PowerPICC SOLO2 Catheter 3295108DRecalls & reported problems
Bard Access Systems, Inc. · Dot ppicc solo, 5fr dl · Class 2 · Implant
All models of PowerPICC SOLO2 Catheter (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741028052), received between Feb 15, 2019 and Sep 22, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Contamination” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Contamination1
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2023Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT THEY ARE HAVING ISSUES WITH PULLING AIR IN PICCS WHEN THEY ARE CHECKING FOR BLOOD RETURN. THERE SEEMS TO BE A LEAK IN THE STIFFENING WIRE LUMEN THAT CAUSES AIR TO PULL AND NOT BLOOD. RESPONSE RECEIVED 13 SEP 2023 : THIS PULLS AIR INTO THE SALINE SYRINGE WHEN ASPIRATING FOR BLOOD RETURN. IT ALSO CAUSES SALINE TO LEAK OUT OF THE RUBBER STOPPER WHERE THE STIFFENING WIRE EXITS. IT ALSO CAUSES THE BLOOD RETURN CHECK TO BE MISLEADING, THEREFORE LEADING THE INSERTER TO BELIEVE THAT THE PICC IS NOT IN THE PROPER LOCATION. WE HAVE NOTICED THIS OCCURRING DURING MULTIPLE INSERTIONS. NO INJURY HAS OCCURRED TO PATIENTS, BUT POTENTIALLY COULD LEAD TO A MALPOSITIONED PICC PLACEMENT OR EXPELLING AIR FROM THE SYRINGE INTO THE PATIENT'S CENTRAL VENOUS SYSTEM. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT. ADDITIONAL INFORMATION RECEIVED: THERE'S NO SAMPLE TO BE RETURNED SINCE THEY HAVE BEEN INSERTED ON PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17800129 ↗MalfunctionFeb 15, 2019Contamination
IT WAS REPORTED THAT THERE WAS AN OPEN SEAM ABOUT ONE INCH BELOW THE LEFT SHOULDER ON THE STERILE GOWN WITHIN THE KIT. ON (B)(6) 2019- ADDITIONAL INFORMATION RECEIVED STATED, " THE CLINICIAN WAS ADVISED TO CHANGE THE GOWN PRIOR TO PROCEEDING WITH INSERTION BUT CHOSE TO UTILIZE THE GOWN FOR THE INSERTION. THE GOWN WAS CONTAMINATED WITH BIOHAZARD AND WAS DISCARDED FOLLOWING COMPLETION OF THE PROCEDURE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8346224 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page