PowerPICC Catheter 3275118Recalls & reported problems

Bard Access Systems, Inc. · PowerPICC Catheter 5F Nurse Basic Tray with Microintroducer (Dual-Lumen) · Class 2 · Implant

All models of PowerPICC Catheter (33)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Bard PowerPICC catheter appears in 266 adverse event reports, with no deaths and 23 injuries. The reports mostly describe fluid leaks, deformed or frayed material, contamination and breaks, along with some reports of blood clots, swelling and redness at the insertion site. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
265
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 265 adverse event reports that list this exact device (DI 00801741027918), received between Jun 10, 2016 and Aug 31, 2026 — classified as 22 injuries, 243 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (74 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 1 reportsAug 2024: 6 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 6 reportsFeb 2025: 3 reportsMar 2025: 5 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 6 reportsJan 2026: 6 reportsFeb 2026: 10 reportsMar 2026: 2 reportsApr 2026: 5 reportsMay 2026: 2 reportsJun 2026: 5 reportsJul 2026: 3 reportsAug 2026: 4 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak74
  • Material Deformation44
  • Contamination /Decontamination Problem26
  • Material Frayed25
  • Defective Component19
  • Break14
  • Corroded12
  • Adverse Event Without Identified Device or Use Problem11

Patient problems

  • No Clinical Signs, Symptoms or Conditions192
  • No Known Impact Or Consequence To Patient25
  • No Consequences Or Impact To Patient16
  • Thrombosis/Thrombus11
  • Pain9
  • Swelling/ Edema8
  • Foreign Body In Patient5
  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 25 most recent of 265 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533107 ↗
  • MalfunctionAug 31, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533105 ↗
  • MalfunctionAug 26, 2026Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50452722 ↗
  • MalfunctionAug 19, 2026Product Quality Problem; Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323165 ↗
  • MalfunctionJul 10, 2026Material Deformation

    IT WAS REPORTED THAT AFTER DILATION, THERE WAS RESISTANCE WHEN INSERTING THE MICROINTRODUCER INTO THE BLOOD VESSEL; UPON INSPECTION, BURRS WERE FOUND AT THE DISTAL END OF THE SHEATH. NO FURTHER DETAILS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25781934 ↗
  • InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED AFTER THE CATHETER WAS INSERTED, THE PATIENT EXPERIENCED REDNESS AND SWELLING AT THE PUNCTURE SITE, SKIN REPAIR. AN ALLERGIC REACTION TO THE DRESSING WAS SUSPECTED. NO FURTHER DETAILS PROVIDED.

    Patient problems: Erythema; Local Reaction; Swelling/ Edema

    MAUDE report 25746736 ↗
  • MalfunctionJul 1, 2026Fluid/Blood Leak

    IT WAS REPORTED A POWERPICC CATHETER RUPTURED. PRODUCT IS STILL BEING USED ON THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679124 ↗
  • MalfunctionJun 9, 2026Contamination /Decontamination Problem

    IT WAS REPORTED DURING THE CATHETERIZATION PROCESS, A NURSE DISCOVERED AN FOREIGN MATERIAL ON THE STYLET WHILE TRIMMING THE CATHETER. THE ISSUE WAS RESOLVED AFTER REPLACING THE PRODUCT. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446088 ↗
  • MalfunctionJun 9, 2026Material Deformation

    IT WAS REPORTED DURING CATHETER PLACEMENT, THE PUNCTURE SHEATH WAS BLUNT, LEADING TO A SPLIT IN THE DELIVERY SHEATH; A NEW CATHETER SET WAS USED TO REINSERT THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25444366 ↗
  • MalfunctionJun 5, 2026Material Deformation

    IT WAS REPORTED DURING CATHETER PLACEMENT, THE PUNCTURE SHEATH WAS BLUNT, LEADING TO A SPLIT IN THE DELIVERY SHEATH; A NEW CATHETER SET WAS USED TO REINSERT THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25411058 ↗
  • MalfunctionJun 2, 2026Material Deformation

