Navarre Opti-Drain NOD12LPTRecalls & reported problems

Bard Access Systems, Inc. · Navarre Opti-Drain Multi-Use Drainage Catheter, Locking Pigtail, 12 Fr x 30 cm · Class 1

All models of Navarre Opti-Drain (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Opti-Drain is a locking pigtail drainage catheter used to drain fluid from places such as the kidney, liver or cysts. There are 67 adverse event reports, mostly describing the catheter breaking or cracking during use, with pieces separating, leaking fluid, a blocked inner channel, and a locking thread that would not hold. The reports mention no deaths and 4 injuries. No recalls name this device, and reports in the past 12 months fell from 30 to 13.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 00801741037061), received between Nov 19, 2018 and May 18, 2026 — classified as 4 injuries, 63 malfunctions.

The most frequently reported specific device problem is “Fracture” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 5 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 5 reportsMar 2025: 2 reportsApr 2025: 6 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 9 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture24
  • Break22
  • Material Separation20
  • Fluid/Blood Leak16
  • Nonstandard Device6
  • Material Integrity Problem5
  • Deformation Due to Compressive Stress4
  • Device Damaged Prior to Use4

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • No Consequences Or Impact To Patient4
  • Pneumothorax3
  • Extravasation1
  • Wound Infection1

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Nonstandard Device

    ON (B)(6) 2026, A PATIENT WAS PREPPED FOR AN ULTRASOUND GUIDED CYST PERCUTANEOUS DRAINAGE PROCEDURE USING THE NAVARRE OPTI DRAIN CATHETER. PRIOR TO THE PROCEDURE, IT WAS NOTED THAT THE LENGTH OF THE TROCAR NEEDLE WAS SHORTER THAN THE CATHETER. THE DRAINAGE CATHETER WAS REPLACED WITH A NEW DRAINAGE CATHETER TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25197364 ↗
  • MalfunctionMay 18, 2026Nonstandard Device

    ON (B)(6) 2026, THE PATIENT WAS PREPPED FOR AN ULTRASOUND GUIDED PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAPHY DRAINAGE (PTCD) CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN CATHETER. PRIOR TO THE PROCEDURE, IT WAS NOTED THAT THE LENGTH OF TROCAR NEEDLE WAS ALLEGEDLY SHORTER THAN THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25197321 ↗
  • MalfunctionMar 23, 2026Failure to Advance; Blocked Connection; Material Integrity Problem

    ON (B)(6) 2026, THE PATIENT UNDERWENT AN ULTRASOUND-GUIDED CYST PERCUTANEOUS DRAINAGE PROCEDURE USING A NAVARRE OPTI DRAIN CATHETER. DURING THE PROCEDURE, THE CLINICIAN INSPECTED THE PACKAGE AND PRODUCT PRIOR TO USE AND CONFIRMED THAT BOTH WERE INTACT. HOWEVER, DURING THE PROCEDURE, IT WAS IDENTIFIED THAT THE MIDDLE CHANNEL OF THE DRAINAGE DEVICE WAS BLOCKED, WHICH PREVENTED SMOOTH INSERTION OF THE GUIDE WIRE. IT WAS FURTHER REPORTED THAT DAMAGE WAS NOTED TO THE TUBE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24659471 ↗
  • MalfunctionFeb 9, 2026Break; Material Separation

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING A NAVARRE OPTI DRAIN CATHETER. POST THE PROCEDURE, THE SURE-LOK TM THREAD HAD DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288725 ↗
  • MalfunctionFeb 9, 2026Break; Material Separation

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING A NAVARRE OPTI DRAIN CATHETER. POST THE PROCEDURE, THE SURE-LOK TM THREAD HAD DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288724 ↗
  • MalfunctionFeb 9, 2026Break; Material Separation

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING A NAVARRE OPTI DRAIN CATHETER. POST THE PROCEDURE, THE SURE-LOK TM THREAD HAD DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288710 ↗
  • MalfunctionJan 20, 2026Nonstandard Device

    ON (B)(6) 2025, A PATIENT UNDERWENT FOR AN ULTRASOUND GUIDED CYST PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. PRIOR TO THE PROCEDURE, THE LENGTH OF THE TROCAR NEEDLE WAS ALLEGEDLY SHORTER THAN THE CATHETER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24113203 ↗
  • MalfunctionJan 20, 2026Nonstandard Device

    ON (B)(6) 2025, A PATIENT PREPPED FOR AN ULTRASOUND-GUIDED CYST DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE LENGTH OF THE TROCAR NEEDLE WAS ALLEGEDLY SHORTER THAN THE CATHETER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24112769 ↗
  • MalfunctionJan 20, 2026Nonstandard Device

    ON (B)(6) 2025, A PATIENT PREPPED FOR AN ULTRASOUND-GUIDED CYST DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE LENGTH OF THE TROCAR NEEDLE WAS ALLEGEDLY SHORTER THAN THE CATHETER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24112493 ↗
  • MalfunctionJan 9, 2026Fluid/Blood Leak; Fracture

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. POST THE PROCEDURE, AFTER RETURNING TO THE WARD POSTOPERATIVELY, LEAKAGE OCCURRED AT THE CONNECTION POINT OF THE CATHETER TIP. THE LEAKING AT THE CONNECTOR WAS CAUSED BY A CRACKED DRAIN TUBING. THE CONNECTING TUBE WAS IMMEDIATELY REMOVED, AND A RUBBER CONNECTOR WAS REAPPLIED TO SEAL THE TEAR AT THE CATHETER TIP. THE ISSUE WAS RESOLVED BY REPLACING THE LEAKING RUBBER CONNECTOR. THERE WERE NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24018820 ↗
  • MalfunctionDec 5, 2025Difficult to Insert; Nonstandard Device; Failure to Advance; Difficult to Insert; Nonstandard Device; Failure to Advance

