PowerPICC SOLO2 FT Catheter 3295108FDRecalls & reported problems
Bard Access Systems, Inc. · DOT PPICC SOLO FT · Class 2 · Implant
All models of PowerPICC SOLO2 FT Catheter (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741028076), received between Mar 5, 2021 and Dec 12, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Failure to Infuse1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2024Failure to Infuse; Failure to Infuse
IT WAS REPORTED BY THE CUSTOMER, CATHETER NOT POWER FLUSHING WHEN IT FLIPS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20919615 ↗MalfunctionNov 20, 2023Material Deformation; Material Deformation
CUSTOMER STATED THAT GUIDEWIRE COILS WHEN BEING INSERTED INTO PATIENT. SHE STATED THIS HAS HAPPENED SEVERAL TIMES, BUT DOES NOT KNOW HOW MANY TIMES OR SPECIFIC DATES. ADDITIONAL INFORMATION PROVIDED 10/30/2023: AFTER GUIDEWIRE COILS, IT MAY OR MAY NOT BE ABLE TO BE USED; MULTIPLE OCCASIONS OF OPENING A SECOND KIT TO REPLACE THE GUIDEWIRE WHICH CAUSES A DELAY OF ACCESS. EVENT DID NOT INVOLVE AN URGENT/LIFE THREATENING MEDICAL SITUATION. EXTENDED TIME FOR INSERTION DUE TO DIFFICULTY AND HAVING TO OPEN ANOTHER KIT. MULTIPLE KITS HAD TO BE USED. NO PATIENTS HARMED. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18175572 ↗MalfunctionNov 20, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT GUIDEWIRE COILS WHEN BEING INSERTED INTO PATIENT. ADDITIONAL INFORMATION PROVIDED 10/30/2023: AFTER GUIDEWIRE COILS, IT MAY OR MAY NOT BE ABLE TO BE USED; OPENING A SECOND KIT TO REPLACE THE GUIDEWIRE WHICH CAUSES A DELAY OF ACCESS. THIS DID NOT CAUSE AN URGENT SITUATION, BUT EXTENDED TIME FOR INSERTION DUE TO DIFFICULTY AND HAVING TO OPEN ANOTHER KIT. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18173890 ↗MalfunctionAug 31, 2021Material Deformation; Material Deformation
IT WAS REPORTED BY HEALTHCARE PROFESSIONAL "OUR INFUSION NURSE WAS INSERTING A PICC THIS MORNING THAT HAD A FAULTY INTRODUCER; THE END IS ROLLED JUST SLIGHTLY. THEY HAD TO OPEN ANOTHER KIT TO COMPLETE PROCEDURE." NO OTHER INFORMATION WAS PROVIDED. (B)(6) 2021: DURING EVALUATION THE RETURNED DEVICE WAS NOTED TO BE CRUMPLED AT THE DISTAL EDGE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12392011 ↗MalfunctionJul 30, 2021Material Deformation; Material Deformation
IT WAS REPORTED A FAULTY INTRODUCER. NO OTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12254993 ↗InjuryMar 5, 2021Material Frayed; Material Frayed
IT WAS REPORTED VIA MEDWATCH "ATTEMPTED TO INSERT PICC LINE IN LUE, ENCOUNTERED SOME RESISTANCE ON THE SUBCLAVIAN REGION WHEN ATTEMPTING TO ADVANCE LINE. PULLED LINE BACK APPROXIMATELY 10 CM, FLUSHED THE LINE AMD ATTEMPTED TO ADVANCE. WHEN THAT ATTEMPT FAILED, THE LINE WAS REMOVED AND PICC WAS PLACED IN THE RIGHT ARM. CHEST X-RAY WAS DONE AND A FOREIGN BODY, PRESUMABLY A WIRE MEASURING ABOUT 3.8CM WAS DETECTED IN THE LEFT SUBCLAVIAN REGION. THE PATIENT HAD BEEN DISCHARGED. THE PATIENT WAS INSTRUCTED TO RETURN TO THE HOSPITAL IMMEDIATELY. THE FOREIGN OBJECT WAS REMOVED ON (B)(6) 2020. "
Patient problems: Foreign Body In Patient
MAUDE report 11424112 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page