Bio-Medicus™ 96530-119 (UDI 00763000947088)Recalls & reported problems
Medtronic, Inc. · CANNULA 96530-119 BIO-MEDICUS 19FR · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000947088), received between Feb 26, 2026 and Apr 16, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K153598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem2
- Material Deformation1
Patient problems
- Aortic dissection2
- Vascular Dissection2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2026Material Deformation
IT WAS REPORTED THAT DURING USE OF THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL CANNULA, THE DEVICE WAS NOT PERFORMING EFFECTIVELY (POOR EFFECT). THE CANNULA WAS REMOVED, AND DEFORMATION OF THE CANNULA BODY WAS OBSERVED AFTER REMOVAL. THE CANNULA WAS REPLACED WITH ANOTHER PRODUCT. IT WAS UNKNOWN IF THERE WAS ANY COMPLICATIONS AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24902634 ↗InjuryMar 9, 2026Mechanical Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL CANNULA, IT WAS REPORTED THAT THE CANNULA TIP WAS SHARP AND RELATIVELY HARD, MAKING IT PRONE TO VASCULAR INJURY DURING PUNCTURE AND INSERTION.THE CANNULA WAS USED TO ESTABLISH CARDIOPULMONARY BYPASS, AND AN AORTIC DISSECTION OCCURRED. UNDER ULTRASOUND GUIDANCE, THE ARTERY WAS CANNULATED USING THE PUNCTURE METHOD. THE PATIENT SUFFERED ARTERIAL INJURY, RESULTING IN AN AORTIC DISSECTION. THE PATIENT WAS URGENTLY TRANSFERRED TO THE OPERATING ROOM FOR DESCENDING AORTIC STENT PLACEMENT. THE PATIENT'S CONDITION GRADUALLY IMPROVED AND THEY WERE DISCHARGED AFTER RECOVERY. THE DEVICE WAS REPLACED. THERE WAS NO FURTHER ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Aortic dissection; Vascular Dissection
MAUDE report 24543641 ↗InjuryFeb 26, 2026Mechanical Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL CANNULA, IT WAS REPORTED THAT AORTIC PERFUSION WAS PERFORMED VIA THE FEMORAL ARTERY CANNULA, AND AORTIC DISSECTION OCCURRED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO FURTHER ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. IT WAS ALSO REPORTED THAT THE EVENT MAY BE RELATED TO THE UNIQUE VASCULAR CONDITIONS IN THIS PARTICULAR CASE AND THE CANNULATION TECHNIQUE. NO QUALITY DEFECTS HAD BEEN IDENTIFIED.
Patient problems: Aortic dissection; Vascular Dissection
MAUDE report 24451655 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K153598. One submission often covers many models, so check the product and lot details in each record.
- Z-2739-2026 · May 25, 2026 · Open, ClassifiedMedtronic Bio-Medicus Adult Cannula Kit, Product Number/CFN 96530-115; Cardiopulmonary bypass vascular cannula
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page