Skip to content

NA BB7L41R27Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB7L41R27 OR ADULT 3/8 · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000920043), received between Oct 29, 2024 and Oct 29, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

25 other recalls cover devices cleared under the same FDA submission (K171979); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2024Obstruction of Flow; Obstruction of Flow

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THE VACUUM RELIEF VALVE (VRV) IN THE YELLOW SUCKER LINE WAS NOT WORKING. IT WOULD NOT LET FLUID PASS WHEN IT WAS TESTED. THE CUSTOMER CHECKED THE PUMP OCCLUSION AND POSITIONING DIRECTION OF THE VALVE. HOWEVER, THEY BOTH WERE NOT THE CAUSE OF THE ISSUE. THERE WAS NO PACKAGE DAMAGE. THE VALVE WAS REPLACED WITH ANOTHER ONE THEY HAD SEPARATE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. . MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE ISSUE WAS DETECTED PRIOR TO USE WHEN THE DEVICE WAS BEING TESTED. THE BLOCKAGE WAS DISCOVERED PRE-BYPASS. THE CUSTOMER STATED THAT NORMOSOL WAS USED FOR PRIMING BUT THIS WAS REALLY NOT APPLICABLE BECAUSE THE SUCKER LINES WERE NOT PRIMED. THE CUSTOMER STATED THAT THEY USED A HMS PLUS TO ASSESS ANTI-COAGULATION (HEPARIN LEVEL).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20557133 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171979. One submission often covers many models, so check the product and lot details in each record.

  • Z-1893-2021 · Apr 29, 2021 · Terminated
    Medtronic BB7D76R13, Custom Pack BB7D76R13 CORR HL, Custom Perfusion System, Sterile, Rx.
  • Z-1878-2021 · Apr 29, 2021 · Terminated
    Medtronic CB1H18R20, Custom Pack CB1H18R20 Accesory, Custom Perfusion System, Sterile, Rx.
  • Z-1883-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10J00R12, Custom Pack HY10J00R12 ECC Small CA, Custom Perfusion System,
  • Z-1895-2021 · Apr 29, 2021 · Terminated
    Medtronic 1A46R11, Custom Pack 1A46R11 Neonatal, Custom Perfusion System, Sterile, Rx.
  • Z-1889-2021 · Apr 29, 2021 · Terminated
    Medtronic CB7J98R7, Custom Pack CB7J98R7, Custom Pack CB7J98R7 3/8 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1872-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8M63R4, Custom Pack HY8M63R4 Adult ECC, Custom Perfusion System, Sterile, Rx.
  • Z-1880-2021 · Apr 29, 2021 · Terminated
    Medtronic HY8R62R13,Custom Pack HY8R62R13 E Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1873-2021 · Apr 29, 2021 · Terminated
    Medtronic BB10R94R6, Custom Pack BB10R94R6 1/4 ECCPK, Custom Perfusion System, Sterile, Rx.
  • Z-1886-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10D90R2, Custom Pack HY10D90R2 3/8 ACC, Custom Perfusion System, Sterile, Rx.
  • Z-1890-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10L28R3, Custom Pack HY10L28R3 1/4 Pack, Custom Perfusion System, Sterile, Rx.
  • Z-1892-2021 · Apr 29, 2021 · Terminated
    Medtronic BB8M63R5, Custom Pack BB8M63R5 Adult ECC, Custom Perfusion system, Sterile, Rx.
  • Z-1885-2021 · Apr 29, 2021 · Terminated
    Medtronic HY10Y54R4, Custom Pack HY10Y54R4 ECC PED 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1887-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10R79R6, Custom Pack CB10R79R6 1/4X1/4 Loop, Custom Perfusion System, Sterile, Rx.
  • Z-1879-2021 · Apr 29, 2021 · Terminated
    Medtronic CB10W63R1, Custom Pack CB10W63R1 HL 20 1/4, Custom Perfusion System, Sterile, Rx.
  • Z-1875-2021 · Apr 29, 2021 · Terminated
    Medtronic HY7M07R6, Custom Pack HY7M07R6 E 1/4, Custom Perfusion System, Sterile, Rx.

…and 10 more.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.