Skip to content

Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314Recalls & reported problems

C. R. Bard, Inc. · Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Bard Lubri-Sil Foley catheter is a silicone urinary tube. Reports often describe holes in the tube, balloons that do not deflate and fluid leaks. The 3,843 reports include 29 injuries and no deaths. There is 1 open recall, started in June 2025, because the tube may become blocked and stop urine draining from the bladder. Reports fell to 221 in the last 12 months from 1,410 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
3,889
Deaths reported
0
Injuries reported
31
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3,889 adverse event reports that list this exact device (DI 00801741039843), received between Apr 19, 2017 and Sep 30, 2026 — classified as 31 injuries, 3,858 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (1,384 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 64 reportsDec 2023: 46 reportsJan 2024: 19 reportsFeb 2024: 31 reportsMar 2024: 42 reportsApr 2024: 36 reportsMay 2024: 32 reportsJun 2024: 64 reportsJul 2024: 61 reportsAug 2024: 71 reportsSep 2024: 41 reportsOct 2024: 79 reportsNov 2024: 71 reportsDec 2024: 96 reportsJan 2025: 93 reportsFeb 2025: 88 reportsMar 2025: 93 reportsApr 2025: 122 reportsMay 2025: 113 reportsJun 2025: 248 reportsJul 2025: 150 reportsAug 2025: 106 reportsSep 2025: 120 reportsOct 2025: 110 reportsNov 2025: 85 reportsDec 2025: 67 reportsJan 2026: 31 reportsFeb 2026: 11 reportsMar 2026: 4 reportsApr 2026: 2 reportsMay 2026: 4 reportsJun 2026: 17 reportsJul 2026: 17 reportsAug 2026: 15 reportsSep 2026: 14 reportsOct 2026: 0 reports202420252026248
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 18, 2025Class IIOpen, Classified

    Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.

    Z-2462-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Puncture/Hole1,384
  • Deflation Problem640
  • Difficult to Remove528
  • Fluid/Blood Leak495
  • Decrease in Pressure280
  • Component Misassembled193
  • Failure to Infuse185
  • Inflation Problem169

Patient problems

  • No Clinical Signs, Symptoms or Conditions3,330
  • No Known Impact Or Consequence To Patient405
  • No Consequences Or Impact To Patient319
  • No Patient Involvement217
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available56
  • Pain8
  • Discomfort7
  • No Information7

Recent adverse event reports

The 25 most recent of 3,889 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Difficult to Remove

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952607 ↗
  • MalfunctionSep 28, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926184 ↗
  • MalfunctionSep 23, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873312 ↗
  • MalfunctionSep 22, 2026Material Rupture

    It was reported that during use, 2 hours after placement, the foley catheter used in the operating room fell out in the Intensive Care Unit (ICU). There is no indication that any device parts remained in the body after the balloon ruptured.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855631 ↗
  • InjurySep 21, 2026Difficult to Remove

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50848696 ↗
  • MalfunctionSep 18, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829602 ↗
  • MalfunctionSep 18, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829600 ↗
  • MalfunctionSep 18, 2026Calcified

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819877 ↗
  • MalfunctionSep 15, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781993 ↗
  • MalfunctionSep 14, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765282 ↗
  • MalfunctionSep 11, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50741620 ↗
  • MalfunctionSep 10, 2026Difficult to Remove

    It was reported that during placement the foley catheter took a long time to drain fluid when removing the catheter. Was able to drain the entire volume after spending some time (Approximately 15 minutes).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717393 ↗
  • MalfunctionSep 2, 2026Fluid/Blood Leak

    It was reported that used during a procedure in the cardiac catheterization lab. Sterile water leaked from the catheter's valve section.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566241 ↗
  • MalfunctionSep 2, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50562029 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole; Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459571 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456186 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456184 ↗
  • MalfunctionAug 26, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456183 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456182 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole

    It was reported that after being placed in OP, it was discovered that the foley catheter had come out in the ward. Similar cases occur frequently. Natural drainage.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456181 ↗
  • MalfunctionAug 26, 2026Material Puncture/Hole; Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456179 ↗
  • MalfunctionAug 21, 2026Material Puncture/Hole; Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50377858 ↗
  • MalfunctionAug 19, 2026Material Rupture

    It was reported that spontaneous detachment due to leak in the balloon section.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50329555 ↗
  • MalfunctionAug 19, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50319309 ↗
  • MalfunctionAug 17, 2026Material Puncture/Hole; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50280938 ↗

Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices

All CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices →