PowerPICC SOLO2 Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 1396108Recalls & reported problems
Bard Access Systems, Inc. · PowerPICC SOLO Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 6F Full Tray · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741034732), received between Mar 15, 2017 and Aug 24, 2017 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Crack1
- Device Operates Differently Than Expected1
- Fluid/Blood Leak1
- Fracture1
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2017Crack
FACILITY REPORTED TO THE SALES REP THAT THE CATHETER APPEARS TO HAVE A CRACK AT THE HUB. NO OTHER INFORMATION WAS REPORTED, BUT HAS BEEN REQUESTED.
Patient problems: No Information
MAUDE report 6817066 ↗MalfunctionAug 10, 2017Device Operates Differently Than Expected
PER SALES REP, THE FACILITY REPORTED A TRIPLE LUMEN PICC WAS BEING PLACED AND ALL THREE LUMEN WERE FLUSHED WITH NO INCIDENT PRIOR TO PLACEMENT. PICC WAS PLACED WITHOUT INCIDENT, RN THEN CHECKED EACH LUMEN FOR ASPIRATION AND FLUSHED. IT WAS STATED THAT DURING FLUSHING THE WHITE SOLO VALVE "POPPED" AND SALINE/BLOOD SPLASHED OVER PLACERS GLOVES, GOWN, AND PATIENT DRAPE. THE LUMEN WAS CLAMPED WITH IV CLAMP. HOSPITALIST WAS NOTIFIED OF POTENTIAL FOR INTRODUCTION OF WHITE HUB/VALVE PARTICLES THROUGH THE LINE, AND THE DECISION WAS MADE THAT THERE WAS MINIMAL RISK, AND THE LINE WAS EXCHANGED OVER THE WIRE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6786084 ↗MalfunctionAug 10, 2017Break
PER SALES REP, THE FACILITY REPORTED A PICC WHITE LUMEN BROKE DURING PRE-PLACEMENT FLUSH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6784711 ↗MalfunctionMar 15, 2017Fluid/Blood Leak; Fracture; Material Fragmentation
ON 2/22/2017 - FACILITY REPORTED TO THE SALES REPRESENTATIVE THAT 2 HOURS POST PLACEMENT, THE CATHETER WAS FLUSHED AND THE HUB "CRUMBLED" IN THE NURSES HAND. DEVICE WAS REMOVED. NO PATIENT INJURY WAS REPORTED. ON 2/23/2017 - ADDITIONAL INFORMATION REPORTED TO THE SALES REP, HEALTHCARE PROFESSIONAL (HCP) PLACED THE PICC LINE AT 5:45 PM, USING 3 CG. HCP STATED THAT THE PLACEMENT WENT WELL AND WAS FUNCTIONING PROPERLY. THE NIGHT ONCOLOGY NURSE WENT TO ACCESS PATENCY OF THE CATHETER AT 8:10 PM, BEFORE INITIATING CHEMOTHERAPY. THE PATIENT STATED THAT BLOOD WAS DRAWN FROM THE FIRST 2 LUMENS AND FLUSHED WITHOUT DIFFICULTY. WHEN THE THIRD (WHITE) LUMEN WAS ACCESSED IT FELL APART IN THE HCP HANDS. THE HCP NOTED THAT THE VALVE AREA WAS COMPRESSED, THE HCP STATED THAT BLOOD WAS EVERYWHERE. THE LINE WAS REMOVED THE FOLLOWING MORNING. THE HCP NOTED THAT THE FRACTURED LINE WAS SECURED WITH A SLIDE CLAMP OFF OF A PERIPHERAL. ALLEGED FRACTURES WERE SEEN ON BOTH SIDES OF THE VALVE HOUSING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6408278 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page