Skip to content

Concerto Versa™ CV-12-50 (UDI 00763000936761)Recalls & reported problems

Medtronic, Inc. · COIL CONCERTO VERSA CV-12-50 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000936761), received between Jul 2, 2026 and Jul 2, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Material Deformation1
  • Physical Resistance/Sticking1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2026Difficult to Remove; Material Deformation; Physical Resistance/Sticking; Premature Separation

    IT WAS REPORTED THAT A 12X50 CONCERTO VERSA COIL WAS USED DURING AN EMBOLIZATION PROCEDURE TO TREAT AN UNDERPERFUSED TIPS STENT IN THE PATIENTS HEPATIC VARIX. MINIMAL VESSEL TORTUOSITY WAS REPORTED. A 5FR NON-MEDTRONIC (KA2 MERIT MEDICAL) DIAGNOSTIC CATHETER AND A 4CM 6FR NON-MEDTRONIC (TERUMO) SHEATH WERE USED. THE DEVICE WAS PREPPED PER THE IFU WITHOUT ISSUE. TRACKING DIFFICULTY WAS NOTED DURING INITIAL ADVANCEMENT, AND RESISTANCE WAS ENCOUNTERED AT THE CATHETER HUB AS THE COIL ADVANCED WITHIN THE (KMP) CATHETER. PHYSICIAN USED THE CHEATER TO ADD SUPPORT. PHYSICIAN MAY HAVE KINKED THE PUSHER/RELEASE WIRE AT SOME POINT BEFORE USING THE CHEATER. THE IMPLANT COIL WAS STRETCHED. A CONTINUOUS HEPARINIZED SALINE FLUSH WAS NOT ADMINISTERED DURING THE PROCEDURE. DURING AN ATTEMPTED SINGLE REPOSITIONING, THE COIL DID NOT PULL BACK WITH THE PUSHER AND THE COIL DETACHED PREMATURELY DURING TRACKING INSIDE THE MICROCATHETER/DIAGNOSTIC CATHETER. PHYSICIAN TRIED TO USE THE CATHETER TO PACK IT IN BUT WASN¿T WORKING WELL, PHYSICIAN THEN TRIED TO JUST PULL THE SYSTEM BACK BUT THE COIL WASN¿T COMING WITH THE CATHETER. THE COIL WAS RETRIEVED AND REMOVED WHILE MAINTAINING SHEATH ACCESS BY USING REPEA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25688443 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates