PowerPICC Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 1275118Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC Catheter with Sherlock 3CG (TPS) Stylet 5F Basic Tray with Microintroducer · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bard PowerPICC catheter, a long-term central line placed through a vein, comes with a Sherlock tip-positioning stylet and dilators in the insertion kit. The 48 reports mostly describe parts not working as intended: dilators buckling at the tip instead of dilating the vein, a blunt trimming razor, caps stuck on the needleless connector, and leaking lock assemblies. Reports include 2 deaths and 4 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 2
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 00801741034589), received between Oct 27, 2017 and Sep 14, 2026 — classified as 2 deaths, 4 injuries, 43 malfunctions.
The most frequently reported specific device problem is “Defective Component” (27 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component27
- Material Deformation9
- Detachment Of Device Component2
- Fluid/Blood Leak2
- Break1
- Device Markings/Labelling Problem1
- Dull, Blunt1
- Expiration Date Error1
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient2
- No Known Impact Or Consequence To Patient2
- Anxiety1
- Death1
- No Consequences Or Impact To Patient1
- Swelling1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50765659 ↗MalfunctionFeb 18, 2026Material Deformation
IT WAS REPORTED THE NEEDLELESS CONNECTOR OF THE PURPLE LUMEN OF A DL PICC WAS NOT BEING ABLE TO BE REMOVED. THE PATIENT REPORTED ONGOING ISSUES WITH THE PURPLE LUMEN CAP GETTING "STUCK" ON IT. WHEN I ASSESSED, THERE WAS CLEAR INDICATION THAT SOMEONE HAD APPLIED A CLAMP TO TRY TO REMOVE THE NEEDLELESS CONNECTOR AS THERE WERE SCORE MARKS ON THE HUB. ULTIMATELY, THE TEAM OPTED TO REMOVE THIS PICC FROM THE PATIENT, BOTH DUE TO THE DAMAGE ON THE HUB AND THE FACT THAT THE NEEDLELESS CONNECTOR WOULD NOT COME LOOSE. THE PICC WAS REPLACED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378812 ↗MalfunctionAug 15, 2025Dull, Blunt; Failure to Cut
WHEN RN WENT TO TRIM THE PICC [PERIPHERALLY INSERTED CENTRAL CATHETER] LINE WITH THE STANDARD GREEN BARD RAZOR THAT COMES IN THE KIT USED TO TRIM THE CATHETER, THE RAZOR DID NOT TRIM THE CATHETER AFTER MULTIPLE ATTEMPTS. THE RAZOR WAS BLUNTED AND A SUTURE REMOVAL KIT AND STERILE SCISSORS WAS USED TO TRIM THE CATHETER TO THE LENGTH NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22814237 ↗MalfunctionAug 1, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT THE "BUCKLING" WAS OCCURRING QUITE FREQUENTLY, AND THEY HAD TO ADAPT THEIR DILATION TECHNIQUE TO ACCOMMODATE FOR IT (LARGER NICK IN THE SKIN, STRETCH AND PAUSE BEFORE CONTINUING, AND HOLDING MUCH CLOSER TO HUB OF DILATOR). MOST OFTEN, THE DILATOR WOULD JUST BUCKLE AND NOT CREATE THE ROUGH EDGE, BUT SOMETIMES IT CREATED THE ROUGH EDGE MAKING IT UNUSABLE. THE DILATORS ARE UNUSABLE AND HAD TO BE REPLACED MID-PROCEDURE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22688244 ↗MalfunctionAug 1, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT THE "BUCKLING" WAS OCCURRING QUITE FREQUENTLY, AND THEY HAD TO ADAPT THEIR DILATION TECHNIQUE TO ACCOMMODATE FOR IT (LARGER NICK IN THE SKIN, STRETCH AND PAUSE BEFORE CONTINUING, AND HOLDING MUCH CLOSER TO HUB OF DILATOR). MOST OFTEN, THE DILATOR WOULD JUST BUCKLE AND NOT CREATE THE ROUGH EDGE, BUT SOMETIMES IT CREATED THE ROUGH EDGE MAKING IT UNUSABLE. THE DILATORS ARE UNUSABLE AND HAD TO BE REPLACED MID-PROCEDURE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22688227 ↗MalfunctionJul 10, 2025Material Deformation; Material Deformation
IT WAS REPORTED "I WAS INSERTING A DUAL LUMEN BARD POWER PICC. WHEN I USED THE DILATOR INCLUDED IN THE PICC KIT, THE EDGE OF THE EXTERNAL SHEATH "BUCKLED" AT THE TIP CREATING A ROUGH EDGE, AND NOT USABLE TO DILATE THE VEIN. A NEW DILATOR WAS USED TO DILATE THE VEIN." FAULTY PRODUCT DID NOT REACH THE PATIENT, NO HARM WAS REPORTED. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22468132 ↗MalfunctionJul 10, 2025Material Deformation; Material Deformation
IT WAS REPORTED THE QUALITY OF THE DILATOR IN THE KITS CHANGED . IT IS NO LONGER AS "STIFF" AS IT ORIGINALLY WAS. THIS "BUCKLING" WAS OCCURRING QUITE FREQUENTLY, AND WE HAD TO ADAPT OUR DILATION TECHNIQUE TO ACCOMMODATE FOR IT (LARGER NICK IN THE SKIN, STRETCH AND PAUSE BEFORE CONTINUING, AND HOLDING MUCH CLOSER TO HUB OF DILATOR). MOST OFTEN, THE DILATOR WOULD JUST BUCKLE AND NOT CREATE THE ROUGH EDGE, BUT SOMETIMES IT CREATED THE ROUGH EDGE MAKING IT UNUSABLE. NO OTHER INFORMATION WAS PROVIDED. A SPECIFIC NUMBER OF AFFECTED DEVICES OR PATIENTS WAS NOT PROVIDED BY THE COMPLAINANT. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22468114 ↗MalfunctionMay 9, 2024Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY. DEVICE WAS USED ON ONE PATIENT. NO HARM REPORTED. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION. THE LEAK WAS DISCOVERED WHEN I WAS FLUSHING THE LINE DURING INSERTION, THE EVENT DID NOT INTERRUPT TREATMENT AND NO SIGNS, SYMPTOMS OR COMPLICATIONS EXPERIENCE BY PATIENT. WE SECURED THE DEVICE WITH A STAT LOCK. THERE WAS NO PATIENT HARM NOTED DUE TO MY INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19281152 ↗MalfunctionFeb 5, 2024Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER ¿I HAVE A DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY THAT WAS USED ON A PATIENT. OCCURRED ON JANUARY 12, 2024. THERE WAS NO PATIENT HARM AND IT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION OR A DELAY IN TREATMENT. THE LEAK WAS DISCOVERED WHEN I PRIMED THE LINE. NO SIGNS, SYMPTOMS, OR COMPLICATIONS EXPERIENCED BY THE PATIENT. I DO NOT HAVE THE NAME OF THE PLANNED MEDICATIONS TO BE INFUSED. WE SECURED THE DEVICE WITH A STATLOCK. NO HARM NOTED DUE TO MY INTERVENTION.¿ NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18643850 ↗MalfunctionDec 13, 2023Defective Component; Defective Component
IT WAS REPORTED THAT CATHETER THAT LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18320648 ↗MalfunctionOct 24, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT HAVE A DOUBLE LUMEN PICC LINE WITH A LEAKING T LOCK ASSEMBLY. THE PRODUCT WAS USED ON A PATIENT. NO PATIENT HARM REPORTED, NO SIGNS, SYMPTOMS OR COMPLICATIONS EXPERIENCE BY PATIENT. EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION. THE LEAK WAS DISCOVERED WHEN I PRIMED THE LINE. THE EVENT DID NOT INTERRUPT TREATMENT. I DO NOT HAVE THE NAME OF PLANNED MEDICATIONS TO BE INFUSED. WE SECURED THE DEVICE WITH A STAT LOCK. NO OTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17995690 ↗MalfunctionSep 26, 2023Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT A CATHETER LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. NO PATIENT HARM REPORTED. NO OTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17821278 ↗MalfunctionSep 8, 2023Defective Component; Defective Component
