NATURA 411803Recalls & reported problems

Convatec, Purchasing Department · SUR-FIT Natura® Moldable Stomahesive® Skin Barrier With Hydrocolloid Flexible Collar [45Mm (1 3/4" ) - Medium] · Class 1

All models of NATURA (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Convatec SUR-FIT Natura skin barrier is a moldable wafer worn around an ostomy opening. Of the 161 reports, all are malfunctions, and most describe a starter hole that is off-center, making some wafers unusable. The reports include no deaths or injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
161
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 161 adverse event reports that list this exact device (DI 00768455105977), received between Apr 10, 2019 and Jun 15, 2026 — classified as 161 malfunctions.

The most frequently reported specific device problem is “Malposition of Device” (157 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 10 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 23 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 8 reportsJun 2026: 16 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device157
  • Positioning Problem14
  • Material Too Rigid or Stiff3
  • Material Disintegration2
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient80
  • No Clinical Signs, Symptoms or Conditions64
  • No Patient Involvement17

Recent adverse event reports

The 25 most recent of 161 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505917 ↗
  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505915 ↗
  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505914 ↗
  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505913 ↗
  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505910 ↗
  • MalfunctionJun 15, 2026Malposition of Device

    THE CONSUMER REPORTED THAT SIX UNITS WERE UNUSABLE DUE TO AN OFF-CENTER STARTER HOLE, WHILE FOUR UNITS FROM THE SAME BOX WERE USABLE WITH MINOR DEVIATION. WHEN PROPERLY CENTERED, THE DEVICE ACHIEVED EXPECTED PERFORMANCE (UP TO 5 DAYS WEAR TIME). THE PRODUCT WAS NOT USED. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505908 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389332 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389331 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389330 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389329 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389327 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389322 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389238 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389237 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389234 ↗
  • MalfunctionJun 4, 2026Malposition of Device

    THE CUSTOMER REPORTED THAT THE PRODUCT OPENINGS IN THE BOX WERE OFF-CENTERED. HE STATED THAT ALL TEN WAFERS IN THE BOX WERE AFFECTED AND NOTED THAT THIS BOX HAD BEEN PROVIDED AS A REPLACEMENT FOR THE SAME ISSUE REPORTED IN A PREVIOUS COMPLAINT. THE CUSTOMER INDICATED THAT THE OPENINGS IN ALL WAFERS WERE MISALIGNED, WITH SOME MORE NOTICEABLY OFF-CENTER THAN OTHERS, RENDERING SEVERAL WAFERS COMPLETELY UNUSABLE FOR HIM. HE CONFIRMED THAT HE HAD NOT USED ANY WAFERS FROM THIS BOX. HE FURTHER EXPLAINED THAT WHEN HE HAD PREVIOUSLY USED PRODUCTS WITH THIS ISSUE, HE HAD BEEN UNABLE TO ACHIEVE A PROPER SEAL. HE EMPHASIZED THAT AN EVENLY CENTERED OPENING WAS CRUCIAL FOR HIM DUE TO THE PRESENCE OF A PARASTOMAL HERNIA. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389228 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240886 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240885 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240884 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240883 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240882 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240881 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240880 ↗
  • MalfunctionMay 21, 2026Malposition of Device

    THE END USER REPORTED THAT THE OPENING IN THE EIGHT MOLDABLE SKIN BARRIERS WERE OFF CENTERED WHICH MADE THE SKIN BARRIER UNUSABLE. CONSUMER WAS A LONG-TIME USER OF THIS PRODUCT; TYPICAL WEAR TIME WAS FIVE DAYS. A PHOTOGRAPH DEPICTING THE ISSUE WAS RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25240879 ↗
  • MalfunctionFeb 7, 2025Malposition of Device; Malposition of Device

    THE DISTRIBUTOR REPORTED THAT THE EDGE OF THE CENTRAL HOLE OF MOLDABLE WAFERS WAS NOT REGULAR, AND THE GLUE OVERFLOWED. THE PRODUCT WAS NOT USED BY THE END USER. THE PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21332703 ↗

FDA clearance and classification

The FDA regulates product code EXE (protector, ostomy) under 21 CFR 876.5900 as a Class I device. 21 CFR 876.5900 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NATURA 411803

Has the NATURA 411803 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00768455105977), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NATURA 411803?

The FDA's MAUDE database lists 161 adverse event reports for this device, received between Apr 10, 2019 and Jun 15, 2026. In those reports, the most common specific device problems are “Malposition of Device” (157), “Positioning Problem” (14) and “Material Too Rigid or Stiff” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NATURA 411803?

None of the reports are classified as deaths, 0 as injuries and 161 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NATURA 411803 FDA approved?

Devices of this type (product code EXE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Convatec, Purchasing Department.

Who makes the NATURA 411803, and is it still sold?

The FDA's device database lists Convatec, Purchasing Department as the company responsible for it (brand NATURA, model 411803). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.