PowerPICC® SV Catheter with Sherlock Tip Location System (TLS) Stylet 9173108Recalls & reported problems
Bard Access Systems, Inc. · DOT PPICC SV 3Fr SL TLS · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741028564), received between Apr 26, 2017 and Dec 11, 2024 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Device Contamination with Chemical or Other Material2
- Material Deformation2
- Break1
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2024Material Deformation; Material Deformation
IT WAS REPORTED SHERLOCK STYLET IS KINKING AND SEE THAT DURING REMOVAL OF STYLET AFTER INSERTION. NO PATIENT HARM REPORTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20908734 ↗MalfunctionNov 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER PICC LEAKING/RUPTURING IN THE AREA WHERE THE CATHETER CONNECTS TO THE SUTURE WING. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15756954 ↗MalfunctionNov 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER PICC LEAKING/RUPTURING WHERE THE EXTENSION LEG CONNECTS TO THE HUB. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15756315 ↗MalfunctionNov 4, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER "PICCS ARE LEAKING AROUND THE SUTURE WING AND THE HUB."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15736767 ↗MalfunctionOct 19, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "INTRODUCER SHEATH WHERE THE SHEATH MEETS THE DILATOR IS ROLLING UP UPON ATTEMPTED ENTRY, CAUSING INABILITY TO ENTER VESSEL"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15633303 ↗InjuryApr 20, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY HEALTH PROFESSIONAL "PATIENT WITH LINE AT HOME FOR TREATMENT. PATIENT INFORMED MOTHER ¿ARM FELT WET¿, MOTHER INITIALLY THOUGHT CAP LOOSE, SO SHE TIGHTENED MORE. A FEW MINUTES LATER, PATIENT INFORMED MOM, STILL WET. MOTHER OBSERVED LEAKING AT DISTAL END OF TUBING WHERE CAP ATTACHES. MOTHER THEN CALLED CLINIC WHO INSTRUCTED MOTHER TO GO TO ER. MILRINONE STILL INFUSING. UPON EXAM IN ER, THIS RN THEN OBSERVED LEAKING AT STATED SITE. DISCUSSED WITH MEDICAL TEAM. PICC WAS LOCKED OFF, PIV STARTED FOR MILRINONE, AND PATIENT ADMITTED FOR PICC REPLACEMENT THE NEXT DAY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11694119 ↗MalfunctionOct 1, 2020Break; Break
IT WAS REPORTED THAT THE SV PICC LINE WAS PLACED ON (B)(6) 2020 AND BROKE ON (B)(6) 2020. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10615104 ↗MalfunctionJul 10, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED HOME HEALTH CAME TO PT'S HOUSE AND PICC WOULD FLUSH BUT NOT DRAW BLOOD. IT WAS STATED THE PATIENT'S SHIRT WAS WET SO BROUGHT HIM TO THE HOSPITAL VIA THE ER ON 6/16 AND THERE WAS A HOLE IDENTIFIED IN THE CATHETER WHERE THE HUB AND EXTENSION LEG MEET. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10259008 ↗MalfunctionDec 5, 2019Crack; Crack
IT WAS REPORTED THAT THE BLUE PORT CRACKED AT THE CONNECTION SITE. ON 11/26/2019-ADDITIONAL INFORMATION: "THE LINE WAS DC'D FROM THE PATIENT AND THROWN AWAY. THERE WAS NO KNOWN PATIENT HARM. NO KNOWN INFECTIOUS DISEASE. THE CONNECTION AT THE LINE AND THE BLUE CAP WAS NOTED TO BE CRACKED AND LEAKING."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9417907 ↗InjuryMay 1, 2018Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THE PICC LINE WAS PLACED IN 2018. IT WAS STATED THAT A METALLIC STRUCTURE WHICH PROJECTS OVER THE RIGHT PROXIMAL FEMUR WAS IDENTIFIED DURING A KUB SCAN. USING ULTRASOUND, THE LINEAR HYPOECHOIC STRUCTURE WAS CONFIRMED IN THE ANTERIOR MUSCULATURE OF THE PROXIMAL RIGHT THIGH, ASSUMED TO REPRESENT A FOREIGN BODY 1.3CM LONG AND 5-9MM DEEP INTO THE SKIN. DOPPLER CONFIRMED THE FOREIGN BODY TO BE EXTRAVASCULAR.
Patient problems: Foreign Body In Patient
MAUDE report 7478339 ↗MalfunctionApr 26, 2017Device Contamination with Chemical or Other Material
THE FACILITY REPORTED TO THE SALES REP THAT THE TRAY WAS OPENED IN THE O.R., AS THE PRODUCTS WERE BEING REMOVED FROM THE TRAY IT WAS VISUALIZED THAT A HAIR WAS PACKAGED BETWEEN THE BLUE WRAP AND THE TRAY. THE KIT WAS DISCARDED AND A NEW KIT WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6524723 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page