Bard® Silastic® Foley Catheter 33616Recalls & reported problems
C. R. Bard, Inc. · Bard® Silastic® Foley Catheter 5cc Balloon · Class 2
All models of Bard Silastic Foley Catheter (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard Silastic Foley catheter is a soft silicone urinary catheter with a small balloon that holds it in place. The 30 reports mostly describe drainage problems: urine not flowing, balloon port or tubing coming apart, balloon not deflating so it had to be removed surgically, catheter collapsing or clogging soon after insertion. One injury was reported and no deaths. One open recall, started April 2026, covers staining on the surface of affected catheters due to material contamination.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00801741020063), received between Oct 23, 2017 and Jan 26, 2026 — classified as 1 injury, 29 malfunctions.
The most frequently reported specific device problem is “Component Missing” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 23, 2026Class IIOpen, Classified
Due to stain present on the surface of affected foley catheters.
Z-2226-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Component Missing4
- Component Misassembled3
- Disconnection3
- Partial Blockage3
- Burst Container or Vessel2
- Deflation Problem2
- Difficult to Remove2
- Inaccurate Flow Rate2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient5
- No Known Impact Or Consequence To Patient5
- Hemorrhage/Blood Loss/Bleeding1
- No Information1
- No Patient Involvement1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2026Inaccurate Flow Rate
IT WAS REPORTED BY A CARE PROVIDER THAT THE 33616 HAS A DEFECTIVE STRUCTURAL INTEGRITY. NOT ALLOWING URINE DRAINAGE WHEN INSERTED. PREVIOUS FOLEY CATHETERS DID NOT CAUSE ANY ISSUE WITH DRAINING WHILE INSERTED. FOLEY CATHETER HAS BEEN REPOSITIONED AND ATTEMPTED TO BE ADJUSTED. NO RESOLVE. FOLEY DEFECTS ARE CAUSING PATIENT TO USE SUPPLIES SOONER. SENDING EXCHANGED ORDER. APOLOGIZED FOR DISCOMFORT EXPERIENCE. IT'S UNCLEAR IF LOT NUMBER WAS REQUESTED. UNCLEAR IF MEDICAL INTERVENTION WAS PROVIDED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24185156 ↗MalfunctionDec 12, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED BY THE CUSTOMER STATED THE CATHETER 33616 (B)(6) 2025 - THE BALLOON WOULD NOT DEFLATE. THE DOCTOR HAD TO CLOSE OFF THE LINE AND LET THE BALLOON DEFLATE FOR 2 HOURS. AND THE CATHETER 33618 (B)(6) 2025 ¿ THE BALLON WOULD NOT DEFLATE. SO, THE DOCTOR USED A SURGICAL INSTRUMENT TO REMOVE THE CATHETER WITH THE BUBBLE INFLATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23790924 ↗MalfunctionDec 10, 2025Disconnection; Disconnection
IT WAS REPORTED THAT AFTER INSERTION OF NEW 16FR SILASTIC FOLEY CATHETER 10CC SYRINGE FILLED WITH STERILE WATER ATTACHED TO RETENTION BALLOON PORT. AS BALLOON HAS BEEN ATTACHED AND INFLATED THE ENTIRE BALLOON PORT CAME OFF CATHETER TUBING. THAT MALFUNCTION OCCURRED WITH 2 DIFFERENT CATHETERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771544 ↗MalfunctionDec 10, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE INDWELLING FOLEY CATHETER MALFUNCTION; AFTER INSERTION OF NEW 16FR SILASTIC FOLEY CATHETER. 10CC SYRINGE FILLED WITH STERILE WATER ATTACH TO RETENTION BALLOON PORT. AS BALLOON WAS BEING ATTACHED AND INFLATED THE ENTIRE BALLOON PORT CAME OFF CATHETER TUBING. THIS MALFUNCTION OCCURRED WITH 2 DIFFERENT CATHETERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771173 ↗MalfunctionNov 24, 2025Inflation Problem; Inflation Problem
