PowerPICC SOLO2 FT Catheter 3295108FDRecalls & reported problems

Bard Access Systems, Inc. · DOT PPICC SOLO FT · Class 2 · Implant

All models of PowerPICC SOLO2 FT Catheter (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741028076), received between Mar 5, 2021 and Dec 12, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation4
  • Failure to Infuse1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2024Failure to Infuse; Failure to Infuse

    IT WAS REPORTED BY THE CUSTOMER, CATHETER NOT POWER FLUSHING WHEN IT FLIPS. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20919615 ↗
  • MalfunctionNov 20, 2023Material Deformation; Material Deformation

    CUSTOMER STATED THAT GUIDEWIRE COILS WHEN BEING INSERTED INTO PATIENT. SHE STATED THIS HAS HAPPENED SEVERAL TIMES, BUT DOES NOT KNOW HOW MANY TIMES OR SPECIFIC DATES. ADDITIONAL INFORMATION PROVIDED 10/30/2023: AFTER GUIDEWIRE COILS, IT MAY OR MAY NOT BE ABLE TO BE USED; MULTIPLE OCCASIONS OF OPENING A SECOND KIT TO REPLACE THE GUIDEWIRE WHICH CAUSES A DELAY OF ACCESS. EVENT DID NOT INVOLVE AN URGENT/LIFE THREATENING MEDICAL SITUATION. EXTENDED TIME FOR INSERTION DUE TO DIFFICULTY AND HAVING TO OPEN ANOTHER KIT. MULTIPLE KITS HAD TO BE USED. NO PATIENTS HARMED. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18175572 ↗
  • MalfunctionNov 20, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER THAT GUIDEWIRE COILS WHEN BEING INSERTED INTO PATIENT. ADDITIONAL INFORMATION PROVIDED 10/30/2023: AFTER GUIDEWIRE COILS, IT MAY OR MAY NOT BE ABLE TO BE USED; OPENING A SECOND KIT TO REPLACE THE GUIDEWIRE WHICH CAUSES A DELAY OF ACCESS. THIS DID NOT CAUSE AN URGENT SITUATION, BUT EXTENDED TIME FOR INSERTION DUE TO DIFFICULTY AND HAVING TO OPEN ANOTHER KIT. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18173890 ↗
  • MalfunctionAug 31, 2021Material Deformation; Material Deformation

    IT WAS REPORTED BY HEALTHCARE PROFESSIONAL "OUR INFUSION NURSE WAS INSERTING A PICC THIS MORNING THAT HAD A FAULTY INTRODUCER; THE END IS ROLLED JUST SLIGHTLY. THEY HAD TO OPEN ANOTHER KIT TO COMPLETE PROCEDURE." NO OTHER INFORMATION WAS PROVIDED. (B)(6) 2021: DURING EVALUATION THE RETURNED DEVICE WAS NOTED TO BE CRUMPLED AT THE DISTAL EDGE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12392011 ↗
  • MalfunctionJul 30, 2021Material Deformation; Material Deformation

    IT WAS REPORTED A FAULTY INTRODUCER. NO OTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12254993 ↗
  • InjuryMar 5, 2021Material Frayed; Material Frayed

    IT WAS REPORTED VIA MEDWATCH "ATTEMPTED TO INSERT PICC LINE IN LUE, ENCOUNTERED SOME RESISTANCE ON THE SUBCLAVIAN REGION WHEN ATTEMPTING TO ADVANCE LINE. PULLED LINE BACK APPROXIMATELY 10 CM, FLUSHED THE LINE AMD ATTEMPTED TO ADVANCE. WHEN THAT ATTEMPT FAILED, THE LINE WAS REMOVED AND PICC WAS PLACED IN THE RIGHT ARM. CHEST X-RAY WAS DONE AND A FOREIGN BODY, PRESUMABLY A WIRE MEASURING ABOUT 3.8CM WAS DETECTED IN THE LEFT SUBCLAVIAN REGION. THE PATIENT HAD BEEN DISCHARGED. THE PATIENT WAS INSTRUCTED TO RETURN TO THE HOSPITAL IMMEDIATELY. THE FOREIGN OBJECT WAS REMOVED ON (B)(6) 2020. "

    Patient problems: Foreign Body In Patient

    MAUDE report 11424112 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates