Bardex® All-Silicone Foley Catheter 165822Recalls & reported problems
C. R. Bard, Inc. · Bardex® All-Silicone Foley Catheter 5cc Balloon · Class 2
All models of Bardex All-Silicone Foley Catheter (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741029783), received between Jan 30, 2018 and Oct 30, 2025 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Material Protrusion/Extrusion3
- Material Puncture/Hole2
- Nonstandard Device2
- Burst Container or Vessel1
- Component Falling1
- Deflation Problem1
- Difficult to Open or Close1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Skin Inflammation/ Irritation2
- No Information1
- Pain1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT THE CUSTOMER IS SEEKING FULL CREDIT ON AN ORDER OF FOLEY CATHETERS DUE TO MANUFACTURING DEFECTS. THEY HAVE ONLY RECENTLY DISCOVERED. DUE TO THE NATURE OF CATHETER PACKAGING, IT IS IMPOSSIBLE TO INSPECT THE CATHETERS UPON RECEIPT TO IDENTIFY SUCH DEFECTS WHILE MAINTAINING THEIR STERILITY. ON ORDER (B)(4) FROM (B)(6) 2024, THE CUSTOMER RECEIVED TWO CASES OF BARDEX FOLEY CATHETER, PART 165822, FROM WHAT APPEARS TO BE A DEFECTIVE LOT. ROUGHLY 7 OF THE 12 CATHETERS HAD EITHER SURFACE DEFECTS (BURRS) OR HAD LOOSE FLAPS OF SILICONE STUCK TO THEM. FIVE OF THESE CATHETERS CAUSED MINOR INTERNAL TRAUMA OR IRRITATION, IN THREE CASES PRODUCING BLOOD UPON REMOVAL. THEY DID NOT USE THE LAST TWO CATHETERS FROM THIS CASE, AND INSTEAD THEY OPENED THEM UP AND FOUND ONE WITH A LOOSE FLAP OF SILICONE (SEE PHOTO). THE DEFECTIVE LOT IS NGJV0914. THE CUSTOMER HAS ONE REMAINING CASE OF 12 CATHETERS FROM THIS LOT. THE CUSTOMER CANNOT USE IT DUE TO THE SUBSTANTIAL MANUFACTURING DEFECTS IN ITS SIBLING CASE. THE PATIENT WISHES TO RETURN THIS CASE TO YOU FOR FULL CREDIT AND SO YOU CAN FORWARD IT TO BARD FOR ANY INVESTIGATION THEY CARE TO PERFORM. DEFECTS IN CATHETERS ARE IMPORTANT ENOUGH TH
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Skin Inflammation/ Irritation
MAUDE report 23427698 ↗MalfunctionOct 30, 2025Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THE CUSTOMER IS SEEKING FULL CREDIT ON AN ORDER OF FOLEY CATHETERS DUE TO MANUFACTURING DEFECTS. THEY HAVE ONLY RECENTLY DISCOVERED. DUE TO THE NATURE OF CATHETER PACKAGING, IT IS IMPOSSIBLE TO INSPECT THE CATHETERS UPON RECEIPT TO IDENTIFY SUCH DEFECTS WHILE MAINTAINING THEIR STERILITY. ON ORDER (B)(4) FROM (B)(6) 2024, THE CUSTOMER RECEIVED TWO CASES OF BARDEX FOLEY CATHETER, PART 165822, FROM WHAT APPEARS TO BE A DEFECTIVE LOT. ROUGHLY 7 OF THE 12 CATHETERS HAD EITHER SURFACE DEFECTS (BURRS) OR HAD LOOSE FLAPS OF SILICONE STUCK TO THEM. FIVE OF THESE CATHETERS CAUSED MINOR INTERNAL TRAUMA OR IRRITATION, IN THREE CASES PRODUCING BLOOD UPON REMOVAL. THEY DID NOT USE THE LAST TWO CATHETERS FROM THIS CASE, AND INSTEAD THEY OPENED THEM UP AND FOUND ONE WITH A LOOSE FLAP OF SILICONE. THE DEFECTIVE LOT IS NGJV0914. THE CUSTOMER HAS ONE REMAINING CASE OF 12 CATHETERS FROM THIS LOT. THE CUSTOMER CANNOT USE IT DUE TO THE SUBSTANTIAL MANUFACTURING DEFECTS IN ITS SIBLING CASE. THE PATIENT WISHES TO RETURN THIS CASE TO YOU FOR FULL CREDIT AND SO YOU CAN FORWARD IT TO BARD FOR ANY INVESTIGATION THEY CARE TO PERFORM. DEFECTS IN CATHETERS ARE IMPORTANT ENOUGH THAT THE FDA T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23427663 ↗MalfunctionOct 30, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT THE PT EMAIL SENT TO CHAH READS I AM SEEKING FULL CREDIT ON AN ORDER OF FOLEY CATHETERS DUE TO MANUFACTURING DEFECTS I HAVE ONLY RECENTLY DISCOVERED. DUE TO THE NATURE OF CATHETER PACKAGING, IT IS IMPOSSIBLE TO INSPECT THE CATHETERS UPON RECEIPT TO IDENTIFY SUCH DEFECTS WHILE MAINTAINING THEIR STERILITY. ON MY ORDER (B)(4) FROM (B)(6)2024, I RECEIVED TWO CASES OF BARDEX FOLEY CATHETER, PART 165822, FROM WHAT APPEARS TO BE A DEFECTIVE LOT. ROUGHLY 7 OF THE 12 CATHETERS HAD EITHER SURFACE DEFECTS (BURRS) (SEE PHOTO) OR HAD LOOSE FLAPS OF SILICONE STUCK TO THEM. FIVE OF THESE CATHETERS CAUSED MINOR INTERNAL TRAUMA OR IRRITATION, IN THREE CASES PRODUCING BLOOD UPON REMOVAL. I DID NOT USE THE LAST TWO CATHS FROM THIS CASE, AND INSTEAD I OPENED THEM UP AND FOUND ONE WITH A LOOSE FLAP OF SILICONE (SEE PHOTO) THE DEFECTIVE LOT IS NGJV0914 (SEE PHOTO). I HAVE ONE REMAINING CASE OF 12 CATHETERS FROM THIS LOT. I CANNOT USE IT DUE TO THE SUBSTANTIAL MANUFACTURING DEFECTS IN ITS SIBLING CASE. I WISH TO RETURN THIS CASE TO YOU FOR FULL CREDIT AND SO YOU CAN FORWARD IT TO BARD FOR ANY INVESTIGATION THEY CARE TO PERFORM. DEFECTS IN CATHETERS ARE IMPORTANT ENOUGH THAT THE FDA T
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Skin Inflammation/ Irritation
MAUDE report 23426687 ↗MalfunctionOct 30, 2025Nonstandard Device; Material Protrusion/Extrusion; Nonstandard Device; Material Protrusion/Extrusion
IT WAS REPORTED THAT THE CUSTOMER IS SEEKING FULL CREDIT ON AN ORDER OF FOLEY CATHETERS DUE TO MANUFACTURING DEFECTS. THEY HAVE ONLY RECENTLY DISCOVERED. DUE TO THE NATURE OF CATHETER PACKAGING, IT IS IMPOSSIBLE TO INSPECT THE CATHETERS UPON RECEIPT TO IDENTIFY SUCH DEFECTS WHILE MAINTAINING THEIR STERILITY. ON ORDER (B)(4) FROM (B)(6) 2024, THE CUSTOMER RECEIVED TWO CASES OF BARDEX FOLEY CATHETER, PART 165822, FROM WHAT APPEARS TO BE A DEFECTIVE LOT. ROUGHLY 7 OF THE 12 CATHETERS HAD EITHER SURFACE DEFECTS (BURRS) OR HAD LOOSE FLAPS OF SILICONE STUCK TO THEM. FIVE OF THESE CATHETERS CAUSED MINOR INTERNAL TRAUMA OR IRRITATION, IN THREE CASES PRODUCING BLOOD UPON REMOVAL. THEY DID NOT USE THE LAST TWO CATHETERS FROM THIS CASE, AND INSTEAD THEY OPENED THEM UP AND FOUND ONE WITH A LOOSE FLAP OF SILICONE. THE DEFECTIVE LOT IS NGJV0914. THE CUSTOMER HAS ONE REMAINING CASE OF 12 CATHETERS FROM THIS LOT. THE CUSTOMER CANNOT USE IT DUE TO THE SUBSTANTIAL MANUFACTURING DEFECTS IN ITS SIBLING CASE. THE PATIENT WISHES TO RETURN THIS CASE TO YOU FOR FULL CREDIT AND SO YOU CAN FORWARD IT TO BARD FOR ANY INVESTIGATION THEY CARE TO PERFORM. DEFECTS IN CATHETERS ARE IMPORTANT ENOUGH THAT THE FDA T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426682 ↗InjuryMay 22, 2025Material Too Soft/Flexible; Material Too Soft/Flexible
