Skip to content

PowerLine 0730615Recalls & reported problems

Bard Access Systems, Inc. · PowerLine Central Venous Catheter. 6F Triple-Lumen Polyurethane Catheter with SureCuff Ingrowth Cuff. Microintroducer Kit · Class 2 · Implant

All models of PowerLine (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00801741036637), received between Aug 8, 2016 and Apr 3, 2026 — classified as 4 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Fracture5
  • Loss of or Failure to Bond3
  • Material Separation3
  • Deformation Due to Compressive Stress2
  • Fluid/Blood Leak2
  • Migration2
  • Air/Gas in Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Bacterial Infection2
  • No Known Impact Or Consequence To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Fungal Infection1
  • No Consequences Or Impact To Patient1
  • Therapeutic Effects, Unexpected1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Fracture; Material Separation; Deformation Due to Compressive Stress

    ON (B)(6) 2026, THE PATIENT WAS PREPPED A CATHETER PLACEMENT PROCEDURE USING THE POWERLINE CV CATHETER PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT TWO TRIPLE LUMEN POWERLINES WERE OPENED AND FOUND TO HAVE DAMAGED TUNNELERS. THE END THAT IS INSERTED INTO THE CATHETER WAS BENT, AND WHEN ATTEMPTS WERE MADE TO STRAIGHTEN THEM, THE TUNNELERS SNAPPED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24774396 ↗
  • MalfunctionApr 1, 2026Fracture; Material Separation; Deformation Due to Compressive Stress

    ON (B)(6) 2026, THE PATIENT WAS PREPPED A CATHETER PLACEMENT PROCEDURE USING THE POWERLINE CV CATHETER PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT TWO TRIPLE LUMEN POWERLINES WERE OPENED AND FOUND TO HAVE DAMAGED TUNNELERS. THE END THAT IS INSERTED INTO THE CATHETER WAS BENT, AND WHEN ATTEMPTS WERE MADE TO STRAIGHTEN THEM, THE TUNNELERS SNAPPED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24754909 ↗
  • MalfunctionDec 20, 2023Failure to Disconnect; Scratched Material; Failure to Disconnect; Scratched Material

    IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT VIA THE LEFT SUBCLAVIAN VEIN, DURING THE ROUTINE NEEDLE-FREE CONNECTORS CHANGE, CLAMPS WERE NEEDED TO BE PLACED DIRECTLY ON THE COLORED PLASTIC PARTS OF THE CATHETER TO REMOVE THE CONNECTORS, WHICH ALLEGEDLY RESULTED IN A STRIPPING DAMAGE TO THE COLORED PLASTIC PIECES. IT WAS FURTHER REPORTED THAT NEEDLE-FREE CONNECTORS WOULD NOT LOOSEN WITHOUT A CLAMP ON THE NEEDLE-FREE CONNECTORS AND ON THE COLORED PLASTIC PIECE OF THE CATHETER AND TURNING THE CLAMPS IN OPPOSITE DIRECTIONS. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18361911 ↗
  • MalfunctionMar 27, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO MONTHS AND THREE DAYS POST CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY LEAKED WHERE THE LINE CONNECTS TO THE INJECTION PORT. IT WAS FURTHER REPORTED THAT THE LINE WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16618303 ↗
  • InjuryNov 16, 2022Migration; Migration

    IT WAS REPORTED THAT TWO DAYS POST CHRONIC CATHETER PLACEMENT PROCEDURE, THE LINE ALLEGEDLY MIGRATED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15803651 ↗
  • MalfunctionNov 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SOMETIME POST CHRONIC CATHETER PLACEMENT PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED INFECTION. REPORTEDLY, THERE IS NO INFORMATION REGARDING THE TREATMENT OF THE INFECTION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15750452 ↗
  • MalfunctionNov 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SOMETIME POST CHRONIC CATHETER PLACEMENT PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED INFECTION. REPORTEDLY, THERE IS NO INFORMATION REGARDING THE TREATMENT OF THE INFECTION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15750369 ↗
  • InjuryOct 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH RESULTS OF A CLINICAL TRIAL THAT APPROXIMATELY THREE MONTHS AND NINETEEN DAYS POST INDEX PROCEDURE COMPLETION USING A CHRONIC CATHETER PLACEMENT, THE SUBJECT DEVELOPED ADVERSE EVENT OF BACTEREMIA. IT WAS FURTHER REPORTED THAT APPROXIMATELY THREE MONTHS AND TWENTY DAYS POST INDEX PROCEDURE, EVENT REQUIRED RE-INTERVENTION AND THE CATHETER WAS REMOVED, AND EXCHANGED. THE CURRENT STATUS OF THE PATIENT WAS NOT PROVIDED.

