Broviac 0600040Recalls & reported problems
Bard Access Systems, Inc. · Broviac CV Catheter 2.7F Single-Lumen CV Catheter with SureCuff Tissue Ingrowth Cuff, 52cm Tip-to-Cuff Length Cutdown Tray · Class 2 · Implant
All models of Broviac (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Broviac single-lumen CV catheter is a long-term implanted central line used for vascular access. Most of the 64 reports describe the line leaking or cracking, sometimes breaking inside the patient or the cuff working loose and the catheter slipping out, along with blockages that are hard to flush. Reports mention 16 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 00801741036309), received between Dec 3, 2015 and Jan 12, 2026 — classified as 16 injuries, 48 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak19
- Break11
- Fracture10
- Difficult to Flush5
- Obstruction of Flow5
- Leak/Splash4
- Expulsion3
- Material Puncture/Hole3
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Known Impact Or Consequence To Patient12
- No Consequences Or Impact To Patient10
- Edema4
- No Information4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Extravasation3
- Swelling3
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
InjuryJan 12, 2026Fracture; Material Protrusion/Extrusion
A PATIENT UNDERWENT A PLACEMENT PROCEDURE OF BROVIAC CV CATHETER. POST PROCEDURE AFTER TWELVE DAYS OF PLACEMENT IT WAS REPORTED THAT A CHRONIC CATHETER APPEARS TO BE BROKEN INTERNALLY. THE RIGHT LEG ALSO APPEARS TO BE PUFFY WHERE THE BROVIAC IS PLACED. FURTHER, REMOVED IN INTERVENTIONAL RADIOLOGY. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24033498 ↗MalfunctionApr 9, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
IT WAS REPORTED THAT DURING INSTALLATION OF A CHRONIC CATHETER PLACEMENT PROCEDURE, THE DEVICE RIGID PORTION WAS ALLEGEDLY FOUND LONGER THAN THE ORIGINAL. IT WAS FURTHER REPORTED THAT THE CUFF IS IN A DIFFERENT LOCATION THAN INDICATED ON THE LABELING. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21805046 ↗MalfunctionNov 23, 2024Partial Blockage; Material Discolored; Restricted Flow rate; Deformation Due to Compressive Stress; Contamination /Decontamination Problem; Partial Blockage; Material Discolored; Restricted Flow rate; Deformation Due to Compressive Stress; Contamination /Decontamination Problem
IT WAS REPORTED THAT TEN MONTHS AND NINETEEN DAYS POST A CHRONIC CATHETER PLACEMENT VIA THE RIGHT EXTERNAL JUGULAR VEIN, THE WHITE PLASTIC PART OF THE CONNECTION AND THE TRANSITION AREA OF THE CATHETER HAD ALLEGEDLY TURNED BLACK, WHICH ALLEGEDLY CAUSED THE NARROWING OF THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20766327 ↗MalfunctionOct 22, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT DURING A PREPARATION OF CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER PERFORATION WAS DISCOVERED DURING DRUG ADMINISTRATION. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20503064 ↗InjuryAug 21, 2024Fracture; Material Separation; Expulsion; Fracture; Material Separation; Expulsion
IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE CUFF WAS ALLEGEDLY EXPOSED, SUTURES NOT ADHERENT. IT WAS FURTHER REPORTED THAT THE CATHETER SLID OUT WITHOUT ANY RESISTANCE. REPORTEDLY, THE TIP OF CATHETER WAS SHORT BUT NOT FRAYED, NOT BEVELED. ADDITIONALLY, REMOVED LINE WAS LOOKING SHORT AND SUSPECTED POSSIBILITY OF BROKEN PIECE OF BROVIAC LINE RETAINED AT THE TIME OF LINE REMOVAL. BROKEN PIECE WAS REMOVED BY SURGERY USING INTERVENTIONAL RADIOLOGY. THE PROCEDURE WAS COMPLETED BY REMOVING AND REPLACING THE LINE. CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20023599 ↗MalfunctionJun 29, 2024Fracture; Obstruction of Flow; Burst Container or Vessel; Fracture; Obstruction of Flow
IT WAS REPORTED THAT THREE DAYS POST A CHRONIC CATHETER PLACEMENT, THE TIP OF THE LINE WAS ALLEGEDLY BLOCKED WHICH ALLEGEDLY RESULTED IN RUPTURE OF LINE ON THE INSIDE THE PATIENT. REPORTEDLY, THE LINE WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19640398 ↗MalfunctionMay 23, 2024Difficult to Flush; Obstruction of Flow; Difficult to Flush; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY TWO WEEKS POST A CHRONIC CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY BLOCKED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY DIFFICULT TO BE FLUSHED. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19379402 ↗MalfunctionMay 23, 2024Difficult to Flush; Obstruction of Flow; Difficult to Flush; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY TWO WEEKS POST A CHRONIC CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY BLOCKED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY DIFFICULT TO BE FLUSHED. