PowerPICC SOLO2 Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 1396108Recalls & reported problems

Bard Access Systems, Inc. · PowerPICC SOLO Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet 6F Full Tray · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741034732), received between Mar 15, 2017 and Aug 24, 2017 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Crack1
  • Device Operates Differently Than Expected1
  • Fluid/Blood Leak1
  • Fracture1
  • Material Fragmentation1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2017Crack

    FACILITY REPORTED TO THE SALES REP THAT THE CATHETER APPEARS TO HAVE A CRACK AT THE HUB. NO OTHER INFORMATION WAS REPORTED, BUT HAS BEEN REQUESTED.

    Patient problems: No Information

    MAUDE report 6817066 ↗
  • MalfunctionAug 10, 2017Device Operates Differently Than Expected

    PER SALES REP, THE FACILITY REPORTED A TRIPLE LUMEN PICC WAS BEING PLACED AND ALL THREE LUMEN WERE FLUSHED WITH NO INCIDENT PRIOR TO PLACEMENT. PICC WAS PLACED WITHOUT INCIDENT, RN THEN CHECKED EACH LUMEN FOR ASPIRATION AND FLUSHED. IT WAS STATED THAT DURING FLUSHING THE WHITE SOLO VALVE "POPPED" AND SALINE/BLOOD SPLASHED OVER PLACERS GLOVES, GOWN, AND PATIENT DRAPE. THE LUMEN WAS CLAMPED WITH IV CLAMP. HOSPITALIST WAS NOTIFIED OF POTENTIAL FOR INTRODUCTION OF WHITE HUB/VALVE PARTICLES THROUGH THE LINE, AND THE DECISION WAS MADE THAT THERE WAS MINIMAL RISK, AND THE LINE WAS EXCHANGED OVER THE WIRE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6786084 ↗
  • MalfunctionAug 10, 2017Break

    PER SALES REP, THE FACILITY REPORTED A PICC WHITE LUMEN BROKE DURING PRE-PLACEMENT FLUSH.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6784711 ↗
  • MalfunctionMar 15, 2017Fluid/Blood Leak; Fracture; Material Fragmentation

    ON 2/22/2017 - FACILITY REPORTED TO THE SALES REPRESENTATIVE THAT 2 HOURS POST PLACEMENT, THE CATHETER WAS FLUSHED AND THE HUB "CRUMBLED" IN THE NURSES HAND. DEVICE WAS REMOVED. NO PATIENT INJURY WAS REPORTED. ON 2/23/2017 - ADDITIONAL INFORMATION REPORTED TO THE SALES REP, HEALTHCARE PROFESSIONAL (HCP) PLACED THE PICC LINE AT 5:45 PM, USING 3 CG. HCP STATED THAT THE PLACEMENT WENT WELL AND WAS FUNCTIONING PROPERLY. THE NIGHT ONCOLOGY NURSE WENT TO ACCESS PATENCY OF THE CATHETER AT 8:10 PM, BEFORE INITIATING CHEMOTHERAPY. THE PATIENT STATED THAT BLOOD WAS DRAWN FROM THE FIRST 2 LUMENS AND FLUSHED WITHOUT DIFFICULTY. WHEN THE THIRD (WHITE) LUMEN WAS ACCESSED IT FELL APART IN THE HCP HANDS. THE HCP NOTED THAT THE VALVE AREA WAS COMPRESSED, THE HCP STATED THAT BLOOD WAS EVERYWHERE. THE LINE WAS REMOVED THE FOLLOWING MORNING. THE HCP NOTED THAT THE FRACTURED LINE WAS SECURED WITH A SLIDE CLAMP OFF OF A PERIPHERAL. ALLEGED FRACTURES WERE SEEN ON BOTH SIDES OF THE VALVE HOUSING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6408278 ↗