Poly RadPICC Catheter 3164155Recalls & reported problems
Bard Access Systems, Inc. · Poly RadPICC Catheter 4F Single-Lumen Polyurethane Catheter Microintroducer Basic Kit (70 cm Guidewire) · Class 2 · Implant
All models of Poly RadPICC Catheter (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Poly RadPICC catheter is a long-term central line placed through a vein in the arm. The 54 reports mostly describe the plastic securing clamp detaching, which let the catheter slide partway out of the vein, and one report mentions a break. No deaths or injuries were reported. There are no recalls for this device, and no reports were filed in the past year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00801741034947), received between Aug 9, 2018 and Jan 28, 2023 — classified as 54 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (54 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component54
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY THIRD DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261353 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." ON 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY SIXTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261349 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE TWENTY NINETH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261344 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY FIFTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261343 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE TWENTY SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261338 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM.". 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY EIGHTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261337 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FIFTEENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261334 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261333 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE EIGHTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261329 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE TWENTY FOURTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261328 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY EIGHTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261323 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE SEVENTEENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261322 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261321 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE TENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261319 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY SIXTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261318 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023: FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY FOURTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261317 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY NINTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261316 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY SEVENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261315 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY FOURTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261314 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY NINTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261312 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY SEVENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261311 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM. ON 1/27/2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTY SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261310 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE FORTIETH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261308 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE TWENTY SEVENTH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261307 ↗MalfunctionJan 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE CUSTOMER "THE STAT LOCK PLASTIC DETACHED AND A PICC WAS PULLED OUT 20CM." (B)(6) 2023 - FORTY-NINE DEVICES WERE RETURNED FOR EVALUATION AND ALL EXHIBITED DETACHMENT. THIS REPORT ADDRESSES THE THIRTY SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261305 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page