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19379401 ↗MalfunctionMay 23, 2024Difficult to Flush; Obstruction of Flow; Difficult to Flush; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY TWO WEEKS POST A CHRONIC CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY BLOCKED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY DIFFICULT TO BE FLUSHED. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19379396 ↗MalfunctionMay 23, 2024Difficult to Flush; Obstruction of Flow; Difficult to Flush; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY TWO WEEKS POST A CHRONIC CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY BLOCKED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY DIFFICULT TO BE FLUSHED. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19379395 ↗MalfunctionDec 15, 2023Expulsion; Expulsion
IT WAS REPORTED IN AN ARTICLE IN THE JOURNAL OF THE ¿JAPANESE SOCIETY OF PEDIATRIC SURGEONS¿ TITLED "COMPLICATIONS AND MANAGEMENT OF CENTRAL VENOUS CATHETERS OF LONG-TERM TOTAL PARENTAL NUTRITION", THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT VIA THE LEFT EXTERNAL JUGULAR VEIN, THE PATIENT ALLEGEDLY EXPERIENCED REDNESS AND SWELLING AT THE CATHETER INSERTION SITE EARLY AFTER THE PLACEMENT. IT WAS FURTHER REPORTED THAT THE CUFF OF THE CATHETER WAS ALLEGEDLY EXPELLED, FOLLOWED BY ACCIDENTAL REMOVAL CAUSED BY NOT FIXED CUFF TO THE PATIENT. FURTHERMORE, THERE WAS NO REPORTED ON TREATMENT FOR THE REDNESS AND SWELLING. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18333590 ↗MalfunctionDec 15, 2023Expulsion; Expulsion
IT WAS REPORTED IN AN ARTICLE IN THE JOURNAL OF THE ¿JAPANESE SOCIETY OF PEDIATRIC SURGEONS¿ TITLED "COMPLICATIONS AND MANAGEMENT OF CENTRAL VENOUS CATHETERS OF LONG-TERM TOTAL PARENTAL NUTRITION", THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT VIA THE RIGHT EXTERNAL JUGULAR VEIN, THE CATHETER WAS ALLEGEDLY REMOVED DUE TO AN ACCIDENTAL REMOVAL CAUSED BY NOT FIXED CUFF TO THE PATIENT. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18333588 ↗MalfunctionNov 10, 2023Fracture; Fracture
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE LINE WAS ALLEGEDLY FOUND TO BE SNAPPED. IT WAS FURTHER REPORTED THAT THE LINE WAS CLAMPED AND THE FLUIDS WERE PAUSED. FURTHERMORE, THE FLUIDS WERE CHANGED TO THE PERIPHERAL VENOUS LINE AND THE END OF THE CATHETER WAS STILL ATTACHED TO THE PATIENT COVERED WITH THE TEGADERM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18109373 ↗InjurySep 12, 2023Component Missing; Component Missing
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CUFF WAS ALLEGEDLY MISSING AT THE PROXIMAL END OF THE CATHETER. REPORTEDLY, THE DEVICE ALLEGEDLY FELL OUT IN THE NEONATAL INTENSIVE CARE UNIT. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 17727508 ↗MalfunctionSep 19, 2022Fluid/Blood Leak; Fracture; Material Puncture/Hole; Fluid/Blood Leak; Fracture; Material Puncture/Hole
IT WAS REPORTED THAT SOMETIMES POST PORT PLACEMENT, THE LINE ALLEGEDLY FOUND TO HAVE A LEAK. IT WAS FURTHER REPORTED THAT CATHETER ALLEGEDLY HAD HOLE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15444388 ↗MalfunctionApr 11, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A CATHETER PLACEMENT PROCEDURE, THE TIP OF THE CATHETER WAS ALLEGEDLY DAMAGED DURING INSERTING THE CATHETER OVER THE GUIDEWIRE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14070819 ↗MalfunctionSep 28, 2021Burst Container or Vessel; Fracture; Burst Container or Vessel; Fracture
IT WAS REPORTED THAT SOME TIME POST A CATHETER PLACEMENT, A CRACK WAS ALLEGEDLY FOUND ON THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12541057 ↗MalfunctionAug 11, 2021Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT AFTER CATHETER PLACEMENT, THE CATHETER ALLEGEDLY BROKEN AND BLOOD LEAKED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12305794 ↗InjuryJul 14, 2021Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT APPROXIMATELY SIX HOURS POST A CATHETER PLACEMENT PROCEDURE IN THE RIGHT INTERNAL JUGULAR, THE PATIENT ALLEGEDLY BLED. UPON INSPECTION, THE CATHETER WAS FOUND TO BE CRACKED. THE DEVICE WAS REMOVED AND REPLACED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 12164821 ↗InjurySep 11, 2020Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT APPROXIMATELY ONE WEEK POST CHRONIC CATHETER IMPLANTATION, THE DEVICE ALLEGEDLY BROKEN. IT WAS FURTHER REPORTED THAT PATIENT DEVELOPED EDEMA UNDER THE SKIN AND INFILTRATION OF THE SKIN. THE DEVICE WAS REMOVED AND REPLACED. THE CURRENT PATIENT STATUS IS UNKNOWN.