CUSTOMER REPORTED CATHETER THAT LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. ADDITIONAL INFORMATION PROVIDED (B)(6): THE CATHETER DID NOT BREAK. EVENT DID NOT INTERRUPT TREATMENT OR CAUSE CLINICALLY SIGNIFICANT DELAY TO TREATMENT. NO OTHER INFORMATION WAS PROVIDED. THERE WERE TWO SAMPLES RETURNED FOR INVESTIGATION, THIS FILE REPRESENTS THE FIRST RETURNED SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17708742 ↗MalfunctionSep 7, 2023Defective Component; Defective Component
CUSTOMER REPORTED CATHETER THAT LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. ADDITIONAL INFORMATION PROVIDED JULY 5: THE CATHETER DID NOT BREAK. EVENT DID NOT INTERRUPT TREATMENT OR CAUSE CLINICALLY SIGNIFICANT DELAY TO TREATMENT. NO OTHER INFORMATION WAS PROVIDED. THERE WERE TWO SAMPLES RETURNED FOR INVESTIGATION, THIS FILE REPRESENTS THE FIRST RETURNED SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17703485 ↗MalfunctionSep 7, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAVE A DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION. THE LEAK WAS DISCOVERED WHEN CUSTOMER PRIMED THE LINE. THE EVENT DID NOT INTERRUPT TREATMENT AND NO SIGNS, SYMPTOMS, HARM OR COMPLICATIONS EXPERIENCE BY PATIENT. CUSTOMER SECURED THE DEVICE WITH A STAT LOCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17700993 ↗MalfunctionSep 7, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAVE A DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY. THE INCIDENT OCCURRED (B)(6) 2023 WHEN THE PRODUCT WAS USED ON A PATIENT. THE LEAK WAS DISCOVERED WHEN THE LINE WAS PRIMED. THE DEVICE WAS SECURED WITH A STATLOCK. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE THREATENING SITUATION AND THERE WAS NO HARM REPORTED. NO TREATMENT WAS INTERRUPTED, NO COMPLICATIONS OR SYMPTOMS EXPERIENCED BY THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17697897 ↗MalfunctionSep 1, 2023Defective Component; Defective Component
IT WAS REPORTED THAT ANOTHER DOUBLE LUMEN PICC LINE WITH A LEAKING T LOCK ASSEMBLY, THE LEAK WAS DISCOVERED WHEN WE PRIMED THE LINE. THE EVENT DID NOT INTERRUPT TREATMENT. SECURED THE DEVICE WITH A STAT LOCK. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION. THERE WAS NO PATIENT HARM NOTED DUE TO INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17673621 ↗MalfunctionAug 28, 2023Defective Component; Defective Component
IT WAS REPORTED THAT DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY. IT WAS USED ON ONE PATIENT. OCCURRED (B)(6) 2023. THE PRODUCT WAS USED ON A PATIENT. NO HARM REPORTED. NO PATIENT INFORMATION AVAILABLE. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING SITUATION. THE LEAK WAS DISCOVERED WHEN THEY PRIMED THE LINE. THE EVENT DID NOT INTERRUPT TREATMENT. NO SIGNS, SYMPTOMS OR COMPLICATIONS EXPERIENCE BY PATIENT. THEY DO NOT HAVE THE NAME OF PLANNED MEDICATIONS TO BE INFUSED. THEY SECURED THE DEVICE WITH A STAT LOCK. THERE WAS NO PATIENT HARM NOTED DUE TO THEIR INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17638808 ↗InjuryAug 14, 2023Material Frayed; Material Frayed
IT WAS REPORTED BY CUSTOMER THAT AFTER ACCESS DURING A PICC PROCEDURE, NURSE FELT RESISTANCE WHEN ADVANCING THE GUIDEWIRE INTO THE PATIENT¿S VESSEL. THE NURSE WITHDREW THE GUIDEWIRE AND LEFT ACCESS NEEDLE IN PLACE AND REASSESSED VESSEL. AFTER VESSEL ASSESSMENT, NURSE REINSERTED GUIDEWIRE INTO PATIENT¿S VESSEL AND CONTINUED TO FEEL RESISTANCE AND WITHDREW AGAIN. ABOUT 2 CM OF THE TIP OF THE GUIDEWIRE WAS DISLODGED INTO THE PATIENT¿S ANATOMY. WHEN EVALUATING GUIDEWIRE AFTER ATTEMPTED INSERTION, NURSE NOTICED THE TIP OF THE GUIDE WIRE UNRAVELED. NO OTHER INFORMATION PROVIDED. ADDITIONAL INFORMATION RECEIVED: NURSE WAS ADVISED TO REMOVE GUIDEWIRE AND NEEDLE AS A UNIT WHEN GUIDEWIRE NEEDS TO BE REMOVED FROM PATIENT PER IFU.