PATIENT ADVISED THE BALLOON IS NOT INFLATING AND THE RED PRODUCT MARKER IS COMING OFF 7 CATHETERS. PATIENT DID NOT PURCHASE THROUGH LMS. IT IS UNCLEAR IF TROUBLESHOOTING WAS PERFORMED. NO MEDICAL INTERVENTION OR PATIENT IMPACT REPORTED. NO ADDITIONAL INFORMATION PROVIDED. PER CUSTOMER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 12NOV2025, YES, THEY HAVE RETAINED ONE OF THE CATHETERS THAT EXHIBITED THE ISSUE, ALONG WITH ITS ORIGINAL OUTER PACKAGING. THEY WOULD BE GLAD TO SEND IT FOR INVESTIGATION. THE LOT AND SERIAL NUMBERS WERE PRESENT ON THE OUTER PACKAGING WHICH THEY HAVE SAVED. THEY COULD PROVIDE THESE DETAILS, OR THEY WOULD BE AVAILABLE ON THE PHYSICAL SAMPLE AND PACKAGING IF THEY PREFER TO HAVE THEM SHIPPED TO THEM. THERE HAS BEEN A SIGNIFICANT AND DISTRESSING IMPACT ON THE PATIENT. THIS WAS, IN FACT, THE SECOND TIME THEY HAVE ENCOUNTERED SEVERE ISSUES WITH THESE CATHETERS. CUSTOMER WAS EXPERIENCING VISIBLE SWELLING AND IRRITATION IN THE AREA. EACH TIME THEY HAD TO CHANGE A FAULTY CATHETER, THEY SCREAM DUE TO THE PAIN. ON ONE PARTICULAR DAY, THEY HAD TO CHANGE THE CATHETER THREE TIMES BECAUSE THE BALLOON WOULD NOT INFLATE ONCE IT WAS PROPERLY INSERTED. OTHER CATHETERS FROM THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23632138 ↗MalfunctionNov 7, 2025Inaccurate Flow Rate; Inaccurate Flow Rate
IT WAS REPORTED THAT URINARY FOLEY CATHETER INSERTED ACCORDING TO PROPER TECHNIQUE WITH 10 ML IN THE BALLOON NO DRAINAGE, IRRIGATION POSSIBLE. CATHETER REMOVED. SECOND CATHETER INSERTED NO DRAINAGE IRRIGATION POSSIBLE; WHEN STAFF ATTEMPT TO REMOVE THE WATER FROM THE BALLOON, CATHETER COLLAPSES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23498374 ↗MalfunctionNov 7, 2025Collapse; Collapse
IT WAS REPORTED THAT URINARY FOLEY CATHETER INSERTED ACCORDING TO PROPER TECHNIQUE WITH 10 ML IN THE BALLOON NO DRAINAGE, IRRIGATION POSSIBLE. CATHETER REMOVED. SECOND CATHETER INSERTED NO DRAINAGE IRRIGATION POSSIBLE; WHEN STAFF ATTEMPT TO REMOVE THE WATER FROM THE BALLOON, CATHETER COLLAPSES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23498373 ↗MalfunctionNov 3, 2025Complete Blockage; Complete Blockage
PATIENT SAID EXPERIENCED LEAKS AND THE FOLEY CLOGGED AND STOPPED DRAINING SOON AFTER PUTTING IT ON. NO MEDICAL INTERVENTION OR PATIENT IMPACT REPORTED. NO ADDITIONAL INFORMATION PROVIDED. PER ADDITIONAL INFORMATION RECEIVED VIA PHONE ON 30DEC2025, IT WAS REPORTED THAT THE CATHETER SEEMED TO BE DEFECTIVE BECAUSE IT WAS CLOGGED AND WOULD NOT ALLOW URINE TO PASS. A REPLACEMENT WAS SENT TO THE CUSTOMER, AND THE DEFECTIVE CATHETER WAS DISCARDED. THEREFORE, NO SAMPLE IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23453179 ↗MalfunctionOct 6, 2025Component Missing; Component Missing
IT WAS REPORTED THAT 1 PIECE OF FOLEY CATHETER WITH EMPTY PACKAGING CODE 33616 LOT NGJW0884 WAS DETECTED. WHEN TRACKING, THERE WAS AN ENTRY OF 29MAY2025 OF THE MENTIONED LOT, ENTRY VOUCHER 5001015684, ORDER (B)(4). THEREFORE, FOLIO RIP-144-2025 WAS GENERATED AND 1 PIECE WAS BLOCKED IN THE SET. ADDITIONAL INFORMATION RECEIVED ON 25 SEP 2025. THE DEFECT WAS DETECTED ON (B)(6) 2025. THE SAMPLE IS AVAILABLE FOR RETURN AND ANALYSIS. THE PACKAGING HAS NO LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23222167 ↗MalfunctionMar 18, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE FOLEY CATHETER HAD POPPED OUT OF THE PATIENT'S BODY AND WAS INSERTED BY A PHYSICIAN. THEY SAID THAT THE BALLOON POPPED AND WAS NEEDING A REPLACEMENT FOLEY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21632623 ↗MalfunctionDec 27, 2024Disconnection; Disconnection