IT WAS REPORTED THAT THE FOLEY CATHETER MATERIAL WAS DIFFERENT (SOFTER) WHICH INCREASED FALSE PATHS, WAS MORE DIFFICULT TO INSERT, WHICH HURT THE PATIENT. CUSTOMER HAD A SITUATION WHERE THE CATHETER HAD MADE A KNOT IN THE BLADDER, AND THEY HAD TO DO A CYSTOSCOPY TO CUT THE CATHETER WITH A LASER TO REMOVE IT (PCN#: 175812, PCN#: 175814, PCN#: 175816, PCN#: 175818, PCN#: 175820, PCN#: 165822 AND PCN#: 175824). PER ADDITIONAL INFORMATION RECEIVED VIA MAIL ON 06MAY2025, THE TIP OF THE CATHETER THAT CONTAINS THE BALLOON IS PROMINENT AND FORMS A RING WHEN THE BALLOON IS DEFLATED. IT WAS ANOTHER ISSUE WITH THE DEFLATION OF THE BALLOON. THE BALLOON REMAINS DEFORMED WHEN DEFLATED, CAUSING INJURY TO THE PATIENT WHEN REMOVING THE INDWELLING CATHETER. THIS WAS APPARENTLY TRUE FOR ALL OUR 100% FOLEY (1658XX) AND LUBRI-SIL (1758XX) MODELS. (PCN#: 165820, PCN#: 165818 AND PCN#: 165816).
Patient problems: Pain
MAUDE report 22069846 ↗MalfunctionJul 25, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THERE IS A PINHOLE IN THE BALLOON DUE TO THAT STERILE WATER LEAKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19824566 ↗MalfunctionJun 6, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE BALLOON ON THE SILICON FOLEY CATHETER WAS LEAKING AND BELIEVED THAT THE ITEMS WERE DEFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19481071 ↗MalfunctionMar 14, 2019Device Dislodged or Dislocated
IT WAS REPORTED THAT THE CATHETER DISLODGED AND A NEW ONE HAD TO BE PLACED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8420087 ↗MalfunctionMar 13, 2019Burst Container or Vessel; Device Dislodged or Dislocated; Burst Container or Vessel
IT WAS REPORTED THAT THE CATHETER DISLODGED AND A NEW ONE HAD TO BE PLACED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8416477 ↗MalfunctionMar 13, 2019Device Dislodged or Dislocated
IT WAS REPORTED THAT THE CATHETER DISLODGED AND A NEW ONE HAD TO BE PLACED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8416097 ↗MalfunctionMar 13, 2018Deflation Problem; Component Falling; Deflation Problem; Leak/Splash
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT DUE TO A DEFLATED BALLOON. WATER LEAKAGE WAS FOUND FROM THE LUMEN.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7335100 ↗MalfunctionJan 30, 2018Difficult to Open or Close
IT WAS REPORTED THAT THE STATLOCK DEVICE WOULD NOT STAY CLOSED OVER THE 22FR SILICONE CATHETERS THE PATIENT USED. THE PATIENT REPORTED THAT THIS CAUSED THE STATLOCK DEVICE TO ROTATE DURING USE AND TWIST THE CATHETER, PREVENTING URINE FLOW.
Patient problems: No Information
MAUDE report 7229012 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page