Skip to content

PowerPICC SOLO2 HF Catheter 1395108QRecalls & reported problems

Bard Access Systems, Inc. · PowerPICC SOLO2 HF Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet Full Tray 5F Triple-Lumen · Class 2 · Implant

All models of PowerPICC SOLO2 HF Catheter (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741034718), received between Sep 9, 2016 and Sep 9, 2016 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Insert1
  • Material Fragmentation1
  • Physical Resistance1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2016Material Fragmentation; Difficult to Insert; Physical Resistance

    ON (B)(6) 2016 - FACILITY REPORTED THAT WHILE NURSE WAS PLACING A POWERPICC SOLO WITH THE SHERLOCK ULTRASOUND DEVICE, SHE MET RESISTANCE. NURSE REQUESTED ASSISTANCE FROM ANOTHER NURSE, STILL MET RESISTANCE AND CALLED FOR THE RADIOLOGIST. DEVICE WAS REMOVED UNDER CT SCAN. HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT A PIECE OF THE WIRE BROKE OFF AND REMAINED IN THE PATIENT. RADIOLOGIST WAS UNABLE TO RETRIEVE THE BROKEN PIECE OF WIRE. HCP ALLEGED THAT THE WIRE MAY HAVE UNRAVELED. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5939755 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page