Skip to content

Hancock™ 150 HC150-14Recalls & reported problems

Medtronic, Inc. · HANCOCK VALVED CONDUIT MO 14MM · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000926472), received between Mar 27, 2025 and Mar 27, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size1
  • Use of Device Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 27, 2025Inadequacy of Device Shape and/or Size; Use of Device Problem; Inadequacy of Device Shape and/or Size; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION THAT THIS 14MM PULMONARY VALVED CONDUIT HAD AN ACTIVATED TEMPERATURE INDICATOR. IT WAS THEN CONFIRMED THAT THE DEVICE WAS IMPLANTED. THE SURGEON WAS INFORMED PRIOR TO USE OF THE DEVICE THAT THE CONDUIT SHOULD NOT BE USED IF THE TEMPERATURE INDICATOR IS ACTIVATED. SUBSEQUENTLY 8 DAYS LATER, THE VALVE WAS EXPLANTED. THE REASON FOR REPLACEMENT WAS REPORTED AS THE VALVE "WAS TOO BIG". IT WAS REPORTED THAT PRIOR TO USE, THE VALVE WAS STORED IN THE OPERATING ROOM WITH CONSTANT TEMPERATURE, ON AN OPEN CART. IT WAS STATED THAT THE OPERATING ROOM WAS THERMOSTAT CONTROLLED AND THE VALVE DID NOT EXPERIENCE KNOWN EXTREME TEMPERATURES. IT WAS NOTICED THAT THE TEMPERATURE INDICATOR HAD BEEN ACTIVATED AFTER BEING ON THE SHELF FOR APPROXIMATELY 3 WEEKS AND SOME OTHER VALVES STORED IN THE SAME LOCATION HAD INDICATORS WHICH STARTED TO ACTIVATE BUT WERE NOT FULLY BLACK. IT WAS REPORTED THAT THE VALVES WITH TEMPERATURE INDICATORS WHICH WERE NOT FULLY BLACK, HAVE BEEN LIKE THAT FOR AWHILE WITH NO CHANGE AND WERE ON THE SELF PRIOR TO THE HANCOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21708279 ↗