    ON (B)(6) 2025, A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. DURING THE PROCEDURE, THE DEVICE WAS ALLEGEDLY TOO SHORT, MAKING IMPOSSIBLE TO INSERT THE GUIDEWIRE SMOOTHLY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23725800 ↗
  • MalfunctionDec 3, 2025Fracture; Fracture

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED HYDROTHORAX PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN CATHETER. POST THE PROCEDURE, A BREAK OCCURRED IN THE BODY OF THE DRAINAGE CATHETER. THE CATHETER WAS REPLACED WITH A NEW DRAINAGE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23704150 ↗
  • InjuryNov 18, 2025Fluid/Blood Leak; Fracture; Material Separation; Air/Gas in Device; Fluid/Blood Leak; Fracture; Material Separation; Air/Gas in Device

    ON (B)(6) 2025, A PATIENT DIAGNOSED WITH THYMOMA UNDERWENT THORACOSCOPIC EXTENDED THYMECTOMY WITH RIGHT LOWER LOBE WEDGE RESECTION USING THE NAVARRE OPTI DRAIN. POST PROCEDURE, A TUBE WAS RETAINED FOR PLEURAL FLUID AND AIR DRAINAGE. LATER A MINOR LEAKAGE WAS OBSERVED AT THE CHEST TUBE CONNECTOR, AND EXAMINATION REVEALED A FRACTURE IN THE RIGID SECTION OF THE CONNECTOR. THE PATIENT REPORTED MILD CHEST TIGHTNESS WITH OXYGEN SATURATION AT 95%, AND AUSCULTATION SHOWED SLIGHTLY DIMINISHED BREATH SOUNDS ON THE RIGHT SIDE. THE DRAINAGE TUBE CONNECTOR WAS REPLACED, AND A CHEST X-RAY REVEALED A SMALL PNEUMOTHORAX. THE PATIENT WAS INSTRUCTED TO COUGH TO EXPEL TRAPPED AIR, AFTER WHICH CHEST TIGHTNESS RESOLVED, OXYGEN SATURATION IMPROVED TO 99%, AND BREATH SOUNDS NORMALIZED. THE PATIENT REMAINED UNDER OBSERVATION AND WAS LATER DISCHARGED WITHOUT FURTHER COMPLICATIONS. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Pneumothorax

    MAUDE report 23580834 ↗
  • MalfunctionOct 30, 2025Fluid/Blood Leak; Fracture; Material Separation; Fluid/Blood Leak; Fracture; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT A CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI-DRAIN DRAINAGE CATHETER. IT WAS REPORTED THAT POST CATHETER PLACEMENT, THE DRAINAGE CATHETER CONNECTOR WAS ALLEGEDLY FOUND FRACTURED. REPORTEDLY, DRAINAGE CATHETER CONNECTOR WAS ALLEGEDLY FOUND LEAKING POINT IS LOCATED AT THE JUNCTION BETWEEN THE CATHETER BODY AND THE WHITE CONNECTOR. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23424245 ↗
  • MalfunctionOct 16, 2025Fluid/Blood Leak; Fluid/Blood Leak

    A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE USING THE NAVARRE OPTI DRAIN. SOMETIMES POST A PROCEDURE ON (B)(6) 2025, A LEAK WAS IDENTIFIED AT THE CONNECTION POINT OF A SINGLE-USE DRAINAGE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23307653 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE USING THE NAVARRE OPTI-DRAIN. POST A PORT PLACEMENT PROCEDURE, IT WAS FOUND THAT THE SURE-LOKTM THREAD WAS ALLEGEDLY FOUND DIGRESSION. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23120569 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. POST THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO REPORTED PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23119810 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. POST THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO REPORTED PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23119809 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND-GUIDED PCN PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE OPTI DRAIN. POST THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO REPORTED PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23119698 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE USING THE NAVARRE OPTI-DRAIN. POST A PORT PLACEMENT PROCEDURE, IT WAS FOUND THAT THE SURE-LOKTM THREAD WAS ALLEGEDLY FOUND DIGRESSION. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23119697 ↗
  • MalfunctionSep 23, 2025Break; Material Separation; Break; Material Separation

    A PATIENT UNDERWENT A DRAINAGE CATHETER PLACEMENT PROCEDURE USING THE NAVARRE OPTI-DRAIN. POST A PORT PLACEMENT PROCEDURE, IT WAS FOUND THAT THE SURE-LOKTM THREAD WAS ALLEGEDLY FOUND DIGRESSION. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23119696 ↗
  • MalfunctionSep 22, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. AFTER THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSIONED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23107969 ↗
  • MalfunctionSep 22, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. AFTER THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23107715 ↗
  • MalfunctionSep 22, 2025Break; Material Separation; Break; Material Separation

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. AFTER THE PROCEDURE, THE SURE-LOKTM THREAD WAS ALLEGEDLY DIGRESSIONED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23107714 ↗
  • MalfunctionJul 23, 2025Fracture; Fracture

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDED ASCITES PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE. IT WAS REPORTED THAT SOMETIMES POST CATHETER PLACEMENT PROCEDURE, THE DRAINAGE CATHETER CONNECTOR WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22574411 ↗