    Patient problems: Bacterial Infection

    MAUDE report 15664891 ↗
  • InjurySep 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL, THAT APPROXIMATELY FOUR MONTHS AND TWENTY-SEVEN DAYS POST CATHETER PLACEMENT, THE SUBJECT DEVELOPED ADVERSE EVENT OF CANDIDEMIA. IT WAS FURTHER REPORTED THAT THE CATHETER WAS REMOVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Fungal Infection

    MAUDE report 15500180 ↗
  • MalfunctionDec 31, 2021Loss of or Failure to Bond; Fluid/Blood Leak; Fracture; Material Separation; Air/Gas in Device; Loss of or Failure to Bond; Fluid/Blood Leak; Fracture; Material Separation; Air/Gas in Device

    IT WAS REPORTED THAT FOUR MONTHS POST CHRONIC CATHETER PLACEMENT, A PIN HOLE SIZED LEAK WAS ALLEGEDLY NOTED ON THE LINES. IT WAS FURTHER REPORTED THAT AIR WAS NOTED TO ENTER WITH INFUSION. THE LINE WAS REPORTED TO BE CLAMPED OFF AND COVERED WITH TEGADERM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13123722 ↗
  • MalfunctionDec 6, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT SOME TIME POST CHRONIC CATHETER PLACEMENT, THE CATHETER CUFF WAS ALLEGEDLY MISSING UPON REMOVAL FORM PATIENT BODY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12932020 ↗
  • InjurySep 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT POST DEVICE IMPLANT, THE PATIENT ALLEGEDLY EXPERIENCED INFECTION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Bacterial Infection

    MAUDE report 12431490 ↗
  • MalfunctionSep 1, 2021Fracture; Fracture

    IT WAS REPORTED THAT POST DEVICE IMPLANT, THE DEVICE ALLEGEDLY FRACTURED AT THE CLAMPING SITES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12405248 ↗
  • MalfunctionDec 29, 2020Loss of or Failure to Bond; Migration; Loss of or Failure to Bond; Migration

    IT WAS REPORTED THAT DURING A CATHETER PLACEMENT PROCEDURE, THE CUFF ALLEGEDLY FELL OFF FROM THE CATHETER. IT WAS FURTHER REPORTED THAT THE CUFF WAS REMOVED. THE PATIENT STATUS WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11085678 ↗
  • MalfunctionApr 14, 2020Fracture; Fracture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0730615 CHRONIC CATHETER ALLEGEDLY EXPERIENCED FRACTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCES. AGE, WEIGHT, AND SEX WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9960426 ↗
  • MalfunctionJun 16, 2017Device Operates Differently Than Expected

