Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000936624), received between Sep 24, 2026 and Sep 24, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Physical Resistance/Sticking1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Difficult to Advance; Physical Resistance/Sticking; Premature Separation

    It was reported that during treatment for ¿Hypo take down,¿ a CONCERTO VERSA coil was prepared as indicated in the instructions for use and tracking difficulty occurred during initial advancement and resistance/sticking occurred in the catheter at the hub. Physician hubbed introducer sheath to catheter and felt resistance advancing into a 2.8 non-Medtronic microcatheter. Physician pulled back and tried to re-advance and felt resistance again. When physician pulled coil back , the coil had prematurely detached inside the microcatheter. The pusher release wire was bent/broken. The detached coil was removed from the patient. A replacement Concerto Versa CV-3-15 coil was used to complete the procedure and the issue was resolved. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880892 ↗