Skip to content

InPen MMT-8061Recalls & reported problems

Medtronic Minimed, Inc. · SW MMT-8061 INPEN 2.0 ANDROID APP US · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K242775); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Nov 13, 2025Open, Classified

    When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

    Z-0958-2026 · Root cause: Software Design Change · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K242775. One submission often covers many models, so check the product and lot details in each record.

  • Z-2496-2025 · Jun 16, 2025 · Open, Classified
    InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)
  • Z-1650-2025 · Mar 4, 2025 · Open, Classified
    InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA

Other Calculator, drug dose devices

All Calculator, drug dose devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page