Skip to content

Bardex® I.C. Foley Catheter 0165SI16Recalls & reported problems

C. R. Bard, Inc. · Bardex® I.C. 5cc/mL Balloon Foley Catheter · Class 2

All models of Bardex I.C. Foley Catheter (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Bardex I.C. Foley catheter is the subject of 1,031 adverse event reports, which mention no deaths and 90 injuries. Many reports describe balloons that will not deflate or inflate, leaks of fluid or urine, holes in the material, and catheters that are hard to remove. Reports fell from 147 in the previous 12 months to 25 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,052
Deaths reported
0
Injuries reported
93
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,052 adverse event reports that list this exact device (DI 00801741016752), received between Mar 29, 2017 and Sep 30, 2026 — classified as 93 injuries, 959 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (169 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 15 reportsJan 2024: 7 reportsFeb 2024: 13 reportsMar 2024: 17 reportsApr 2024: 10 reportsMay 2024: 8 reportsJun 2024: 6 reportsJul 2024: 8 reportsAug 2024: 6 reportsSep 2024: 12 reportsOct 2024: 13 reportsNov 2024: 6 reportsDec 2024: 13 reportsJan 2025: 12 reportsFeb 2025: 8 reportsMar 2025: 14 reportsApr 2025: 18 reportsMay 2025: 9 reportsJun 2025: 22 reportsJul 2025: 12 reportsAug 2025: 5 reportsSep 2025: 11 reportsOct 2025: 17 reportsNov 2025: 8 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 3 reportsJul 2026: 3 reportsAug 2026: 5 reportsSep 2026: 14 reportsOct 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem169
  • Failure to Infuse146
  • Material Puncture/Hole134
  • Fluid/Blood Leak127
  • Decrease in Pressure98
  • Difficult to Remove95
  • Burst Container or Vessel91
  • Material Rupture71

Patient problems

  • No Clinical Signs, Symptoms or Conditions674
  • No Known Impact Or Consequence To Patient240
  • No Consequences Or Impact To Patient205
  • No Patient Involvement68
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available45
  • Patient Problem/Medical Problem32
  • Foreign Body In Patient29
  • No Code Available13

Recent adverse event reports

The 25 most recent of 1,052 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952605 ↗
  • MalfunctionSep 30, 2026Disconnection

    It was reported that the Foley catheter was leaking and discovered that the collection port was missing. The detached collection port was found on the patient absorbent pad.As per customer follow up information received on 08Sep2026, physical samples are available. There was no medical intervention required.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951862 ↗
  • MalfunctionSep 29, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940874 ↗
  • MalfunctionSep 29, 2026Difficult to Remove

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940505 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899366 ↗
  • MalfunctionSep 18, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829597 ↗
  • MalfunctionSep 18, 2026Inflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819870 ↗
  • MalfunctionSep 18, 2026Material Puncture/Hole

    It was reported that during use a balloon catheter was inserted during surgery on (b)(6). It came out spontaneously in the ward on (b)(6). Suspecting a malfunction, the foley catheter balloon was inflated and expanded without issue. There were no abnormalities for about three hours, but six hours later, the balloon had deflated. The area near the tip was wet.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819868 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    It was reported that the Regular catheter was very uncomfortable and got infection. No medical intervention was reported.

    Patient problems: Discomfort; Unspecified Infection

    MAUDE report 50802928 ↗
  • MalfunctionSep 14, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764413 ↗
  • MalfunctionSep 11, 2026Material Puncture/Hole

    It was reported that this was a case of spontaneous dislodgement of the Foley Catheter. On August 31, following placement, while the patient was sitting on the toilet seat in the wheelchair-accessible restroom on September 3, the catheter spontaneously dislodged; therefore, its use was discontinued, and a different device was used for insertion. At the time of dislodgement, the catheter¿s cuff had deflated; although it was inflated with air and expanded, it reportedly continued to deflate even though there was no leakage from the cuff.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735892 ↗
  • MalfunctionSep 2, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50573358 ↗
  • MalfunctionSep 2, 2026Difficult to Remove

