Skip to content

SUR-FIT NATURA 125271Recalls & reported problems

Convatec, Purchasing Department · SUR-FIT Natura® Stomahesive® Flexible Pre-Cut 4" X 4" Wafer Tan 1 Stoma, Flange Size 1 3/4 · Class 1

All models of SUR-FIT NATURA (43)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00768455101849), received between Dec 2, 2020 and Jan 23, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Malposition of Device” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device8
  • Sharp Edges1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient1
  • Skin Inflammation/ Irritation1
  • Urticaria1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2025Malposition of Device; Malposition of Device

    THE CONSUMER REPORTED THAT FIVE WAFERS FROM ONE MARKET UNIT (MU) HAD OFF CENTERED STARTER HOLE ISSUE. THE CONSUMER ALSO STATED THAT THE OFF CENTERED HOLES WERE IN VARYING DEGREES LIKE SOME WERE MILD AND SOME WERE MORE SEVERE. THE END USER ALSO REPORTED THAT SHE WAS ABLE TO WEAR SOME OF THE AFFECTED WAFERS, BUT WEAR TIME DECLINED FROM TWO DAYS TO LESS THAN ONE DAY. SHE STATED THAT ONE OF THE AFFECTED WAFERS WERE SO BAD THAT SHE WAS UNABLE TO ATTACH IT TO THE COMPANY¿S KNOWN POUCH. ALSO, MENTIONED THAT ONE DAY SHE HAD TO CHANGE THE PRODUCT THREE TIMES DUE TO THE OFF CENTERED HOLES. SHE STATED THAT EMOTIONALLY SHE WAS VERY UPSET AND HAD LOST FAITH IN THE PRODUCT. THE PRODUCT WAS USED BY PATIENT. NO PHOTO WAS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21215766 ↗
  • MalfunctionJan 23, 2025Malposition of Device; Malposition of Device

    TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11 (ADDL MFG NARRATIVE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21215764 ↗
  • MalfunctionJan 23, 2025Malposition of Device; Malposition of Device

    THE CONSUMER REPORTED THAT FIVE WAFERS FROM ONE MARKET UNIT (MU) HAD OFF CENTERED STARTER HOLE ISSUE. THE CONSUMER ALSO STATED THAT THE OFF CENTERED HOLES WERE IN VARYING DEGREES LIKE SOME WERE MILD AND SOME WERE MORE SEVERE. THE END USER ALSO REPORTED THAT SHE WAS ABLE TO WEAR SOME OF THE AFFECTED WAFERS, BUT WEAR TIME DECLINED FROM TWO DAYS TO LESS THAN ONE DAY. SHE STATED THAT ONE OF THE AFFECTED WAFERS WERE SO BAD THAT SHE WAS UNABLE TO ATTACH IT TO THE COMPANY¿S KNOWN POUCH. ALSO, MENTIONED THAT ONE DAY SHE HAD TO CHANGE THE PRODUCT THREE TIMES DUE TO THE OFF CENTERED HOLES. SHE STATED THAT EMOTIONALLY SHE WAS VERY UPSET AND HAD LOST FAITH IN THE PRODUCT. THE PRODUCT WAS USED BY PATIENT. NO PHOTO WAS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21215763 ↗
  • MalfunctionJan 23, 2025Malposition of Device; Malposition of Device

    TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11 (ADDL MFG NARRATIVE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21215762 ↗
  • MalfunctionJan 23, 2025Malposition of Device; Malposition of Device

    TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11 (ADDL MFG NARRATIVE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21215761 ↗
  • MalfunctionNov 5, 2024Sharp Edges; Sharp Edges

    THE END USER REPORTED THAT HE EXPERIENCED RED BUMPS ALONG PERIPHERY OF TAPE COLLAR EDGE. HE NORMALLY USED OTHER COMPANY¿S PRODUCTS AND RECENTLY SWITCHED TO OUR COMPANY¿S PRODUCT DUE TO POUCH OPTION PREFERENCE. HE HAD NOT PREVIOUSLY TRIALED OUR COMPANY¿S PRODUCT. HE STATED HE DEVELOPED RED BUMPS ALONG THE PERIPHERY OF THE TAPE COLLAR. HE ALSO, STATED THAT THE TAPE EDGES FELT ROUGH. THERE WAS NO IRRITATION UNDER THE TAPE OR MASS. HE APPLIED THE WAFER IN A SEMI-RECLINED POSITION BEFORE SITTING UP. HE USED THE COMPANY¿S BARRIER WIPES PRIOR TO APPLICATION AND ALLOWED IT TO DRY. AFTER REMOVING THE SKIN BARRIER, HE APPLIED HIS USUAL OTHER COMPANY¿S PRODUCT, AND THE BUMPS WERE RESOLVED BY THE NEXT CHANGE THREE TO FOUR DAYS LATER. NO PHOTOGRAPH WAS AVAILABLE AT THIS TIME.

    Patient problems: Skin Inflammation/ Irritation; Urticaria

    MAUDE report 20607922 ↗
  • MalfunctionFeb 20, 2024Malposition of Device; Malposition of Device

    IT WAS REPORTED THAT STARTER HOLE OF PRE-CUT WAFER WAS OVAL. THE PRODUCT WAS NOT USED. NO PHOTO WAS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18737402 ↗
  • MalfunctionFeb 20, 2024Malposition of Device; Malposition of Device

    IT WAS REPORTED THAT STARTER HOLE OF PRE-CUT WAFER WAS OVAL. THE PRODUCT WAS NOT USED. NO PHOTO WAS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18737401 ↗
  • MalfunctionDec 2, 2020Malposition of Device; Malposition of Device

    TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10933970 ↗

Other Protector, Ostomy devices

All Protector, Ostomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page