Patient problems: Foreign Body In Patient
MAUDE report 17547048 ↗MalfunctionAug 9, 2023Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT THE CATHETER LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. NO PATIENT HARM. RESPONSE RECEIVED 26 JULY 2023: THE EVENT DID NOT INVOLVE AN URGENT OR LIFE THREATENING MEDICAL SITUATION, AND THERE WAS NO INTERRUPTION IN TREATMENT. THE DEVICE WAS SECURED WITH A STATLOCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17492115 ↗MalfunctionAug 2, 2023Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT THE CATHETER LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. NO PATIENT HARM. NO OTHER INFORMATION PROVIDED. ADDITIONAL INFORMATION RECEIVED 26 JULY 2023: NO URGENT/LIFE THREATENING MEDICAL SITUATION. DID NOT INTERRUPT TREATMENT, OR CAUSE A CLINICALLY SIGNIFICANT DELAY IN TREATMENT, THAT NEGATIVELY IMPACTED THE PATIENT. STAT LOCK USED FOR CATHETER SECUREMENT. NO PATIENT HARM, INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17443403 ↗MalfunctionAug 2, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT "PICC TEAM DISCOVERED KINKED 5F DUAL LUMEN PICC WHEN X RAY OF PATIENTS ARM AFTER FLOOR NURSE COULD NOT FLUSH LINE". NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17439851 ↗MalfunctionAug 1, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAVE DL PICC LINE WITH A LEAKING T LOCK ASSEMBLY. IT WAS USED ON ONE PATIENT. THERE WAS NO HARM REPORTED. THE EVENT DID NOT INVOLVE AN URGENT OR LIFE THREATENING SITUATION. THE LEAK WAS DISCOVERED WHEN THE CLINICIAN PRIMED THE LINE. THE EVENT DID NOT INTERRUPT TREATMENT. NO SIGNS, SYMPTOMS OR COMPLICATIONS EXPERIENCE BY PATIENT. CLINICIAN DO NOT HAVE THE NAME OF PLANNED MEDICATIONS TO BE INFUSED. CATHETER WAS SECURED THE DEVICE WITH A STATLOCK. THERE WAS NO PATIENT HARM NOTED DUE TO CLINICIAN INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17437135 ↗MalfunctionJul 31, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT CATHETER LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. ADDITIONAL INFORMATION RECEIVED ON 17 JULY 2023: LEAK DISCOVERED DURING INSERTION. CATHETER SECUREMENT WAS STAT LOCK. IT WAS STATED THE EVENT DID NOT INVOLVE AN URGENT OR LIFE THREATENING MEDICAL SITUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17424298 ↗MalfunctionJul 25, 2023Defective Component; Defective Component
IT WAS REPORTED BY CUSTOMER THAT THE CATHETER THAT LEAKED FROM THE RUBBER-STOPPER WHEN PICC LINE WAS BEING INSERTED. ADDITIONAL INFORMATION FROM CUSTOMER ON 07-12-2023. THIS EVENT DID NOT INVOLVE A LIFE THREATENING SITUATION. THE LEAK WAS DISCOVERED DURING INSERTION. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17389896 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page