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50563774 ↗
  • MalfunctionSep 2, 2026Inflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50563773 ↗
  • MalfunctionAug 28, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50506896 ↗
  • MalfunctionAug 28, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501980 ↗
  • InjuryAug 24, 2026Patient Device Interaction Problem

    It was reported that a catheter was placed on (b)(6) for a patient hospitalized in the (b)(6) Department due to a cerebral infarction. It was temporarily removed on (b)(6), but the patient developed urinary retention on (b)(6), and approximately 700 mL of urine was ejected via catheterization. On (b)(6), a Foley catheter was placed, but the patient complained of abdominal pain. A contrast-enhanced CT scan revealed that the catheter had penetrated the bladder wall by approximately 5 cm and was protruding outside the bladder. The patient has developed sepsis due to a urinary tract infection (UTI), which has led to peritonitis.

    Patient problems: Abdominal Pain; Perforation; Urinary Retention

    MAUDE report 50409215 ↗
  • MalfunctionAug 6, 2026Material Puncture/Hole; Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50166664 ↗
  • MalfunctionAug 4, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50131790 ↗
  • MalfunctionJul 15, 2026Calcified

    IT WAS REPORTED BY CUSTOMER THAT THEY WERE USING LATEX FOLEY 0165SI16 BUT WAS HAVING ISSUES WITH SEDIMENT SO THEY SWITCHED TO A SILICONE FOLEY: 175816. THEY STATE WHEN LOOKING AT THESE PRODUCTS THE EXTERNAL DIAMETERS LOOK DIFFERENT. THE SILICONE LOOKS A GOOD BIT SMALLER THAN THE LATEX FOLEY. VERBAL RESPONSE NOTES: EXPLAINED THAT THE SILICONE FOLEY SHOULD HAVE A LARGER INTERNAL DIAMETER THAN LATEX, BUT NOT THE OUTER DIAMETER. EXPLAINED HOW 16FR IS; 16FR NO MATTER THE SUBSTRATE. LEAKING CAN BE CAUSED BY A FEW ISSUES. SUGGESTED SPEAKING WITH HIS UROLOGIST TO IDENTIFY THE ROOT CAUSE. EXPLAINED HE CAN DISCUSS A DIFFERENT SIZE WITH HIS HCP, BUT DO NOT RECOMMEND CHANGING WITHOUT DISCUSSING FIRST. THEY HAVE STATED THAT THEY WILL DISCUSS THESE ISSUES WITH THEIR DOCTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25833860 ↗
  • MalfunctionJul 14, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25816391 ↗
  • MalfunctionJul 7, 2026Fracture

    It was reported via email that during routine care and flushing of a (SP) Foley catheter, a mechanical issue was identified with the catheter flush valve. Staff noticed that the flush port was cracking more frequently during flushing. Due to the malfunctioning valve, the catheter could not be flushed properly, which resulted in several catheters needing to be replaced. We are able to reposition valve plastic some of the time for proper flushing but have had to replace the foley catheter several times.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25736196 ↗
  • MalfunctionJun 17, 2026Material Rupture

    IT WAS REPORTED THAT FOLEY CATHETER BALLOON WAS DAMAGED DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25540152 ↗
  • InjuryJun 16, 2026Difficult to Remove

    IT WAS REPORTED THAT THEY RECENTLY ENCOUNTERED A CONCERNING SITUATION WHERE A NURSE WAS UNABLE TO PULL OUT STERILE WATER FROM THE FOLEY CATHETER, OR THE CATHETER BALLOON AND THE FOLEY COULD NOT BE REMOVED. THE PATIENT ULTIMATELY HAD TO BE SENT TO THE HOSPITAL FOR REMOVAL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25520745 ↗
  • MalfunctionJun 10, 2026Material Puncture/Hole

    IT WAS REPORTED THAT FOLEY CATHETER FELL OUT ON ITS OWN IMMEDIATELY AFTER BEING INSERTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25462560 ↗

Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices

All CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices →