Bard® Silastic® Foley Catheter 33422Recalls & reported problems
C. R. Bard, Inc. · Bard® Silastic® Foley Catheter 30cc Balloon · Class 2
All models of Bard Silastic Foley Catheter (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741020018), received between Apr 21, 2017 and Jun 8, 2026 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 23, 2026Class IIOpen, Classified
Due to stain present on the surface of affected foley catheters.
Z-2226-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Burst Container or Vessel1
- Collapse1
- Crack1
- Deflation Problem1
- Difficult to Remove1
- Failure to Infuse1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THE FOLEYS HE RECEIVED IN THIS ORDER WERE DAMAGED, CAUSING LEAKAGE. HE ALREADY USED 2 OUT OF THE 3. SENT AN EMAIL TO HAVE CUSTOMER SERVICE CALL BACK. NO ADDITIONAL INFORMATION PROVIDED. IT'S UNCLEAR IF TROUBLESHOOTING WAS PROVIDED. (PREPPING AND PLACEMENT TIPS SHARED)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25437691 ↗InjuryMar 18, 2026Material Rupture
IT WAS REPORTED THAT FOLEY CATHETER BEEN REPORTED AS DEFECTIVE, ACCORDING TO THE PATIENT THE BALLOON BURST INSIDE AND HAS TO BE HOSPITALIZED FOR 5 MONTHS, ANOTHER INSTANCE WAS THE BALLOON WAS LEFT INSIDE. THEY SEEK MEDICAL ATTENTION.
Patient problems: Foreign Body In Patient
MAUDE report 24628320 ↗MalfunctionSep 12, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE PT SAID THAT THE BALLOON WAS NOT EXPANDING. UNCLEAR IF MEDICAL INTERVENTION WAS PROVIDED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038325 ↗MalfunctionJun 30, 2025Collapse; Collapse
IT WAS REPORTED BY THE CUSTOMER THAT THE FOLEY #5733422, LOT#NGJP2382 KEPT COLLAPSING AND THE URINE BACKS UP AND SPILLS OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22358117 ↗MalfunctionAug 11, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE CUSTOMER HAD MULTIPLE ISSUES WITH THE FOLEY CATHETER BALLOON. THE BALLOON HAD BURST INSIDE THE PATIENT SEVERAL TIMES. THE BALLOON ALSO DID NOT BLOW UP ON SOME OCCASIONS. THE BALLOON WAS ALSO LEAKING. THIS HAD HAPPENED WITH ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17510636 ↗MalfunctionAug 11, 2023Failure to Infuse; Failure to Infuse
IT WAS REPORTED THAT THE CUSTOMER HAD MULTIPLE ISSUES WITH THE FOLEY CATHETER BALLOON. THE BALLOON HAD BURST INSIDE THE PATIENT SEVERAL TIMES. THE BALLOON ALSO DID NOT BLOW UP ON SOME OCCASIONS. THE BALLOON WAS ALSO LEAKING. THIS HAD HAPPENED WITH ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17509033 ↗MalfunctionAug 10, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE CUSTOMER HAD MULTIPLE ISSUES WITH THE FOLEY CATHETER BALLOON. THE BALLOON HAD BURST INSIDE THE PATIENT SEVERAL TIMES. THE BALLOON ALSO DID NOT BLOW UP ON SOME OCCASIONS . THE BALLOON WAS ALSO LEAKING. THIS HAD HAPPENED WITH ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17497363 ↗MalfunctionNov 5, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE FOLEY CATHETER HAD URINE LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10793598 ↗MalfunctionOct 26, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PATIENT HAD A BYPASS ISSUE. SUPPOSEDLY THE LUBRICATH WAS SUBSTITUTED FOR THE SAME FRENCH OF BALLOON SIZE AS ORIGINALLY RECEIVED HOWEVER THERE WAS STILL URINATING THROUGH THE BYPASS. THE PATIENT HAD A SIGNIFICANT ENLARGED PROSTATE. IT WAS FURTHER REPORTED THAT THE PATIENT HAD A PROSTATE CANCER AND THE INSERTION ISSUES MIGHT BE RELATED TO THIS. THE MEDICAL INFORMATION LEAD ADDED THAT IN THE PATIENT EXPERIENCE. THE BYPASS WAS COMMONLY RELATED TO BLADDER SPASMS.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10734325 ↗InjuryJan 3, 2020Deflation Problem; Deflation Problem; Difficult to Remove
IT WAS REPORTED THAT THE SILASTIC FOLEY CATHETER WAS DIFFICULT TO REMOVE AND WAS GETTING STUCK IN THE URINARY TRACT. ONE PATIENT REPORTEDLY HAD TO HAVE THE CATHETER REMOVED IN INTERVENTIONAL RADIOLOGY. PER ADDITIONAL INFORMATION RECEIVED FROM THE COMPLAINANT VIA EMAIL ON 10JAN2020, IT WAS NOTE THAT THE PATIENT REQUIRED CT GUIDED PUNCTURE OF THE BALLOON THROUGH THE ABDOMEN.
Patient problems: Device Embedded In Tissue or Plaque; Patient Problem/Medical Problem
MAUDE report 9544562 ↗MalfunctionApr 21, 2017Crack
IT WAS REPORTED THAT THE RUBBER WAS CRACKING AROUND THE MALE END THAT CONNECTS TO THE URINARY BAG.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6512251 ↗
FDA clearance and classification
The FDA regulates product code EZL (catheter, retention type, balloon) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Bard® Silastic® Foley Catheter 33422
Has the Bard® Silastic® Foley Catheter 33422 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2226-2026 (Class II), began on Apr 23, 2026: Due to stain present on the surface of affected foley catheters. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Bard® Silastic® Foley Catheter 33422?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Apr 21, 2017 and Jun 8, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (3), “Burst Container or Vessel” (1) and “Collapse” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (1), “Foreign Body In Patient” (1) and “Patient Problem/Medical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Bard® Silastic® Foley Catheter 33422?
None of the reports are classified as deaths, 2 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Bard® Silastic® Foley Catheter 33422, and is it still sold?
The FDA's device database lists C. R. Bard, Inc. as the company responsible for it (brand Bard® Silastic® Foley Catheter, model 33422). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON487 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death381 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON235 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON168 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON152 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page