Skip to content

PowerPICC® HF Catheter 3385355QRecalls & reported problems

Bard Access Systems, Inc. · STC PPICC, 5Fr TL, IR · Class 2 · Implant

All models of PowerPICC HF Catheter (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741028205), received between Jun 1, 2016 and Feb 12, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Hole In Material3
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Flush1
  • Failure to Obtain Sample1
  • Kinked1
  • Material Puncture/Hole1
  • Obstruction of Flow1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Fungal Infection1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Swelling/ Edema1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED, "CNS AT BEDSIDE WITH RN TO DO DRESSING CHANGE. WHEN NURSE WENT TO TURN PATIENT, SHE LIFTED PATIENT'S LEFT ARM AND NOTED DISCOLORATION AND EDEMA BELOW THE PATIENT'S PICC, NO OTHER LINES PRESENT. PROVIDER WAS CALLED TO BEDSIDE. ALL INFUSIONS WERE STOPPED, CLAMPED. PLASTICS CONSULT PLACED STAT. CONSULTING SURGEON CAME TO BEDSIDE. ADVISED RN NOT TO APPLY ANY COMPRESS OR ADMINISTER ANY ANTIDOTE, DUE TO MULTIPLE INFUSIONS AND DISCOLORATION ALREADY APPEARING. SERVICE WILL FOLLOW PATIENT TO ASSESS FOR ANY WOUND EVOLVEMENT. SUSPECTING PICC PRODUCT FAILURE, DUE TO PICC PLACEMENT CONFIRMED ON MORNING CHEST X-RAY." ADDITIONAL INFORMATION PROVIDED: PATIENT IMPACT: TEMPORARY PHYSICAL HARM, AS WELL AS AN EXTRA XRAY TO DETERMINE THE PICC HAD FAILED. PATIENT PRESENTED ODDLY WITH WHAT APPEARED TO BE AN EXTRAVASATION. AFTER FURTHER DIAGNOSTIC WORKUP, A DVT WAS DISCOVERED, ALONG WITH AN ANGIO-INVASIVE FUNGAL INFECTION OF THE SOFT TISSUE. WE NOW DO NOT BELIEVE THERE WAS AN EXTRAVASATION PRESENT. MEDICATION USED TO TREAT EDEMA INCLUDES MULTIPLE MEDICATIONS INCLUDING VASOACTIVES, TPN, AND ANTIBIOTICS. CATHETER WAS NOT REPLACED. PATIENT IS NOT TO HAVE ANY IV PLACED IN THE LUE THAT NOW REQ

    Patient problems: Fungal Infection; Swelling/ Edema; Thrombosis/Thrombus

    MAUDE report 21365874 ↗
  • MalfunctionApr 5, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "AFTER INSERTION OF TRIPLE LUMEN PICC, ASPIRATED AND FLUSHED LINE AND AIR WAS PULLED BACK." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16686257 ↗
  • MalfunctionMay 10, 2019Material Puncture/Hole

    IT WAS REPORTED THAT A HOLE WAS DISCOVERED IN THE PICC IN THE CT PORT AFTER THE PATIENT HAD A CT SCAN. ON (B)(6) 2019 - ADDITIONAL INFORMATION RECEIVED STATED, "A TRIPLE LUMEN PICC LINE WAS EXCHANGED ON THIS DATE DUE TO THE MIDDLE, CT APPROVED PORT SHOWED A SIZABLE HOLE IN THE CATHETER AFTER A CT INJECTION OF CONTRAST. THE ORIGINAL PICC LINE WAS PLACED ON (B)(6) 2019. THE PATIENT WAS NOT HARMED, HOWEVER, REQUIRED A NEW LINE TO BE PLACED.".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8600055 ↗
  • MalfunctionJul 1, 2016Fluid/Blood Leak; Hole In Material; Kinked

    PER SALES REPRESENTATIVE, NURSE REPORTED THAT THEIR VASCULAR ACCESS TEAM WAS CALLED TO CONSULT ON A PATIENT REGARDING DIFFICULTY WITH PICCS. THEY COULD NOT GET BLOOD RETURN AND HAD DIFFICULTY FLUSHING. TEAM CHECKED FOR MECHANICAL OBSTRUCTION, FLUSHED THE PICC, AND NOTICED SALINE LEAKING FROM THE INSERTION SITE. THE TEAM RETRACTED THE PICC A LITTLE MORE, FLUSHED TWICE AND WAS ABLE TO IDENTIFY A KINK AND PUNCTURE TO CATHETER. PICC WAS REMOVED AT THE TIME THIS WAS IDENTIFIED. NO PATIENT HARM REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5765561 ↗
  • MalfunctionJun 1, 2016Fluid/Blood Leak; Difficult to Flush; Hole In Material; Obstruction of Flow; Failure to Obtain Sample

    ON 05/12/2016- PER SALES REPRESENTATIVE USER FACILITY REPORTED THAT AFTER A PICC WAS PLACED THEY WERE UNABLE TO DRAW BLOOD AND HAD DIFFICULTY FLUSHING. IT WAS SAID THAT THEY CHECKED FOR A MECHANICAL OBSTRUCTION, FLUSHED THE PICC AND THEN NOTICED SALINE LEAKING FROM INSERTION SITE. THE PICC WAS FURTHER RETRACTED, FLUSHED TWICE AND REPORTEDLY A PUNCTURE WAS NOTED. THE PICC WAS REMOVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5693058 ↗
  • MalfunctionJun 1, 2016Fluid/Blood Leak; Hole In Material

    ON (B)(6) 2016- PER SALES REPRESENTATIVE, FACILITY REPORTED THAT AFTER A PICC WAS PLACED THEY WERE UNABLE TO DRAW BLOOD AND HAD DIFFICULTY FLUSHING. IT WAS SAID THAT THEY CHECKED FOR A MECHANICAL OBSTRUCTION, FLUSHED THE PICC AND THEN NOTICED SALINE LEAKING FROM INSERTION SITE. THE PICC WAS FURTHER RETRACTED, FLUSHED TWICE AND REPORTEDLY A PUNCTURE WAS NOTED. THE PICC WAS REMOVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5693057 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page