Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316Recalls & reported problems
C. R. Bard, Inc. · Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741024931), received between Oct 4, 2021 and Jul 23, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Invagination” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 25, 2026Open, Classified
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
Z-3115-2026 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Material Invagination2
- Burst Container or Vessel1
- Gel Leak1
- Material Fragmentation1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT SYRINGE IN FOLEY TRAY WAS SMASHED (PCN#947316). DRAIN BAG OPENING WHERE THE TIP FOR EMPTYING SITS WAS ALL SMASHED WHICH ALLOWED URINE TO LEAK ALL OVER THE OPERATING ROOM FLOOR (PCN#154002). FIRST ONE THE FOLEY CATHETER BALLOON POPPED (PCN#175816) AND THE SECOND TRY, THE BALLOON LEAKED (PCN#175816). PER FOLLOW UP RESPONSE VIA EMAIL ON 16JUL2024, CUSTOMER STATED THAT THEY WERE ALL THE SAME LOT#NGHZ3646. NO ADDITIONAL PATIENT IMPACT FROM THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19805825 ↗MalfunctionNov 17, 2023Gel Leak; Nonstandard Device; Gel Leak; Nonstandard Device
IT WAS REPORTED THAT THE INSIDE THIS PRODUCT COMES AN ACCESSORY, CALLED URINARY CATHETER LUBRICATION GEL. THIS ACCESSORY WAS CHEMICALLY CHANGED, THAT IS, THE CLIENT INDICATES THAT IT CAME DRY INSIDE THE PACKAGING, FOR THIS REASON IT COULD NOT BE USED, THAT IS, THIS LUBRICANT WAS USED SO THAT THE URINARY CATHETER PASSES THROUGH THE URINARY MEATUS WITHOUT RESISTANCE AND MORE SOFT. IT WAS STATED THAT NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18159736 ↗MalfunctionMar 24, 2022Material Invagination; Material Invagination
IT WAS REPORTED THAT UPON DEFLATING THE BALLOON, IF THERE WAS ANY PRESSURE BETWEEN THE BALLOON AND THE BLADDER WALL, A RIDGE GETS FORMED ON THE SILICONE CATHETER PREVENTING THE CATHETER FROM BEING REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13897184 ↗MalfunctionNov 18, 2021Material Invagination; Material Invagination
IT WAS REPORTED THAT UPON REMOVING THE FOLEY CATHETER, THE BALLOON WAS DEFLATED BUT IF THERE WAS ANY PRESSURE AGAINST THE BALLOON DURING THIS STEP, A FOLD WILL OCCUR IN THE SILICONE CIRCUMFERENTIALLY. THIS PREVENTED THE FOLEY CATHETER FROM BEING REMOVED. THIS HAD HAPPENED MULTIPLE TIMES WITH BOTH THE TRAY PRODUCT NUMBER 897216 AND WITH TRAY PRODUCT NUMBER 947316. AND ONE PATIENT NEEDED TO GO TO HAVE A PROCEDURE TO REMOVE THE CATHETER. THE CUSTOMER STATED THAT THEY HAVE A LARGE NUMBER OF THESE FOLEY TRAYS THAT WERE NOT WORKING FOR THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12836629 ↗MalfunctionOct 4, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT TWO SILICONE FOLEY CATHETER BALLOON WERE POPPED. PER FOLLOW UP RECEIVED VIA EMAIL ON (B)(6) 2021, CUSTOMER STATED THAT THE BALLOON WAS NOT INFLATED PRIOR TO INSERTION ON ANY OF THE FOLEY CATHETERS AND THE INCIDENT HAPPENED AT THE TIME OF INSERTION WHEN THE NURSE INJECTED 10CC OF FLUID INTO THE BALLOON. THE NURSE DISCARDED BOTH FAILED FOLEYS AS THEY WERE INSIDE THE PATIENT AT THE TIME OF FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12568172 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.