · Terminated
    Medtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
  • Z-1681-2015 · Apr 10, 2015 · Terminated
    Medtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
  • Z-1061-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
  • Z-1062-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
  • Z-1060-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow

…and 25 more.

Other Pump, infusion, implanted, programmable devices

All Pump, infusion, implanted, programmable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.