PowerPICC® Catheters 1174108DRecalls & reported problems
Bard Access Systems, Inc. · PowerPICC Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 4F · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard PowerPICC catheter has 222 adverse event reports, including 213 malfunctions and 9 injuries, and no deaths. Many reports describe the catheter leaking, bending or deforming, and some report breaks or a hole in the tubing. There is 1 open recall, begun in March 2025, over tubing that may develop leaks over time and let fluid leak into surrounding tissue. Reports fell from 18 in the prior year to 4 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 222
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 222 adverse event reports that list this exact device (DI 00801741034459), received between Nov 3, 2015 and May 13, 2026 — classified as 9 injuries, 213 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (107 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak107
- Material Deformation54
- Material Twisted/Bent17
- Break12
- Material Frayed9
- Physical Resistance/Sticking6
- Contamination /Decontamination Problem5
- Device Contamination with Chemical or Other Material5
Patient problems
- No Clinical Signs, Symptoms or Conditions135
- No Known Impact Or Consequence To Patient42
- No Consequences Or Impact To Patient21
- Pain5
- Insufficient Information3
- No Patient Involvement3
- Device Embedded In Tissue or Plaque2
- Discomfort2
Recent adverse event reports
The 25 most recent of 222 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT IRREGULAR EDGE ON DILATOR IN PICC KIT THAT WILL NOT PUSH THROUGH THE VEIN. HAD TO PULL NEW DILATOR AND FINISHED THE LINE - UNKNOWN IF ANY DAMAGE DONE TO THE VESSEL. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25164097 ↗MalfunctionFeb 2, 2026Fluid/Blood Leak
IT WAS REPORTED THAT 4FR DL SV WITH A HOLE. THE HOLE IS RIGHT BELOW THE HUB. NO FURTHER DETAILS AVAILABLE OR PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24235310 ↗MalfunctionNov 25, 2025Material Frayed; Material Twisted/Bent; Material Frayed; Material Twisted/Bent
IT WAS REPORTED GUIDEWIRE NOTED TO BE DIFFICULT TO REMOVE AND REPLACE IN LINE PRIOR TO PLACEMENT IN PATIENT. UPON REMOVAL, GUIDEWIRE CURLED LIKE A RIBBON AND MADE THE LINE UNUSABLE. PATIENT'S SKIN WAS SHREDDED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Abrasion
MAUDE report 23646606 ↗MalfunctionNov 25, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ADDITIONAL INFORMATION: WE WERE NOT INFORMED OF ANY LEAKAGE THAT OCCURRED PRIOR TO THE INCIDENT OF ATTEMPTED REMOVAL. THE PATIENT WAS ON DAPTOMYCIN 450MG ONCE DAILY INFUSED OVER 30MIN, ZOSYN 3.375GM Q8HR INFUSED OVER 4H, AND NS AT 100ML/HR CONTINUOUSLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645915 ↗MalfunctionAug 5, 2025Break; Break
IT WAS REPORTED THAT A PICC BROKE. PATIENT WAS WHEELING HER WHEELCHAIR AND HEARD SOMETHING HIT THE FLOOR...IT WAS THE TOP OF HER PICC. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22714557 ↗MalfunctionJul 14, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY CUSTOMER THAT SAFETY NEEDLE MALFUNCTION, GREEN CAP CAME COMPLETELY OFF THE END. NO INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22500360 ↗MalfunctionJul 10, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER DURING SINGLE LUMEN 4FR PICC INSERTION, RESISTANCE MET BUT ABLE TO FULLY PASS CATHETER. MULTIPLE PASSES MADE WITH PATIENT PERFORMING VAGAL MANEUVERS AND POSITION CHANGES AS PICC WOULD NOT DROP. WHEN 3CG WIRE WAS REMOVED FROM CATHETER AND INSPECTED, KINK NOTED AND CATHETER INTEGRITY APPEARED TO BE COMPROMISED. THIS WIRE WAS IMMEDIATELY REMOVED FROM THE PROCEDURE AND A NEW KIT WAS USED FOR THE REST OF THE PROCEDURE AND TIP OF CATHETER WAS PRESENT UPON INSPECTION. NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22468119 ↗MalfunctionJul 1, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT PICC LINE HAS A HOLE AND LEAKS AT THE 5CM MARK. PATIENT NEEDED NEW LINE PLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22378954 ↗MalfunctionJun 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT 4FR SINGLE LUMEN PICC FRACTURED. THE TEAR WAS BETWEEN THE 2CM AND 3CM MARK WITH ZERO BEING TRUE ZERO. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22196114 ↗InjuryMay 22, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED; LEAKING LINE OCCURRED OVER THE WEEKEND. THE EXTERNAL LENGTH OF THE LINE HAD A FRACTURE/HOLE AND WAS CONTINUOUSLY LEAKING (A MORPHINE DRIP WAS RUNNING). NURSING STAFF HAS ¿PLUGGED¿ THE HOLE BY WRAPPING A TEGADERM AROUND IT. THE NEGATIVE OUTCOME FOR THIS INCIDENT WAS THE DELAY IN CARE. PATIENT¿S CONTINUOUS MORPHINE DRIP HAD TO BE HELD FROM THE TIME THE LINE WAS REMOVED UNTIL THE NEW LINE WAS INSERTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22071723 ↗MalfunctionMay 21, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THE PICC EXTERNAL TUBING WAS FOUND TO BE LEAKING ON MONDAY (B)(6) 2025; A HOLE/CRACK WAS PRESENT. A MORPHINE DRIP WAS RUNNING. THE NURSING STAFF "PLUGGED" THE HOLE BY WRAPPING A TEGADERM AROUND IT (NURSING STAFF HAS SINCE BEEN EDUCATED). THERE WAS NO HARM TO THE PATIENT, BUT A NEW PICC HAD TO BE PLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22063410 ↗MalfunctionMay 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED 4F SL PICC WITH HOLE - "PATIENT HAD PICC PLACED ON 3/20 (LOT #REJW2363). DISCHARGED FROM HOSPITAL THEN CAME BACK FOR TROUBLESHOOTING (NOT DRAWING BACK/HARD TO FLUSH). CHEST X-RAY SHOWED IT WAS IN APPROPRIATE POSITION BUT WE HAD TO REPLACE D/T MALFUNCTION. AFTER REMOVAL, TEAM ASSESSED AND NOTICED FLUSH WAS EXITING THRU HOLE/CUT BETWEEN 5-6 CM MARKS." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22030783 ↗MalfunctionApr 30, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THEY REMOVED A PICC LINE FROM A PATIENT TODAY (IT WAS PLACED ON (B)(6) 2025. THE EXPOSED CATHETER HAD A FRACTURE/HOLE AND WAS LEAKING. THE PATIENT STATES THERE WAS NO UNUSUAL ACTIVITY ON HIS BEHALF AND HE DID NOT SEE ANY STAFF USING ANY SHARP INSTRUMENTS NEAR THE LINE. I¿M NOT SURE IF THIS WAS AN ISSUE WITH THE PRODUCT ITSELF. THE PICC WAS REMOVED DUE TO THE ISSUE WITH LEAKING. IT WAS NOTICED DURING ROUTINE FLUSHING OF THE LINE. THE LEAK WAS IN THE EXTERNAL LINE, AT ABOUT THE 1 CM MARK. THERE WERE 5 CM EXTERNAL UPON PLACEMENT AND THE LINE WAS IN PLACE FOR 30 DAYS BEFORE THE LEAK WAS REPORTED. THERE WAS NOTHING CONTINUOUSLY INFUSING AT THE TIME OF THE REPORTED LEAK. THE PATIENT WAS ON ABX THERAPY. THE EVENT DID NOT CAUSE A DELAY IN TREATMENT AS A NEW PIV WAS PLACED WHEN THE LINE WAS REMOVED. THE PATIENT RECEIVED THEIR REMAINING ABX THROUGH PERIPHERAL IVS. THERE WERE NO OTHER SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE LEAK. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21940144 ↗MalfunctionApr 4, 2025Activation Problem; Activation Problem
IT WAS REPORTED WHEN PLACING A PICC LINE YESTERDAY I HAD THE ECHOGENIC NEEDLE WITHIN THE KIT MALFUNCTION. WHILE REMOVING THE NEEDLE FROM THE SKIN THERE IS A SAFETY MECHANISM THAT IS NORMALLY ABLE TO BE ACTIVATED AS YOU WITHDRAW THE NEEDLE FROM THE SKIN SO THAT YOU ARE NEVER EXPOSED TO A CONTAMINATED SHARP. AS I WAS WITHDRAWING THE NEEDLE OVER THE GUIDEWIRE POST PLACEMENT, I ATTEMPTED TO ACTIVATE THE SAFETY MECHANISM AND IT DETACHED COMPLETELY FROM THE END OF THE CONTAMINATED NEEDLE. I WAS ABLE TO PLACE THE CONTAMINATED SHARP UNDER THE KIT IN MY STERILE FIELD UNTIL THE PROCEDURE WAS COMPLETED AND THEN PLACE IT SAFELY IN A SHARPS CONTAINER. NO PATIENT OR CLINICIAN HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21779515 ↗MalfunctionMar 26, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THE INTRODUCER COLLAPSED ON ITSELF WHEN TRYING TO ADVANCE, ALMOST LIKE A MUSHROOM SHAPE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21700604 ↗MalfunctionMar 20, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT THEY PLACED A SINGLE LUMEN PICC ON (B)(6) (LOT #REJW2342) THAT KINKED/LOOPED JUST BEYOND THE INSERTION SITE, WHICH WE DETERMINED ON UE XRS TODAY WHILE TROUBLESHOOTING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21653553 ↗MalfunctionMar 20, 2025Material Deformation; Material Deformation
ADDITIONAL INFORMATION RECEIVED ON MARCH 26, 2025: LINE DID NOT WORK (UNABLE TO USE FOR BLOOD DRAWS, TPA GIVEN), REQUIRED REMOVAL AND REPLACEMENT. NO SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21653328 ↗MalfunctionFeb 19, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED BY THE CUSTOMER THAT THE INSIDE OVER FOR THE TRAY WAS RIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21411502 ↗MalfunctionFeb 19, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE PATIENT PAIN IN UPPER ARM WITH ALL APPROPRIATE STUDIES SHOWING DONE AND NORMAL. WHILE REMOVING THE OLD PICC A HOLE WAS DISCOVERED AT APPROXIMATELY 3 CM. THE PICC WAS REPLACED. NO OTHER INFORMATION WAS PROVIDED AND NO PATIENT HARM REPORTED.
Patient problems: Pain
MAUDE report 21407817 ↗MalfunctionFeb 6, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED PICC WAS PLACED (B)(6) 2024 AND DISCONTINUED ON (B)(6) 2024 AFTER THE PATIENT COMPLAINED OF MODERATE PAIN DURING FLUSHING AND INFUSIONS. ALSO COMPLAINED OF LEAKING OF FLUIDS AT INSERTION SITE. WHEN REMOVED THE LINE WAS FOUND TO HAVE 2 BREAKS TO THE PICC LINE. THEY APPEARED TO BE HORIZONTAL BREAKS LOCATED AT THE 11 AND 13CM MARKS. TEMPORARY PAIN AND MILD SWELLING TO UPPER ARM IN REGARDS TO IVF INFILTRATION. NO PERMANENT DAMAGE DISCOVERED.
Patient problems: Pain; Swelling/ Edema
MAUDE report 21328921 ↗MalfunctionJan 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER, PICC KINKED. PLACED (B)(6) 2024 IN THE MIDDLE 1/3 OF RUE, BASILIC VEIN. CATHETER TRIMMED 45CM, EXTERNAL LENGTH 1CM, BREAK AT 6CM, SECURED WITH AT STATLOCK, PICC WAS NOT TPA'D, AND DWELLED FOR 9 DAYS. PICC EXCHANGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21265297 ↗MalfunctionJan 23, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY THE CUSTOMER THE GREEN ACCESS NEEDLE PLASTIC BROKE AT THE TOP. THIS HAPPENED AS THE CLINICIAN WAS INSERTING THE NEEDLE INTO THE PATIENT. THE END USER WAS SUCCESSFUL WITH PLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21219757 ↗MalfunctionOct 18, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT RN STATES PICC IS LEAKING. UPON ASSESSMENT PICC WOULD NOT FLUSH EASILY, AND WHEN I TRIED TO DRAW BLOOD, AIR CAME INTO SYRINGE. I IMMEDIATELY REMOVED THE DRESSING AND UPON FLUSHING I SAW A STREAM OF FLUID COMING FROM THE PLACE IN THE IMAGE. ITS NOT A SLIT, BUT A PINHOLE. I DC ED PICC INTACT AND PLANNING ON PLACING NEW PICC LATER TODAY. NO PATIENT INJURY. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20488087 ↗MalfunctionSep 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, ENCOUNTERED AN ISSUE WITH A BARD PICC LEAKING WHEN FLUSHED. IT WAS INDEED LEAKING. WHEN D/C¿ING THE PICC I FOUND THAT IT WAS PARTIALLY SEVERED AT APPROXIMATELY THE 5CM MARK. TO COMPLICATE MATTERS, I WAS NOT ABLE TO FULLY D/C THE PICC. I WAS ONLY ABLE TO REMOVE IT HALF WAY. MET LOTS OF RESISTANCE. TRIED A FEW DIFFERENT TECHNIQUES WITHOUT SUCCESS. THE FACILITY WAS GOOD TO SEND THE PATIENT TO THE HOSPITAL FOR FURTHER EVALUATION. PICC PLACED 07/05/2024 IN THE RIGHT BRACHIAL. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20164232 ↗MalfunctionJul 25, 2024Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT THE PICC 3CG WIRE WAS BENT. MISSED STICK ATTEMPT AND DISCOMFORT DURING STICK. NO OTHER INFORMATION WAS PROVIDED. 07/03/2024 - THE DEVICE RETURNED FOR EVALUATION WAS KINKED ON THE STYLET.
Patient problems: Discomfort
MAUDE report 19826700 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.