    IT WAS REPORTED THAT DURING CATHETER PLACEMENT, THE PUNCTURE SHEATH WAS BLUNT, LEADING TO A SPLIT IN THE DELIVERY SHEATH; A NEW CATHETER SET WAS USED TO REINSERT THE CATHETER. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25360137 ↗
  • MalfunctionJun 1, 2026Material Deformation

    IT WAS REPORTED THE SHEATH TORE DURING CATHETER INSERTION, PREVENTING SUCCESSFUL PUNCTURE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25355433 ↗
  • MalfunctionMay 26, 2026Material Deformation

    IT WAS REPORTED THAT DURING USE, THE VASCULAR SHEATH HAS RUPTURED AND THE DEVICE IS UNUSABLE. NO FURTHER DETAILS AVAILABLE. IT WAS REPORTED THIS OCCURRED WITH 3 DEVICES. THIS REPORT ADDRESSES ALL THREE DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25293044 ↗
  • MalfunctionMay 4, 2026Contamination /Decontamination Problem

    IT WAS REPORTED WHITE SPOTS ON THE GUIDEWIRE, PREVENTING INSERTION INTO THE BLOOD VESSEL. NO FURTHER INFORMATION WAS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH EIGHT DEVICES. THIS REPORT ADDRESSES ALL EIGHT DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073364 ↗
  • MalfunctionApr 28, 2026Material Frayed

    IT WAS REPORTED THAT THE GUIDEWIRE INSIDE THE CATHETER IS EXTREMELY SOFT AND FRAYED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25019903 ↗
  • MalfunctionApr 22, 2026Difficult or Delayed Separation

    IT WAS REPORTED THE PUNCTURE BRIDGE COULD NOT BE FULLY TORN OPEN, PREVENTING THE CATHETER FROM DETACHING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24952853 ↗
  • MalfunctionApr 13, 2026Detachment of Device or Device Component

    IT WAS REPORTED THE PATIENT UNDERWENT PICC PLACEMENT UNDER ULTRASOUND GUIDANCE ON (B)(6). AROUND 10:00 A.M. ON JANUARY 20, A NURSE DISCOVERED THAT THE PICC CONNECTOR HAD BECOME DISCONNECTED FROM THE CATHETER DURING A ROUTINE REPOSITIONING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24865103 ↗
  • MalfunctionApr 2, 2026Fluid/Blood Leak

    IT WAS REPORTED FLUID LEAKAGE WAS OBSERVED AT THE PUNCTURE SITE WHILE THE CATHETER WAS IN PLACE. UPON REMOVAL AND EXAMINATION, FLUID LEAKAGE WAS FOUND APPROXIMATELY 15 CM FROM THE CATHETER TIP. THE CATHETER WAS REMOVED, REPLACED, AND REINSERTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765523 ↗
  • MalfunctionApr 2, 2026Material Deformation

    IT WAS REPORTED THAT DURING CATHETER PLACEMENT, THE SEDINGGER SHEATH DID NOT ADVANCE SMOOTHLY. UPON WITHDRAWAL, NO ABNORMALITIES WERE OBSERVED, SO IT WAS REINSERTED. THE PATIENT REPORTED DISCOMFORT AT THE PUNCTURE SITE. UPON WITHDRAWAL, A TEAR WAS DISCOVERED IN THE TEAR-AWAY SHEATH AT THE CONNECTION POINT WITH THE DILATOR. A NEW SET WAS IMMEDIATELY OPENED, AND THE CATHETER WAS SUCCESSFULLY PLACED IN THE PATIENT. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Discomfort

    MAUDE report 24765044 ↗
  • MalfunctionMar 12, 2026Fluid/Blood Leak

    IT WAS REPORTED, OPENED A 3275118 CATHETER DURING PLACEMENT. UPON INSPECTING THE CATHETER'S INTEGRITY, A PINHOLE DEFECT WAS DISCOVERED, RENDERING IT UNUSABLE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584747 ↗
  • MalfunctionMar 9, 2026Restricted Flow rate

