1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741024931), received between Oct 4, 2021 and Jul 23, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Invagination” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 25, 2026Class IIOpen, Classified

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

    Z-3115-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Invagination2
  • Burst Container or Vessel1
  • Gel Leak1
  • Material Fragmentation1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT SYRINGE IN FOLEY TRAY WAS SMASHED (PCN#947316). DRAIN BAG OPENING WHERE THE TIP FOR EMPTYING SITS WAS ALL SMASHED WHICH ALLOWED URINE TO LEAK ALL OVER THE OPERATING ROOM FLOOR (PCN#154002). FIRST ONE THE FOLEY CATHETER BALLOON POPPED (PCN#175816) AND THE SECOND TRY, THE BALLOON LEAKED (PCN#175816). PER FOLLOW UP RESPONSE VIA EMAIL ON 16JUL2024, CUSTOMER STATED THAT THEY WERE ALL THE SAME LOT#NGHZ3646. NO ADDITIONAL PATIENT IMPACT FROM THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19805825 ↗
  • MalfunctionNov 17, 2023Gel Leak; Nonstandard Device; Gel Leak; Nonstandard Device

    IT WAS REPORTED THAT THE INSIDE THIS PRODUCT COMES AN ACCESSORY, CALLED URINARY CATHETER LUBRICATION GEL. THIS ACCESSORY WAS CHEMICALLY CHANGED, THAT IS, THE CLIENT INDICATES THAT IT CAME DRY INSIDE THE PACKAGING, FOR THIS REASON IT COULD NOT BE USED, THAT IS, THIS LUBRICANT WAS USED SO THAT THE URINARY CATHETER PASSES THROUGH THE URINARY MEATUS WITHOUT RESISTANCE AND MORE SOFT. IT WAS STATED THAT NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18159736 ↗
  • MalfunctionMar 24, 2022Material Invagination; Material Invagination

    IT WAS REPORTED THAT UPON DEFLATING THE BALLOON, IF THERE WAS ANY PRESSURE BETWEEN THE BALLOON AND THE BLADDER WALL, A RIDGE GETS FORMED ON THE SILICONE CATHETER PREVENTING THE CATHETER FROM BEING REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13897184 ↗
  • MalfunctionNov 18, 2021Material Invagination; Material Invagination

    IT WAS REPORTED THAT UPON REMOVING THE FOLEY CATHETER, THE BALLOON WAS DEFLATED BUT IF THERE WAS ANY PRESSURE AGAINST THE BALLOON DURING THIS STEP, A FOLD WILL OCCUR IN THE SILICONE CIRCUMFERENTIALLY. THIS PREVENTED THE FOLEY CATHETER FROM BEING REMOVED. THIS HAD HAPPENED MULTIPLE TIMES WITH BOTH THE TRAY PRODUCT NUMBER 897216 AND WITH TRAY PRODUCT NUMBER 947316. AND ONE PATIENT NEEDED TO GO TO HAVE A PROCEDURE TO REMOVE THE CATHETER. THE CUSTOMER STATED THAT THEY HAVE A LARGE NUMBER OF THESE FOLEY TRAYS THAT WERE NOT WORKING FOR THEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12836629 ↗
  • MalfunctionOct 4, 2021Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT TWO SILICONE FOLEY CATHETER BALLOON WERE POPPED. PER FOLLOW UP RECEIVED VIA EMAIL ON (B)(6) 2021, CUSTOMER STATED THAT THE BALLOON WAS NOT INFLATED PRIOR TO INSERTION ON ANY OF THE FOLEY CATHETERS AND THE INCIDENT HAPPENED AT THE TIME OF INSERTION WHEN THE NURSE INJECTED 10CC OF FLUID INTO THE BALLOON. THE NURSE DISCARDED BOTH FAILED FOLEYS AS THEY WERE INSIDE THE PATIENT AT THE TIME OF FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12568172 ↗

FDA clearance and classification

The FDA regulates product code EZL (catheter, retention type, balloon) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316

Has the Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3115-2026 (Class II), began on Jun 25, 2026: Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 4, 2021 and Jul 23, 2024. In those reports, the most common specific device problems are “Material Invagination” (2), “Burst Container or Vessel” (1) and “Gel Leak” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter 947316, and is it still sold?

The FDA's device database lists C. R. Bard, Inc. as the company responsible for it (brand Bard® Foley Tray With Lubri-Sil® All-Silicone Foley Catheter, model 947316). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, RETENTION TYPE, BALLOON devices

All CATHETER, RETENTION TYPE, BALLOON devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page