    NURSE REPORTED TO MS&S, THAT THEY TYPICALLY USE THE DUAL POWERLINE AND AS PART OF THEIR PROTOCOL THEY ADMINISTER A DRUG CALLED PROGRAF (TACROLIMUS) IV AND THEN DRAW DRUG LEVELS OVER A SERIES OF WEEKS FROM A DIFFERENT LUMEN AND HAVE NOT HAD ANY ISSUES WITH ¿SKEWED VALUES¿. THE FACILITY RECENTLY TRIED THE TRIPLE LUMEN POWERLINE WITH TWO PATIENTS AND THEY HAD EXTREMELY ELEVATED DRUG LEVELS WITH BLOOD DRAW THROUGH THE LINE EVEN THREE WEEKS AFTER THE INFUSION. THE NURSE STATED THAT IN ONE CASE THE DRUG LEVEL AS 30, WHICH IS EXTREMELY HIGH AND WHEN THEY DID A PERIPHERAL BLOOD DRAW IT WAS ONLY 2. THE NURSE WAS WONDERING IF THE DRUG IS LEECHING THROUGH THE POLYURETHANE MATERIAL DUE TO THE WALL THICKNESS OR ALCOHOL CONTENT. IT ALSO HAS CASTOR OIL IN THE DRUG WHICH COULD LEAD TO IT COATING THE TUBING; HOWEVER, THEY ARE DRAWING BLOOD FROM A DIFFERENT LUMEN NOT THE INFUSION LUMEN. ANOTHER POSSIBILITY IS THAT A SMALL FIBRIN TAIL OR SHEATH COULD BE DIRECTING THE DRUG FROM ONE LUMEN INTO THE OTHER SINCE THEY ARE VERY CLOSE TOGETHER. THE FACILITY HAS NOT EXPERIENCED THIS ISSUE WITH THE DUAL LUMEN POWERLINES. NO REPORTED PATIENT HARM. THIS FILE ADDRESSES PATIENT TWO.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6647699 ↗
  • MalfunctionJun 16, 2017Unexpected Therapeutic Results; High Test Results

    NURSE REPORTED TO MS&S, THAT THEY TYPICALLY USE THE DUAL POWERLINE AND AS PART OF THEIR PROTOCOL THEY ADMINISTER A DRUG CALLED PROGRAF (TACROLIMUS) IV AND THEN DRAW DRUG LEVELS OVER A SERIES OF WEEKS FROM A DIFFERENT LUMEN AND HAVE NOT HAD ANY ISSUES WITH ¿SKEWED VALUES¿. THE FACILITY RECENTLY TRIED THE TRIPLE LUMEN POWERLINE WITH TWO PATIENTS AND THEY HAD EXTREMELY ELEVATED DRUG LEVELS WITH BLOOD DRAW THROUGH THE LINE EVEN THREE WEEKS AFTER THE INFUSION. THE NURSE STATED THAT IN ONE CASE THE DRUG LEVEL AS 30, WHICH IS EXTREMELY HIGH AND WHEN THEY DID A PERIPHERAL BLOOD DRAW IT WAS ONLY 2. THE NURSE WAS WONDERING IF THE DRUG IS LEECHING THROUGH THE POLYURETHANE MATERIAL DUE TO THE WALL THICKNESS OR ALCOHOL CONTENT. IT ALSO HAS CASTOR OIL IN THE DRUG WHICH COULD LEAD TO IT COATING THE TUBING; HOWEVER, THEY ARE DRAWING BLOOD FROM A DIFFERENT LUMEN NOT THE INFUSION LUMEN. ANOTHER POSSIBILITY IS THAT A SMALL FIBRIN TAIL OR SHEATH COULD BE DIRECTING THE DRUG FROM ONE LUMEN INTO THE OTHER SINCE THEY ARE VERY CLOSE TOGETHER. THE FACILITY HAS NOT EXPERIENCED THIS ISSUE WITH THE DUAL LUMEN POWERLINES. NO REPORTED PATIENT HARM. THIS FILE ADDRESSES PATIENT ONE.

    Patient problems: Therapeutic Effects, Unexpected

    MAUDE report 6647581 ↗
  • MalfunctionAug 8, 2016Leak/Splash

    IT WAS REPORTED THAT A POWERLINE WAS INSERTED ON (B)(6) 2016 BUT A HOLE WAS NOTICED AT A FOLLOW-UP. THE RADIOLOGY DEPARTMENT WAS INFORMED AND THE LINE WAS REMOVED ON (B)(6) 2016. THE LEAK WAS IDENTIFIED A COUPLE OF CENTIMETERS ABOVE THE CUFF IN THE SUBCUTANEOUS TUNNEL. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5855216 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page