    IT WAS REPORTED THE CONTRAST PUMP WAS NOT SUPPORTING THE FLOW RATE OF 4.5 ML, LOCKING UP AND REDUCING IT TO 3.5 ML, DAMAGING THE IMAGES OF THE ANGIOTOMY EXAMS CAUSING INCORRECT TEST RESULTS. THE DEVICE REMAINS IMPLANTED IN THE PATIENT. NO FURTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24545342 ↗
  • MalfunctionFeb 25, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT 25 DAYS AFTER CATHETER PLACEMENT, DRAINAGE WAS OBSERVED AT THE PUNCTURE SITE. UPON REMOVAL AND INSPECTION, LEAKAGE WAS DETECTED APPROXIMATELY 17 CM FROM THE CATHETER TIP. THE CATHETER WAS REMOVED, REPLACED, AND REINSERTED. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446660 ↗
  • InjuryFeb 24, 2026Fluid/Blood Leak

    ADDITIONAL INFORMATION RECEIVED ON 26-FEB-2026 CUSTOMER CONFIRMED 1 UNIT AFFECTED. PATIENT PRESENTED WITH CHEMICAL PHLEBITIS AFTER USING CONTRAST. HE PRESENTED WITH HYPEREMIA, REDNESS AND PAIN. THE PRODUCT WAS DISCARDED.

    Patient problems: Erythema; Extravasation; Vasculitis

    MAUDE report 24434804 ↗
  • MalfunctionFeb 24, 2026Product Quality Problem

    IT WAS REPORTED THAT THE PATIENT WAS ADMITTED FOR FURTHER TREATMENT FOLLOWING A DIAGNOSIS OF ACUTE MYELOID LEUKEMIA. ON (B)(6) 2026, AT (B)(6), A BEDSIDE PICC CATHETER WAS INSERTED USING A PERIPHERAL-TO-CENTRAL CATHETER. A SELDINGER TIP BARB WAS ENCOUNTERED, CAUSING DIFFICULTY IN SHEATH REMOVAL AND DELAYING THE PATIENT'S TREATMENT. USE WAS IMMEDIATELY DISCONTINUED. THE PRODUCT WAS REPLACED WITH AN IDENTICAL, UNCOMPROMISED ITEM, AND TREATMENT CONTINUED. NO FURTHER HARM RESULTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430171 ↗
  • MalfunctionFeb 20, 2026Restricted Flow rate

    IT WAS REPORTED THAT THE CUSTOMER CONTACTED ME ON TUESDAY, FEBRUARY 3, CLAIMING THAT THERE HAD BEEN PROBLEMS WITH THE CONTRAST PUMP, WHICH WAS UNABLE TO SUPPORT A FLOW RATE OF 4.5 ML, LOCKING UP AND REDUCING TO 3.5 ML, COMPROMISING THE IMAGES FROM THE ANGIOTOME EXAMS. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403543 ↗

FDA clearance and classification

The FDA regulates product code LJS (catheter,intravascular,therapeutic,long-term greater than 30 days) under 21 CFR 880.5970 as a Class II device. 21 CFR 880.5970 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PowerPICC Catheter 3275118

Has the PowerPICC Catheter 3275118 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00801741027918), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PowerPICC Catheter 3275118?

The FDA's MAUDE database lists 265 adverse event reports for this device, received between Jun 10, 2016 and Aug 31, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (74), “Material Deformation” (44) and “Contamination /Decontamination Problem” (26); the most common patient problems are “Thrombosis/Thrombus” (11), “Pain” (9) and “Swelling/ Edema” (8). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PowerPICC Catheter 3275118?

None of the reports are classified as deaths, 22 as injuries and 243 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PowerPICC Catheter 3275118 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the PowerPICC Catheter 3275118, and is it still sold?

The FDA's device database lists Bard Access Systems, Inc. as the company responsible for it (brand PowerPICC Catheter, model 3275118). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →