Skip to content

PowerPICC® Catheter 3275108DRecalls & reported problems

Bard Access Systems, Inc. · Dot ppicc, 5fr DL · Class 2 · Implant

All models of PowerPICC Catheter (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741027871), received between Apr 15, 2022 and Apr 11, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Defective Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Component2
  • Fluid/Blood Leak1
  • Material Deformation1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2024Defective Component; Defective Component

    IT WAS REPORTED BY THE CUSTOMER THE PICC WAS USED FOR AQUADEX PEDIATRIC PATIENT AND PLACED IN THE IJ. THIS LINE DWELLED IN THE PATIENT FROM (B)(6) 2023-(B)(6) 2024. THE LINE WAS DISCONTINUED AFTER THE PATIENT'S STATUS IMPROVED AND NO LONGER NEEDED AQUADEX. THE SUTURE WINGS CRACKED OFF. THEY WERE ABLE TO RE-SUTURE THE PICC AND USE IT FOR ANOTHER 2 MONTHS. THE TREATMENT WAS COMPLETED AND THE LINE WAS DISCONTINUED ON (B)(6) 2024. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19091921 ↗
  • MalfunctionApr 11, 2024Defective Component; Defective Component

    IT WAS REPORTED BY THE CUSTOMER PICC USED FOR AQUADEX PEDIATRIC PATIENT AND PLACED IN THE IJ ON (B)(6) 2021 AND THE WING CRACKED ON 3/14/24. THE LINE WAS DISCONTINUED BY PROVIDERS ON 3/19 AFTER THE PATIENT'S STATUS IMPROVED AND NO LONGER NEEDED AQUADEX. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19091073 ↗
  • InjuryMar 7, 2024Material Frayed; Material Frayed

    IT WAS REPORTED BY CUSTOMER THAT A PERIPHERALLY INSERTED CENTRAL CATHETER (PICC LINE) WAS INSERTED, AND THE GUIDE WIRE WAS LOST IN THE VEIN. WHEN IDENTIFIED, THE GUIDE WIRE HAD APPEARED BROKEN IN TWO AND REMOVED IN TWO PIECES. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18856499 ↗
  • MalfunctionNov 7, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THAT PICC PULLS AIR WHEN CONNECTED TO EXTENSION SET. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18088824 ↗
  • MalfunctionApr 15, 2022Material Deformation; Material Deformation

    IT WAS REPORTED "REPORTED AN ISSUE WITH THE TRANSITION ON THE SHEATH BEING NOT AS SMOOTH AS IT HAS BEEN IN THE PAST." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14118389 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page