Patient problems: Edema
MAUDE report 10520179 ↗InjuryJun 5, 2020Loss of or Failure to Bond; Fluid/Blood Leak; Loss of or Failure to Bond; Fluid/Blood Leak
IT WAS REPORTED THAT DAY AFTER POST CATHETER PLACEMENT, THE CENTRAL VENOUS CATHETER ALLEGEDLY LEAKED THE CONTRAST MEDIUM. IT WAS FURTHER REPORTED THAT, IT WAS CRACKED DURING REMOVAL. THE PATIENT STATUS IS UNKNOWN.
Patient problems: Edema
MAUDE report 10122742 ↗InjuryJun 5, 2020Loss of or Failure to Bond; Fluid/Blood Leak; Loss of or Failure to Bond; Fluid/Blood Leak
IT WAS REPORTED THAT DAY AFTER POST CATHETER PLACEMENT, THE CENTRAL VENOUS CATHETER WAS ALLEGEDLY LEAKED THE CONTRAST MEDIUM. IT WAS FURTHER REPORTED THAT THE CUFF WAS CRACKED DURING REMOVAL. THE PATIENT STATUS IS UNKNOWN.
Patient problems: Edema
MAUDE report 10122381 ↗InjuryMay 22, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED APPROXIMATELY SIX WEEKS POST CHRONIC CATHETER IMPLANTATION THE DEVICE ALLEGEDLY LEAKED. THE PATIENT EXPERIENCED SWELLING AND WAS TREATED WITH ANTIBIOTICS, HOWEVER THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Swelling
MAUDE report 10084214 ↗MalfunctionJan 14, 2020Fluid/Blood Leak; Failure to Infuse; Fluid/Blood Leak; Failure to Infuse
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0600040 CHRONIC CATHETER ALLEGEDLY EXPERIENCED SUBCUTANEOUS LEAK WITHOUT EMBOLIZATION AND FAILURE TO INFUSE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED ONE PATIENT WITH NO CONSEQUENCES. THE PATIENT'S AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9586490 ↗MalfunctionOct 16, 2019Fluid/Blood Leak; Fluid/Blood Leak
THIS EVENT SUMMARIZES ONE MALFUNCTION. THE INFORMATION REVIEWED INDICATED THAT MODEL 0600040 CHRONIC CATHETER ALLEGEDLY EXPERIENCED A SPLIT AND LEAKED. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THE PATIENT'S AGE, WEIGHT, AND GENDER WERE NOT PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9199607 ↗
FDA clearance and classification
The FDA regulates product code LJS (catheter,intravascular,therapeutic,long-term greater than 30 days) under 21 CFR 880.5970 as a Class II device. 21 CFR 880.5970 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Broviac 0600040
Has the Broviac 0600040 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741036309), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Broviac 0600040?
The FDA's MAUDE database lists 64 adverse event reports for this device, received between Dec 3, 2015 and Jan 12, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (19), “Break” (11) and “Fracture” (10); the most common patient problems are “Edema” (4), “Extravasation” (3) and “Swelling” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Broviac 0600040?
None of the reports are classified as deaths, 16 as injuries and 48 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Broviac 0600040 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Broviac 0600040, and is it still sold?
The FDA's device database lists Bard Access Systems, Inc. as the company responsible for it (brand Broviac, model 0600040). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,692 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,507 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall398 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls244 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall263 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days276 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page