IT WAS REPORTED THAT THE FOLEY CATHETER HUB DISCONNECTED AND REMAINED ATTACHED TO THE SYRINGE WHEN BALLOON INFLATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21026679 ↗MalfunctionNov 27, 2024Component Missing; Component Missing
IT WAS REPORTED THAT CUSTOMER RECEIVED A BOX AND ONE MISSING FOLEY CATHETER JUST AN EMPTY PACKAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20796251 ↗MalfunctionSep 4, 2024Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE PROBE DID NOT HAVE THE ORANGE FILLING VALVE, WHERE THE SYRINGE WAS CONNECTED TO INSUFFLATE THE FOLEY CATHETER BALLOON, THE PROBE WAS NOT USED IN PATIENT, IT WAS NOT OPENED, THE TRANSPARENT PACKAGING ALLOWED TO SEE THAT THE PROBE WAS MISSING AN ACCESSORY. PER FOLLOW UP INFORMATION RECEIVED VIA IBC ON 23AUG2024, CUSTOMER CONFIRMED THAT THE INFLATION VALVE WAS MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20137523 ↗MalfunctionJul 2, 2024Device Damaged Prior to Receipt by User; Component Missing; Device Damaged Prior to Receipt by User; Component Missing
IT WAS REPORTED THAT WHEN THE CUSTOMER OPENED THE BOX OF 33616 CATHETERS, THERE WERE ONLY 9 CATHETERS INSIDE AND 1 WAS MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19655350 ↗MalfunctionFeb 1, 2024Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A DEFECTIVE FOLEY CATHETER WHICH WAS CLOGGED AND NEEDED TO FLUSH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18621503 ↗MalfunctionNov 7, 2023Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE CUSTOMER HAD RECEIVED A DEFECTIVE ITEM. THE ISSUE WAS THAT THE CATHETER WAS GETTING CLOGGED AND THEY NEEDED TO FLUSH. AS PER FOLLOW-UP INFORMATION RECEIVED FROM CUSTOMER ON 03NOV2023, STATED THAT THEY HAD AN ISSUE WITH THE LEG BAG AND ISSUE WAS LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18083856 ↗MalfunctionJul 25, 2023Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THERE WAS NO ORANGE PART IN THE BEGINNING OF THE FOLEY CATHETER. THAT IS, THE PROBE WAS MISSING A PIECE. THE PROBE WAS NOT USED IN PATIENT, THE PROBE WAS INSIDE THE ORIGINAL CLOSED PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17389396 ↗InjuryDec 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THESE SILASTIC FOLEY CATHETERS CAUSED INFECTIONS TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED. IT WAS UNKNOWN IF THE DEVICE CONTRIBUTED TO THE INFECTION AT THIS TIME AND MEDICAL INTERVENTION WAS UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 16062171 ↗MalfunctionNov 16, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT BALLOONS OF ALL THE 3 FOLEY CATHETERS EXPLODED WITHIN AN HOUR. PER FOLLOW UP VIA PHONE ON (B)(6) 2022, CUSTOMER STATED THAT THEY WENT THROUGH 3 FOLEY CATHETERS WITHIN AN HOUR DUE TO EACH ONE BURSTING AFTER A FEW MINUTES OF USE. STATED THAT THEY INSERTED THE CATHETER AND FILLED WITH 10CC THEN WENT TO SIT TO WATCH TELEVISION AND WITHIN A FEW MINUTES, THEY NOTICED THE BURST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15810570 ↗MalfunctionMay 19, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE NURSE COULD NOT GET THE FOLEY CATHETERS IN AND HAD TO USE A FEW. STATED THAT THE CATHETERS HAD KINKS IN THEM WHICH WAS CAUSED THE PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14433968 ↗MalfunctionApr 6, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE NURSE WENT TO CHANGE A PATIENT¿S CATHETER AND THERE WERE NO ISSUES. A FEW HOURS LATER, THE PATIENT FELT A SPRING AND SNAP IN THEIR BLADDER, BUT THE CATHETER SEEMED TO BE DRAINING STILL. THEN SHORTLY LATER THE PATIENT REPORTED WETNESS AND CATHETER BEGAN LEAKING. NURSE RETURNED TO THE PATIENT AND DISCOVERED THAT THE CATHETER HAD BURST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14021417 ↗MalfunctionMay 13, 2021Component Misassembled; Missing Information; Component Misassembled; Missing Information
