inpen recallZ-0958-2026 · Medtronic MiniMed, Inc. · initiated Nov 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
FDA root cause category: Software Design Change
What the company is doing
On 11/13/2025, correction notices were emailed to customers who were asked to update to version 8.0.1 or higher as soon as possible to resolve this issue. 1.Verify your app version: -On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. 2. If you are running versions 8.0.0, update to the latest version 8.0.1 or higher via the Google Play Store: - Open the Google Play Store. - Tap your profile icon at the top of the screen, then select Manage apps & device. - Tap Updates available or search for the InPen app. - Tap Update next to the InPen app icon. Electronically acknowledge that you have read and understand this notification and have followed the actions listed For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option #1. The Customer Confirmation Form can also be emailed to the firm at [email protected]
Recall details
- Recall number
- Z-0958-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 13, 2025
- Posted by the FDA
- Dec 18, 2025
- Recalling firm
- Medtronic MiniMed, Inc.
- Quantity
- 8251
- Product code
- NDC
- FDA submissions
- K242775
- Distribution
- US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
Product description
InPen App, Model/CFN Number: MMT-8061 (Android Users)
Code information (models, lots, serial numbers)
Software Versions: 7.5.0, 7.5.1, and 8.0.0. UDI-DI: 00763000974596
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- InPen MMT-8061Medtronic Minimed, Inc. · Calculator, drug dose1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page