PowerPICC Catheter 3175135Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC Catheter 5F Single-Lumen Polyurethane Catheter Radiology Basic Tray (135 cm Guidewire) · Class 2 · Implant
All models of PowerPICC Catheter (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741027673), received between Mar 28, 2016 and Oct 15, 2025 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Unraveled Material2
- Break1
- Cut In Material1
- Difficult to Insert1
- Entrapment of Device1
- Fluid/Blood Leak1
- Kinked1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Foreign Body In Patient2
- No Information1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2025Break; Break
IT WAS REPORTED, BROKEN LINE. ADDITIONAL INFORMATION PROVIDED 09-OCT-2025: CATHETER BREAK WAS EXTERNAL TO THE PATIENT. IT IS UNKNOWN IF THE BREAK WAS COMPLETE OR PARTIAL. THE PATIENT DID NOT HAVE ANY SYMPTOMS ASSOCIATED WITH THE BREAK. NOTHING WAS INFUSING AT THAT POINT OF TIME. INTERVENTION INCLUDED REMOVAL AND REPLACEMENT OF NEW PICC. THERE WAS NO DELAY IN TREATMENT. CATHETER WAS SECURED WITH PICC STATLOCK AND DRESSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23294816 ↗MalfunctionOct 15, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THE STATLOCK NOT STICKY EVEN THOUGH IT IS PUT ON FOR 1 DAY ONLY. CUSTOMER CONFIRMED SKIN PREP PAD WAS USED PRIOR TO APPLYING THE STATLOCK. THE EVENT DID NOT CAUSE THE CATHETER TO DISLODGE. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23285721 ↗MalfunctionDec 22, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "5FR GUIDEWIRE FOLDS OVER ON ITSELF IN IR PLACEMENT." ADDITIONAL INFORMATION PROVIDED: IT WAS REPORTED BY THE CUSTOMER, THAT THE CATHETER WAS SUTURED IN PLACE AFTER INSERTION. NO HARM TO THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16034985 ↗InjuryOct 25, 2022Material Frayed; Material Frayed
IT WAS REPORTED BY THE CUSTOMER THAT GUIDEWIRE TIP BROKE OFF. "WHAT ARE THE DETAILS OF THE INCIDENT?": PICC LINE WIRE INSERTED THROUGH NEEDLE ONCE NEEDLE WAS IN THE VEIN. WIRE HAD DIFFICULTY ADVANCING. AS GUIDEWIRE WAS BEING RETRACTED THROUGH THE SKIN, TIP BROKE OFF AND WAS LEFT INSIDE THE PATIENTS ARM. THE BROKEN PIECE WAS REMOVED OPERATIVELY. FOUND NOT TO BE INTRA-VASCULAR, NO REPORTED COMPLICATIONS POST-OP. DATE OF INCIDENT: (B)(6) 2022. "WHAT WAS THE TYPE OF INCIDENT/PROBLEM?" FAILURE (I.E. WHILE PREPARING FOR, IN USE, AFTER USE). "WHAT WAS THE LEVEL OF HARM?" : HARM OR CLINICALLY SERIOUS ADVERSE EVENT. "DID THIS PROBLEM/INCIDENT RESULT IN UNEXPECTED OR PROLONGED CARE?" : YES "WHAT IS THE FREQUENCY OF THIS PROBLEM?" : FIRST TIME. "WAS THIS AN INVASIVE PROCEDURE?" : YES. "WHAT PROCEDURE?" : PICC LINE INSERTION WITH REMOVAL OF RETAINED PIECE.
Patient problems: Foreign Body In Patient
MAUDE report 15671204 ↗MalfunctionJun 2, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT GUIDEWIRE WAS FOUND KINK DURING INSERTION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14581037 ↗MalfunctionSep 9, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LUMEN WAS LEAKING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 10512086 ↗InjuryAug 3, 2018Entrapment of Device; Retraction Problem; Unraveled Material
IT WAS REPORTED THAT AFTER THE PUNCTURE IN THE LEFT BASILICA VEIN WITH THE PUNCTURE NEEDLE, THE GUIDE WAS INSERTED. IT WAS STATED DURING THE GUIDE WITHDRAWAL IT DIDN'T COME OUT. AFTER READJUSTMENTS, THE DOCTOR WAS STILL TRYING TO WITHDRAWAL IT AND DECIDED TO REMOVE IT WITH THE PUNCTURE NEEDLE. THE NEEDLE WAS REMOVED SUCCESSFULLY, HOWEVER IT WAS REPORTED THE GUIDE WAS STILL STUCK INSIDE THE VEIN AND DURING ITS WITHDRAWAL THE TIP FELL APART, LIKE IT WAS UNRAVELING, LEAVING PART OF THE MATERIAL (THIN LINES OF THE NEEDLE TIP) INSIDE THE PATIENT'S VEIN, WHICH WAS CONFIRMED BY RADIOSCOPY (RADIOPAQUE MATERIAL IN THE PATIENT'S ARM AFTER THE WITHDRAWAL OF THE NEEDLE AND GUIDE). IT WAS NECESSARY THE SURGICAL WITHDRAWAL OF THE MATERIAL.
Patient problems: Foreign Body In Patient
MAUDE report 7748557 ↗MalfunctionSep 6, 2017Cut In Material
WHILE PREPARING FOR THE PICC PROCEDURE, IT WAS FOUND THAT THE GUIDEWIRE WAS CUT AND COULD NOT BE USED FOR THE PROCEDURE.
Patient problems: No Information
MAUDE report 6847268 ↗MalfunctionMar 28, 2016Difficult to Insert; Kinked; Sticking; Unraveled Material
ON (B)(6) 2016- IT WAS REPORTED THAT THE PATIENT WAS PREPPED FOR PICC INSERTION AND THE DOCTOR INSERTED THE GUIDEWIRE AND ENCOUNTERED DIFFICULTY WHILE THREADING THE WIRE THROUGH THE VESSEL. IT REPORTEDLY KINKED AND WAS STUCK AND WHEN HE TRIED TO MOVE IT, IT UNRAVELED. THE DOCTOR WAS ABLE TO REMOVE ALL THE WIRE FROM THE VESSEL. A NEW PICC SET WAS OPENED AND THE GUIDEWIRE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5530628 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page