IT WAS REPORTED THAT A PIECE OF THE FOLEY CATHETER ARRIVED INCOMPLETE BECAUSE IT LACKED THE IDENTIFICATION COLOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11820057 ↗MalfunctionMay 28, 2020Failure to Infuse; Inflation Problem; Failure to Infuse
IT WAS REPORTED THAT THE CATHETER BALLOON WOULD NOT INFLATE. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 10095909 ↗MalfunctionMar 12, 2020Material Rupture; Component Missing
IT WAS REPORTED THAT THE BALLOON OF THE CATHETER ALLEGEDLY RUPTURED DURING PATIENT USE. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 27 FEB 2020 FROM IBC REPRESENTATIVE, A PIECE OF THE BALLOON WAS MISSING. THE PIECE COULD NOT BE FOUND. THERE WAS NO REPORTED PATIENT INJURY. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 2 MARCH 2020 FROM IBC REPRESENTATIVE, THERE WAS NO PROCEDURE PERFORMED TO LOCATE THE MISSING PIECE. THERE IS NO INJURY TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9824571 ↗MalfunctionJan 3, 2020Difficult to Remove; Physical Resistance/Sticking
IT WAS REPORTED THAT THE SILASTIC FOLEY CATHETER WAS DIFFICULT TO REMOVE AND WAS GETTING STUCK IN THE URINARY TRACT. THE COMPLAINANT REPORTED THAT THE CATHETER WAS CUT IN ORDER FOR THE BALLOON TO BE PROPERLY DEFLATED, AND THE CATHETER WAS REMOVED FROM THE PATIENT. THE PATIENT REPORTEDLY EXPERIENCED PAIN. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient; Pain
MAUDE report 9544569 ↗
FDA clearance and classification
The FDA regulates product code EZL (catheter, retention type, balloon) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Bard® Silastic® Foley Catheter 33616
Has the Bard® Silastic® Foley Catheter 33616 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2226-2026 (Class II), began on Apr 23, 2026: Due to stain present on the surface of affected foley catheters. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Bard® Silastic® Foley Catheter 33616?
The FDA's MAUDE database lists 30 adverse event reports for this device, received between Oct 23, 2017 and Jan 26, 2026. In those reports, the most common specific device problems are “Component Missing” (4), “Component Misassembled” (3) and “Disconnection” (3); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (1), “Pain” (1) and “Unspecified Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Bard® Silastic® Foley Catheter 33616?
None of the reports are classified as deaths, 1 as an injury and 29 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Bard® Silastic® Foley Catheter 33616, and is it still sold?
The FDA's device database lists C. R. Bard, Inc. as the company responsible for it (brand Bard® Silastic® Foley Catheter, model 33616). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON487 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death381 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON235 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON168 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON152 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page